Potential and frequency effects on fretting corrosion of Ti6Al4V and CoCrMo surfaces

被引:71
|
作者
Swaminathan, Viswanathan [1 ,2 ]
Gilbert, Jeremy L. [1 ,2 ]
机构
[1] Syracuse Univ, Dept Biomed & Chem Engn, Syracuse, NY 13244 USA
[2] Syracuse Univ, Syracuse Biomat Inst, Syracuse, NY USA
关键词
fretting corrosion; metal-on-metal; friction; modularity; tribocorrosion; ORTHOPEDIC IMPLANTS; TITANIUM; BEHAVIOR; ALLOY; SPECTROSCOPY; PRODUCTS; XPS;
D O I
10.1002/jbm.a.34564
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fretting corrosion has been reported at the metal-metal interfaces of a wide range of medical devices, including total joint replacements, spinal devices, and overlapping cardiovascular stents. Currently, the fretting corrosion phenomenon associated with metal-on-metal contacts is not fully understood. This study investigated the effect of potential and fretting frequency on the fretting corrosion performance of Ti6Al4V/Ti6Al4V, Ti6Al4V/CoCrMo, and CoCrMo/CoCrMo alloy combinations at fixed normal load and displacement conditions using a custom built fretting corrosion test system. The results showed that the fretting current densities increased with increases in potential and were highest for Ti6Al4V/Ti6Al4V couple (1.5 mA/cm(2) at 0 V vs. Ag/AgCl). The coefficient of friction varied with potential and was about two times higher for Ti6Al4V/Ti6Al4V (0.71 V at 0 V vs. Ag/AgCl). In most of the potential range tested, the fretting corrosion behavior of CoCrMo/Ti6Al4V and CoCrMo/CoCrMo was similar and dominated by the CoCrMo surface. Increase in applied fretting frequency linearly increased the fretting current densities in the regions where the passive film is stable. Also, the model-based fretting current densities were in excellent agreement with the experimental results. Overall, Ti6Al4V/Ti6Al4V couple was more susceptible to fretting corrosion compared with other couples. However, the effects of these processes on the biological system were not assessed. (c) 2013 Wiley Periodicals, Inc. J Biomed Mater Res Part A: 101A: 2602-2612, 2013.
引用
收藏
页码:2602 / 2612
页数:11
相关论文
共 50 条
  • [31] Improved thrombogenicity on oxygen etched Ti6Al4V surfaces
    Riedel, Nicholas A.
    Smith, Barbara S.
    Williams, John D.
    Popat, Ketul C.
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2012, 32 (05): : 1196 - 1203
  • [32] Oxidation and corrosion protection by halide treatment of powder metallurgy Ti and Ti6Al4V alloy
    Tsipas, S. A.
    Gordo, E.
    Jimenez-Morales, A.
    CORROSION SCIENCE, 2014, 88 : 263 - 274
  • [33] Tribology and hydrophobicity of a biocompatible GPTMS/PFPE coating on Ti6Al4V surfaces
    Panjwani, Bharat
    Sinha, Sujeet. K.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 15 : 103 - 111
  • [34] A Novel Treatment to Selectively Harden Ti6Al4V Surfaces
    Wei, Qi
    Zhang, Zhenxue
    ACS OMEGA, 2023, 8 (27): : 24552 - 24560
  • [35] Selective modification of Ti6Al4V surfaces for biomedical applications
    Rodriguez, Gabriela Melo
    Bowen, James
    Zelzer, Mischa
    Stamboulis, Artemis
    RSC ADVANCES, 2020, 10 (30) : 17642 - 17652
  • [36] Study of the Effect of Nanosecond Laser Texturing on the Corrosion Behavior of Ti6Al4V and Ti6Al4V Parts Produced by Powder Bed Fusion
    Queiroz, Fernanda Martins
    Ribeiro, Gleicy de Lima Xavier
    de Castro, Renato Spacini
    dos Santos, Rogerio Goes
    Vieira, Alexandre
    Terada, Maysa
    Bugarin, Aline de Fatima Santos
    de Rossi, Wagner
    Costa, Isolda
    MATERIALS PERFORMANCE AND CHARACTERIZATION, 2023, 12 (03) : 341 - 350
  • [37] Prediction of crack initiation and propagation during fretting corrosion of novel Ti6Al4V specimens in saline and physiological environments
    Anjum, Zeeshan
    Shah, Masood
    JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2023, 37 (01) : 119 - 126
  • [38] Effects of harmonic structure on the electrochemical behavior of biomedical Ti6Al4V
    Reyes Leiva, Karla Miriam
    dos Santos, Bruno Borges L.
    Vicente Junior, Nerio
    Bruno Marino, Claudia Eliana
    MATERIALS TODAY-PROCEEDINGS, 2020, 33 : 1804 - 1808
  • [39] Fretting corrosion behaviour of Ti-6Al-4V reinforced with zirconia in foetal bovine serum
    Semetse, Lerato
    Obadele, Babatunde Abiodun
    Raganya, Lerato
    Geringer, Jean
    Olubambi, Peter Apata
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2019, 100
  • [40] Prediction of crack initiation and propagation during fretting corrosion of novel Ti6Al4V specimens in saline and physiological environments
    Zeeshan Anjum
    Masood Shah
    Journal of Mechanical Science and Technology, 2023, 37 : 119 - 126