Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes

被引:178
作者
Achenbach, S [1 ]
Moshage, W [1 ]
Diem, B [1 ]
Bieberle, T [1 ]
Schibgilla, V [1 ]
Bachmann, K [1 ]
机构
[1] UNIV ERLANGEN NURNBERG,INST BIOMED TECHN,D-91054 ERLANGEN,GERMANY
关键词
D O I
10.1016/S0002-8703(97)70083-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In phantom studies we investigated the effects of magnetic resonance imaging (MRI) on pacemakers and electrodes. Twenty-five electrodes were exposed to MRI in a 1.5T scanner with continuous registration of the temperature at the electrode tip. Eleven pacemakers (five single chamber and six dual chamber) were exposed to MRI. Pacemaker output was monitored to detect malfunction in VOO/DOO and WI/DDD modes, A temperature increase at the electrode tip of up to 63.1 degrees C was observed during 90 seconds of scanning. In seven electrodes the temperature increase exceeded 15 degrees C. Although no pacemaker malfunctions were observed in asynchronous pacing mode (VOO/D00), inhibition and rapid pacing were observed during spin-echo imaging if the pacemakers were set to VVI or DDD mode. Pacemaker function was not impaired during scanning with gradient-echo sequences. Next to pacemaker dysfunction, electrode heating has to be considered a possible adverse effect when exposing patients with pacemakers to MRI.
引用
收藏
页码:467 / 473
页数:7
相关论文
共 20 条
  • [11] NUCLEAR-MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A PACEMAKER
    INBAR, S
    LARSON, J
    BURT, T
    MAFEE, M
    EZRI, MD
    [J]. AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 1993, 305 (03) : 174 - 175
  • [12] JOHNSON D, 1994, PACE, V17, P772
  • [13] EFFECTS OF NUCLEAR-MAGNETIC-RESONANCE IMAGING ON CARDIAC-PACEMAKERS
    LAUCK, G
    VONSMEKAL, A
    WOLKE, S
    SEELOS, KC
    JUNG, W
    MANZ, M
    LUDERITZ, B
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (08): : 1549 - 1555
  • [14] LAUCK G, 1995, PACE S, V18, P1168
  • [15] MOSHLEY I, 1994, BRIT MED J, V308, P1181
  • [16] THE EFFECTS OF NUCLEAR MAGNETIC-RESONANCE ON PATIENTS WITH CARDIAC-PACEMAKERS
    PAVLICEK, W
    GEISINGER, M
    CASTLE, L
    BORKOWSKI, GP
    MEANEY, TF
    BREAM, BL
    GALLAGHER, JH
    [J]. RADIOLOGY, 1983, 147 (01) : 149 - 153
  • [17] VENTRICULAR-FIBRILLATION DURING TESTING OF A DEMAND PACEMAKER
    SEIPEL, L
    BUB, E
    DRIWAS, S
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1975, 100 (47) : 2439 - 2442
  • [18] POLICIES, GUIDELINES, AND RECOMMENDATIONS FOR MR IMAGING SAFETY AND PATIENT-MANAGEMENT
    SHELLOCK, FG
    KANAL, E
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (01): : 97 - 101
  • [19] TOBISCH RJ, 1993, BIOMED TECH BERL S, V38, P435
  • [20] ZIMMERMANN BH, 1984, DIAGN IMAG CLIN MED, V53, P53