Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes

被引:179
作者
Achenbach, S [1 ]
Moshage, W [1 ]
Diem, B [1 ]
Bieberle, T [1 ]
Schibgilla, V [1 ]
Bachmann, K [1 ]
机构
[1] UNIV ERLANGEN NURNBERG,INST BIOMED TECHN,D-91054 ERLANGEN,GERMANY
关键词
D O I
10.1016/S0002-8703(97)70083-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In phantom studies we investigated the effects of magnetic resonance imaging (MRI) on pacemakers and electrodes. Twenty-five electrodes were exposed to MRI in a 1.5T scanner with continuous registration of the temperature at the electrode tip. Eleven pacemakers (five single chamber and six dual chamber) were exposed to MRI. Pacemaker output was monitored to detect malfunction in VOO/DOO and WI/DDD modes, A temperature increase at the electrode tip of up to 63.1 degrees C was observed during 90 seconds of scanning. In seven electrodes the temperature increase exceeded 15 degrees C. Although no pacemaker malfunctions were observed in asynchronous pacing mode (VOO/D00), inhibition and rapid pacing were observed during spin-echo imaging if the pacemakers were set to VVI or DDD mode. Pacemaker function was not impaired during scanning with gradient-echo sequences. Next to pacemaker dysfunction, electrode heating has to be considered a possible adverse effect when exposing patients with pacemakers to MRI.
引用
收藏
页码:467 / 473
页数:7
相关论文
共 20 条
  • [1] Abart J, 1995, Aktuelle Radiol, V5, P376
  • [2] ACHENBACH S, 1995, Z KARDIOL S, V84, P119
  • [3] NUCLEAR MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A DDD PACEMAKER
    ALAGONA, P
    TOOLE, JC
    MANISCALCO, BS
    GLOVER, MU
    ABERNATHY, GT
    PRIDA, XE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04): : 619 - 619
  • [4] De Cock Carel C., 1995, European Heart Journal, V16, P401
  • [5] EFFECT OF MAGNETIC-RESONANCE-IMAGING ON DDD PACEMAKERS
    ERLEBACHER, JA
    CAHILL, PT
    PANNIZZO, F
    KNOWLES, RJR
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) : 437 - 440
  • [6] THE EFFECTS OF NUCLEAR MAGNETIC-RESONANCE IMAGERS ON EXTERNAL AND IMPLANTABLE PULSE GENERATORS
    FETTER, J
    ARAM, G
    HOLMES, DR
    GRAY, JE
    HAYES, DL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (04): : 720 - 727
  • [7] GIMBEL JR, 1995, J AM COLL CARDIOL, V25, pA11
  • [8] EFFECT OF 1.5 TESLA NUCLEAR-MAGNETIC-RESONANCE IMAGING SCANNER ON IMPLANTED PERMANENT PACEMAKERS
    HAYES, DL
    HOLMES, DR
    GRAY, JE
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (04) : 782 - 786
  • [9] HEIKN JP, 1991, MANUAL CLIN MRI PRAC
  • [10] IBERER F, 1987, HERZ, V7, P196