Development and Initial Application of a Fully Integrated Photoacoustic Micro-Ultrasound System

被引:168
作者
Needles, Andrew [1 ]
Heinmiller, Andrew [2 ]
Sun, John [1 ]
Theodoropoulos, Catherine [1 ]
Bates, David [1 ]
Hirson, Desmond [1 ]
Yin, Melissa [3 ,4 ]
Foster, F. Stuart [3 ,4 ]
机构
[1] VisualSonics Inc, Toronto, ON, Canada
[2] VisualSonics Inc, Vevo LAZR Photoacoust Syst, Applicat Res & Dev, Toronto, ON, Canada
[3] Univ Toronto, Sunnybrook Hlth Sci Ctr, Toronto, ON, Canada
[4] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
基金
加拿大健康研究院;
关键词
IN-VIVO; TOMOGRAPHY; RESOLUTION; SATURATION; BRAIN;
D O I
10.1109/TUFFC.2013.2646
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Photoacoustic (PA) imaging for biomedical applications has been under development for many years. Based on the many advances over the past decade, a new photoacoustic imaging system has been integrated into a micro-ultrasound platform for co-registered PA-ultrasound (US) imaging. The design and implementation of the new scanner is described and its performance quantified. Beamforming techniques and signal processing are described, in conjunction with in vivo PA images of normal subcutaneous mouse tissue and selected tumor models. In particular, the use of the system to estimate the spatial distribution of oxygen saturation (sO(2)) in blood and co-registered with B-mode images of the surrounding anatomy are investigated. The system was validated in vivo against a complementary technique for measuring partial pressure of oxygen in blood (pO(2)). The pO(2) estimates were converted to sO(2) values based on a standard dissociation curve found in the literature. Preliminary studies of oxygenation effects were performed in a mouse model of breast cancer (MDA-MB-231) in which control mice were compared with mice treated with a targeted antiangiogenic agent over a 3 d period. Treated mice exhibited a >90% decrease in blood volume, an 85% reduction in blood wash-in rate, and a 60% decrease in relative tissue oxygenation.
引用
收藏
页码:888 / 897
页数:10
相关论文
共 18 条
[1]  
[Anonymous], [No title captured], Patent No. [5, 840, 023, 5840023]
[2]   A NEW 15-50 MHz ARRAY-BASED MICRO-ULTRASOUND SCANNER FOR PRECLINICAL IMAGING [J].
Foster, F. Stuart ;
Mehi, James ;
Lukacs, Marc ;
Hirson, Desmond ;
White, Chris ;
Chaggares, Chris ;
Needles, Andrew .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (10) :1700-1708
[3]   Noninvasive label-free imaging of microhemodynamics by optical-resolution photoacoustic microscopy [J].
Hu, Song ;
Maslov, Konstantin ;
Wang, Lihong V. .
OPTICS EXPRESS, 2009, 17 (09) :7688-7693
[5]   PHOTOACOUSTIC ULTRASOUND (PAUS) - RECONSTRUCTION TOMOGRAPHY [J].
KRUGER, RA ;
LIU, PY ;
FANG, YR ;
APPLEDORN, CR .
MEDICAL PHYSICS, 1995, 22 (10) :1605-1609
[6]   PHOTOACOUSTIC ULTRASOUND [J].
KRUGER, RA .
MEDICAL PHYSICS, 1994, 21 (01) :127-131
[7]   Photoacoustic angiography of the breast [J].
Kruger, Robert A. ;
Lam, Richard B. ;
Reinecke, Daniel R. ;
Del Rio, Stephen P. ;
Doyle, Ryan P. .
MEDICAL PHYSICS, 2010, 37 (11) :6096-6100
[8]   In vitro measurements of absolute blood oxygen saturation using pulsed near-infrared photoacoustic spectroscopy:: accuracy and resolution [J].
Laufer, J ;
Elwell, C ;
Delpy, D ;
Beard, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (18) :4409-4428
[9]   Photoacoustic tomography and sensing in biomedicine [J].
Li, Changhui ;
Wang, Lihong V. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (19) :R59-R97
[10]   Rapid vascular regrowth in tumors after reversal of VEGF inhibition [J].
Mancuso, Michael R. ;
Davis, Rachel ;
Norberg, Scott M. ;
O'Brien, Shaun ;
Sennino, Barbara ;
Nakahara, Tsutomu ;
Yao, Virginia J. ;
Inai, Tetsuichiro ;
Brooks, Peter ;
Freimark, Bruce ;
Shalinsky, David R. ;
Hu-Lowe, Dana D. ;
McDonald, Donald M. .
JOURNAL OF CLINICAL INVESTIGATION, 2006, 116 (10) :2610-2621