Design of a Cyclic Pressure Bioreactor for the Ex Vivo Study of Aortic Heart Valves

被引:1
|
作者
Schipke, Kimberly J. [1 ]
To, S. D. Filip [1 ]
Warnock, James N. [1 ]
机构
[1] Mississippi State Univ, Dept Agr & Biol Engn, Mississippi State, MS 39762 USA
来源
关键词
Bioengineering; Issue; 54; Mechanobiology; Bioreactor; Aortic Heart Valve; Organ Culture; POSTOPERATIVE ARRHYTHMIAS; DISEASE; SURGERY;
D O I
10.3791/3316
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The aortic valve, located between the left ventricle and the aorta, allows for unidirectional blood flow, preventing backflow into the ventricle. Aortic valve leaflets are composed of interstitial cells suspended within an extracellular matrix (ECM) and are lined with an endothelial cell monolayer. The valve withstands a harsh, dynamic environment and is constantly exposed to shear, flexion, tension, and compression. Research has shown calcific lesions in diseased valves occur in areas of high mechanical stress as a result of endothelial disruption or interstitial matrix damage(1-3). Hence, it is not surprising that epidemiological studies have shown high blood pressure to be a leading risk factor in the onset of aortic valve disease(4). The only treatment option currently available for valve disease is surgical replacement of the diseased valve with a bioprosthetic or mechanical valve(5). Improved understanding of valve biology in response to physical stresses would help elucidate the mechanisms of valve pathogenesis. In turn, this could help in the development of non-invasive therapies such as pharmaceutical intervention or prevention. Several bioreactors have been previously developed to study the mechanobiology of native or engineered heart valves(6-9). Pulsatile bioreactors have also been developed to study a range of tissues including cartilage(10), bone(11) and bladder(12). The aim of this work was to develop a cyclic pressure system that could be used to elucidate the biological response of aortic valve leaflets to increased pressure loads. The system consisted of an acrylic chamber in which to place samples and produce cyclic pressure, viton diaphragm solenoid valves to control the timing of the pressure cycle, and a computer to control electrical devices. The pressure was monitored using a pressure transducer, and the signal was conditioned using a load cell conditioner. A LabVIEW program regulated the pressure using an analog device to pump compressed air into the system at the appropriate rate. The system mimicked the dynamic transvalvular pressure levels associated with the aortic valve; a saw tooth wave produced a gradual increase in pressure, typical of the transvalvular pressure gradient that is present across the valve during diastole, followed by a sharp pressure drop depicting valve opening in systole. The LabVIEW program allowed users to control the magnitude and frequency of cyclic pressure. The system was able to subject tissue samples to physiological and pathological pressure conditions. This device can be used to increase our understanding of how heart valves respond to changes in the local mechanical environment.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Design of a new pulsatile bioreactor for tissue engineered aortic heart valve formation
    Dumont, K
    Yperman, J
    Verbeken, E
    Segers, P
    Meuris, B
    Vandenberghe, S
    Flameng, W
    Verdonck, PR
    ARTIFICIAL ORGANS, 2002, 26 (08) : 710 - 714
  • [32] Rethinking mechanical heart valves in the aortic position: new paradigms in design and testing
    Chakraborty, Sreyashi
    Simon, Melinda G.
    Bellofiore, Alessandro
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2025, 11
  • [33] Design and fabrication of a mock circulatory system for reliability tests on aortic heart valves
    Singh, Dhruv
    Singhal, Abhishek
    PROCEEDING OF THE ASME SUMMER BIOENGINEERING CONFERENCE - 2007, 2007, : 753 - 754
  • [34] Pressure dependency of aortic pulse wave velocity in vivo is not affected by vasoactive substances that alter aortic wall tension ex vivo
    Butlin, Mark
    Lindesay, George
    Viegas, Kayla D.
    Avolio, Alberto P.
    AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2015, 308 (10): : H1221 - H1228
  • [35] The mechanical environment regulates the extent of aortic valve calcification and thickness: an ex vivo whole mouse heart study
    El Ouraoui, M.
    Wu, H. W.
    Jukema, J. W.
    Marsan, N. Ajmone
    Kruithof, B. P. T.
    EUROPEAN HEART JOURNAL, 2024, 45
  • [36] Cyclic pressure and shear stress regulate matrix metalloproteinases and cathepsin activity in porcine aortic valves
    Platt, Manu O.
    Xing, Yun
    jo, Hanong Jo
    Yoganathan, Ajit P.
    JOURNAL OF HEART VALVE DISEASE, 2006, 15 (05): : 622 - 629
  • [37] Comparative study of mechanical and transcatheter heart valves for aortic valve replacement
    Ng, F. C.
    Abas, A.
    Razak, N.
    Abu Bakar, H.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2021, 345 : 16 - 16
  • [38] Experimental in vivo and ex vivo models for the study of human aortic dissection: promises and challenges
    Jiang, Ding-Sheng
    Yi, Xin
    Zhu, Xue-Hai
    Wei, Xiang
    AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2016, 8 (12): : 5125 - 5140
  • [39] Cardiovascular magnetic resonance assessment of the aortic valve stenosis: an in vivo and ex vivo study
    Buchner, Stefan
    Debl, Kurt
    Schmid, Franz-Xaver
    Luchner, Andreas
    Djavidani, Behrus
    BMC MEDICAL IMAGING, 2015, 15
  • [40] Cardiovascular magnetic resonance assessment of the aortic valve stenosis: an in vivo and ex vivo study
    Stefan Buchner
    Kurt Debl
    Franz-Xaver Schmid
    Andreas Luchner
    Behrus Djavidani
    BMC Medical Imaging, 15