Fundamental information content accessible with medical x-ray scatter imaging

被引:9
作者
Leclair, RJ [1 ]
Johns, PC [1 ]
机构
[1] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
来源
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 1999年 / 3659卷
关键词
x-ray scatter imaging; coherent scatter; Compton scatter; contrast; signal-to-noise ratio (SNR); spectral blur; experimental apparatus;
D O I
10.1117/12.349547
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Although scattered radiation is generally regarded as a nuisance in radiological imaging, many innovative imaging concepts that use the scattered field have been demonstrated. A systematic approach, however, for analyzing the medical applications of x-ray scatter imaging has been lacking. We have therefore formulated a simple semi-analytic model that consists of imaging a target object against a background material of the same dimensions when both are situated within a water phantom. The target and background objects have small cross-sectional areas (1.0 mm(2)) to allow the omission of self-attenuation and multiple scatter within the objects. The incident energy fluence is kept constant so that similar doses are delivered by the various photon beams. For imaging white brain matter versus gray brain matter in a 15 cm thick water phantom, the maximum signal-to-noise ratio, over all photon energies, for images obtained with the forward scatter between 2 degrees - 12 degrees exceeds that of primary images for all object thicknesses less than or equal to 40 mm. The penalty in dose as a result of spectral blur is generally moderate. For example, using an 80 kV beam for the previous imaging task would require approximately a 24% dose increase relative to using a monoenergetic beam. A high-precision experimental apparatus has been assembled to validate our predictions.
引用
收藏
页码:672 / 681
页数:10
相关论文
共 20 条
[1]   STATUS AND OUTLOOK OF COHERENT-X-RAY SCATTER IMAGING [J].
HARDING, G ;
KOSANETZKY, J .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :933-944
[2]   X-RAY-DIFFRACTION COMPUTED-TOMOGRAPHY [J].
HARDING, G ;
KOSANETZKY, J ;
NEITZEL, U .
MEDICAL PHYSICS, 1987, 14 (04) :515-525
[3]   ELASTIC SCATTER COMPUTED-TOMOGRAPHY [J].
HARDING, G ;
KOSANETZKY, J ;
NEITZEL, U .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (02) :183-186
[4]   A FLYING SPOT X-RAY SYSTEM FOR COMPTON BACKSCATTER IMAGING [J].
HERR, MD ;
MCINERNEY, JJ ;
LAMSER, DG ;
COPENHAVER, GL .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (03) :461-469
[5]  
Jacobs A. M., 1979, Proceedings of the Society of Photo-Optical Instrumentation Engineers, V206, P129
[6]  
JOHNS HE, 1983, PHYSICS RADIOLOGY
[7]   COHERENT SCATTER IN DIAGNOSTIC-RADIOLOGY [J].
JOHNS, PC ;
YAFFE, MJ .
MEDICAL PHYSICS, 1983, 10 (01) :40-50
[8]   X-RAY-DIFFRACTION MEASUREMENTS OF SOME PLASTIC MATERIALS AND BODY-TISSUES [J].
KOSANETZKY, J ;
KNOERR, B ;
HARDING, G ;
NEITZEL, U .
MEDICAL PHYSICS, 1987, 14 (04) :526-532
[9]   A semianalytic model to investigate the potential applications of x-ray scatter imaging [J].
Leclair, RJ ;
Johns, PC .
MEDICAL PHYSICS, 1998, 25 (06) :1008-1020
[10]  
LECLAIR RJ, UNPUB MED PHYS