Evaluation of a Novel Commercial Assay for the Determination of Cyclosporine A, Tacrolimus, Sirolimus, and Everolimus by Liquid Chromatography-Tandem Mass Spectrometric Assay

被引:23
作者
Becker, Susen [1 ]
Thiery, Joachim [1 ]
Ceglarek, Uta [1 ]
机构
[1] Univ Hosp Leipzig, Inst Lab Med Clin Chem & Mol Diagnost, D-04103 Leipzig, Germany
关键词
tandem mass spectrometry; cyclosporine A; tacrolimus; sirolimus; everolimus; HPLC-MS/MS METHOD; SIMULTANEOUS QUANTIFICATION; HUMAN BLOOD; IMMUNOSUPPRESSANTS; ROBUST;
D O I
10.1097/FTD.0b013e318274827d
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective: The aim of this study was the evaluation of the first commercially available in-vitro diagnostic (IVD)-mass spectrometric immunosuppressant assay from Chromsystems (MassTox Immunosuppressants ONEMINUTE Test) and the comparison to a routinely used online solid phase extraction liquid chromatography-tandem mass spectrometric assay method for the measurement of cyclosporine A, everolimus, sirolimus, and tacrolimus in patient whole blood samples. Methods: An API 4000 [tandem mass spectrometer (AB SCIEX)] combined with a CTC Pal autosampler (CTC Analytics AG) and a Shimadzu ultra-fast liquid chromatography (UFLC) system were applied for the direct liquid chromatography-tandem mass spectrometric assay analysis using electrospray ionization in positive ion mode. Isotope-labeled internal standards were used for the commercial assay. Within-and between-day variation, accuracy, and limits of quantification were determined. Samples from external international proficiency testing schemes were measured to assess the accuracy of the commercial assay. About 100 patient samples were reanalyzed with the commercial test kit, and the results were compared with our in-house results. Results: The limit of quantification for the commercial assay was 0.5 ng/mL for everolimus, sirolimus, and tacrolimus and 5 ng/mL for cyclosporine A. The coefficient of variation for all immunosuppressants was lower than 7% (within day) and 12% (between days) for all 5 concentration levels. Accuracy ranged between 82% and 111% for quality control samples and between 89% and 112% for samples from the external quality assurance program. Both methods showed a very good agreement (r > 0.91) in patient samples over the whole concentration range for all immunosuppressants. Conclusion: The commercial immunosuppressant assay from Chromsystems represents a standardized IVD-certified alternative to our in-house developed assay.
引用
收藏
页码:129 / 132
页数:4
相关论文
共 50 条
  • [21] Evaluation of QMS Everolimus Assay Using Hitachi 917 Analyzer: Comparison With Liquid Chromatography/Mass Spectrometry
    Dasgupta, Amitava
    Davis, Bonnet
    Chow, Loretta
    THERAPEUTIC DRUG MONITORING, 2011, 33 (02) : 149 - 154
  • [22] Analytical Performance Evaluation of a New Cobas Tacrolimus Assay on Cobas e411 Analyzer: Comparison of Values Obtained by the CMIA Tacrolimus Assay and a Liquid Chromatography Combined with Tandem Mass Spectrometric Method
    Dasgupta, Amitava
    Khalil, Samir A.
    Johnson-Davis, Kamisha L.
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2016, 46 (02) : 204 - 208
  • [23] Is the automated Elecsys tacrolimus assay on the Roche cobas e 602 analyzer an acceptable replacement for a liquid chromatography-tandem mass spectrometry-based assay?
    Kanegusuku, Anastasia Gant
    Yeo, Kiang-Teck J.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2024, 161 (01) : 97 - 106
  • [24] Liquid chromatography-tandem mass spectrometric assay for the quantitation in human plasma of the novel indenoisoquinoline topoisomerase I inhibitors, NSC 743400 and NSC 725776
    Holleran, Julianne L.
    Parise, Robert A.
    Yellow-Duke, Archibong E.
    Egorin, Merrill J.
    Eiseman, Julie L.
    Covey, Joseph M.
    Beumer, Jan H.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2010, 52 (05) : 714 - 720
  • [25] Determination of sirolimus in rabbit arteries using liquid chromatography separation and tandem mass spectrometric detection
    Zhang, Jun
    Rodila, Ramona
    Watson, Pamela
    Ji, Qin
    El-Shourbagy, Tawakol A.
    BIOMEDICAL CHROMATOGRAPHY, 2007, 21 (10) : 1036 - 1044
  • [26] Validation of an assay for voriconazole in serum samples using liquid chromatography-tandem mass spectrometry
    Keevil, BG
    Newman, S
    Lockhart, S
    Howard, SJ
    Moore, CB
    Denning, DW
    THERAPEUTIC DRUG MONITORING, 2004, 26 (06) : 650 - 657
  • [27] Enantioselective assay of nimodipine in human plasma using liquid chromatography-tandem mass spectrometry
    Isse, Fadumo Ahmed
    Le, Tyson
    Mahmoud, Sherif Hanafy
    BIOMEDICAL CHROMATOGRAPHY, 2021, 35 (02)
  • [28] Liquid chromatography-tandem mass spectrometric assay for the quantitation of the novel radiation protective agent and radiation mitigator JP4-039 in murine plasma
    Christner, Susan
    Guo, Jianxia
    Parise, Robert A.
    Ringeval, Melanie
    Hoye, Adam T.
    Wipf, Peter
    Epperly, Michael W.
    Greenberger, Joel S.
    Beumer, Jan H.
    Eiseman, Julie L.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 150 : 169 - 175
  • [29] Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood
    Lee, Yong-Wha
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2013, 6 (06) : 1535 - 1539
  • [30] COMPARISON OF ANTIBODY-CONJUGATED MAGNETIC IMMUNOASSAY AND LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY FOR THE MEASUREMENT OF CYCLOSPORINE AND TACROLIMUS IN WHOLE BLOOD
    Cangemi, G.
    Barco, S.
    Bonifazio, P.
    Maffia, A.
    Agazzi, A.
    Melioli, G.
    INTERNATIONAL JOURNAL OF IMMUNOPATHOLOGY AND PHARMACOLOGY, 2013, 26 (02) : 419 - 426