Effectiveness and safety of a model predictive control (MPC) algorithm for an artificial pancreas system in outpatients with type 1 diabetes ( T1D): systematic review and meta-analysis

被引:11
作者
Kang, Su Lim [1 ]
Hwang, Yoo Na [1 ]
Kwon, Ji Yean [1 ]
Kim, Sung Min [1 ,2 ]
机构
[1] Dongguk Univ Seoul, Dept Med Device & Healthcare, 26,Pil Dong 3 Ga, Seoul 04620, South Korea
[2] Dongguk Univ Seoul, Dept Med Device Regulatory Sci, 26,Pil Dong 3 Ga, Seoul 04620, South Korea
关键词
Artificial pancreas; Algorithm; Model predictive control; Hypoglycemia; Type; 1; diabetes; CLOSED-LOOP CONTROL; OVERNIGHT GLUCOSE CONTROL; INSULIN-PUMP THERAPY; HOME-USE; GLYCEMIC CONTROL; BIONIC PANCREAS; CROSSOVER TRIAL; ADOLESCENTS; CHILDREN; DELIVERY;
D O I
10.1186/s13098-022-00962-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The purpose of this study was to assess the effectiveness and safety of a model predictive control (MPC) algorithm for an artificial pancreas system in outpatients with type 1 diabetes. Methods: We searched PubMed, EMBASE, Cochrane Central, and the Web of Science to December 2021. The eligibility criteria for study selection were randomized controlled trials comparing artificial pancreas systems (MPC, PID, and fuzzy algorithms) with conventional insulin therapy in type 1 diabetes patients. The heterogeneity of the overall results was identified by subgroup analysis of two factors including the intervention duration (overnight and 24 h) and the follow-up periods (< 1 week, 1 week to 1 month, and > 1 month). Results: The meta-analysis included a total of 41 studies. Considering the effect on the percentage of time maintained in the target range between the MPC-based artificial pancreas and conventional insulin therapy, the results showed a statistically significantly higher percentage of time maintained in the target range in overnight use (10.03%, 95% CI [7.50, 12.56] p < 0.00001). When the follow-up period was considered, in overnight use, the MPC-based algorithm showed a statistically significantly lower percentage of time maintained in the hypoglycemic range (-1.34%, 95% CI [-1.87, -0.81] p < 0.00001) over a long period of use (> 1 month). Conclusions: Overnight use of the MPC-based artificial pancreas system statistically significantly improved glucose control while increasing time maintained in the target range for outpatients with type 1 diabetes. Results of subgroup analysis revealed that MPC algorithm-based artificial pancreas system was safe while reducing the time maintained in the hypoglycemic range after an overnight intervention with a long follow-up period (more than 1 month).
引用
收藏
页数:12
相关论文
共 50 条
[31]   Telemedicine in Improving Glycemic Control Among Children and Adolescents With Type 1 Diabetes Mellitus: Systematic Review and Meta-Analysis [J].
Zhang, Kun ;
Huang, Qiyuan ;
Wang, Qiaosong ;
Li, Chengyang ;
Zheng, Qirong ;
Li, Zhuoyue ;
Xu, Dan ;
Xie, Cuiling ;
Zhang, Mingqi ;
Lin, Rongjin .
JOURNAL OF MEDICAL INTERNET RESEARCH, 2024, 26
[32]   Periodontal risk markers in children and adolescents with type 1 diabetes: A systematic review and meta-analysis [J].
Jensen, Emilija ;
Allen, Gabrielle ;
Bednarz, Jana ;
Couper, Jennifer ;
Pena, Alexia .
DIABETES-METABOLISM RESEARCH AND REVIEWS, 2021, 37 (01)
[33]   Intranasal versus injectable glucagon for hypoglycemia in type 1 diabetes: systematic review and meta-analysis [J].
Pontiroli, Antonio E. ;
Tagliabue, Elena .
ACTA DIABETOLOGICA, 2020, 57 (06) :743-749
[34]   Subclinical arterial damage in children and adolescents with type 1 diabetes: A systematic review and meta-analysis [J].
Giannopoulou, Eleni Z. ;
Doundoulakis, Ioannis ;
Antza, Christina ;
Christoforidis, Athanasios ;
Haidich, Anna-Bettina ;
Kotsis, Vasilios ;
Stabouli, Stella .
PEDIATRIC DIABETES, 2019, 20 (06) :668-677
[35]   Efficacy and safety of carbohydrate restriction in patients with type 1 diabetes: A systematic review and meta-analysis [J].
Stamati, Athina ;
Sotiriou, Georgia ;
Karagiannis, Thomas ;
Christoforidis, Athanasios .
DIABETES OBESITY & METABOLISM, 2023, 25 (09) :2770-2773
[36]   The effect of automated insulin delivery system use on diabetes distress in people with type 1 diabetes and their caregivers: A systematic review and meta-analysis [J].
Canha, Dulce ;
McMahon, Virginia ;
Schmitz, Susanne ;
De Beaufort, Carine ;
Alzaid, Fawaz ;
Reznik, Yves ;
Riveline, Jean-Pierre ;
Fagherazzi, Guy ;
Aguayo, Gloria A. .
DIABETIC MEDICINE, 2025, 42 (04)
[37]   Efficacy of carbohydrate counting in people with type 1 and type 2 diabetes mellitus: a systematic review and meta-analysis [J].
Sakane, Naoki ;
Domichi, Masayuki ;
Suganuma, Akiko .
DIABETOLOGY INTERNATIONAL, 2025, 16 (03) :546-558
[38]   Sympathetic neural abnormalities in type 1 and type 2 diabetes: a systematic review and meta-analysis [J].
Grassi, Guido ;
Biffi, Annalisa ;
Dell'Oro, Raffaella ;
Quarti Trevano, Fosca ;
Seravalle, Gino ;
Corrao, Giovanni ;
Perseghin, Gianluca ;
Mancia, Giuseppe .
JOURNAL OF HYPERTENSION, 2020, 38 (08) :1436-1442
[39]   Thyroid autoimmunity in Type 1 diabetes: systematic review and meta-analysis [J].
Shun, C. B. ;
Donaghue, K. C. ;
Phelan, H. ;
Twigg, S. M. ;
Craig, M. E. .
DIABETIC MEDICINE, 2014, 31 (02) :126-135
[40]   Decreased Flow-Mediated Dilatation in Children With Type 1 Diabetes: A Systematic Review and Meta-Analysis [J].
Cao, Lei ;
Hou, Miao ;
Zhou, Wanping ;
Sun, Ling ;
Shen, Jie ;
Chen, Ye ;
Tang, Yunjia ;
Wang, Bo ;
Li, Xuan ;
Lv, Haitao .
ANGIOLOGY, 2021, 72 (10) :908-915