Preconcentration and voltammetric determination of mercury(II) at a chemically modified glassy carbon electrode

被引:62
作者
Sousa, MDB
Bertazzoli, R
机构
关键词
D O I
10.1021/ac950969j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work, the organic compound 2-mercaptobenzimidazole was covalently bound on the surface of a glassy carbon rod, via silanization, yielding a material capable of selectively complexing Hg2- ions. This material was applied as an electrode for voltammetric determination of mercury(II) following its nonelectrolytic preconcentration. After exchanging the medium, the voltammetric measurements were carried out by anodic stripping in the differential pulse mode (pulse amplitude, 50 mV; scan rate, 1.25 mV s(-1)) using 10(-2) mol L(-1) NaSCN solution as supporting electrolyte. An anodic stripping peak was obtained at 0.06 V (vs SCE) by scanning the potential from -0.3 to +0.3 V, After a 5 min preconcentration period in a pH 4.0 Hg2+ solution, this electrode shows increasing voltammetric response in the range 0.1-2.2 mu g mL(-1), with a relative standard deviation of 5% and a practical detection limit of 0.1 mu g mL(-1) (5.0 x 10(-7) mol dm(-3)). Compared with the conventional stripping approach, this chemically modified glassy carbon electrode procedure presented good discrimination against interference from Cu(II) in up to 10-fold molar excess.
引用
收藏
页码:1258 / 1261
页数:4
相关论文
共 15 条
[1]   DEVELOPMENT OF A CHEMICALLY MODIFIED ELECTRODE BASED ON CARBON PASTE AND FUNCTIONALIZED SILICA-GEL FOR PRECONCENTRATION AND VOLTAMMETRIC DETERMINATION OF MERCURY(II) [J].
ALEIXO, LM ;
SOUZA, MDB ;
GODINHO, OES ;
NETO, GD ;
GUSHIKEM, Y ;
MOREIRA, JC .
ANALYTICA CHIMICA ACTA, 1993, 271 (01) :143-148
[2]   DETERMINATION OF MERCURY AT ELECTRODES MODIFIED WITH POLYMERIC FILMS OF [RU(V-BPY)3](2+) INCORPORATING AMINO-ACIDS [J].
CHA, SK ;
AHN, BK ;
HWANG, JU ;
ABRUNA, HD .
ANALYTICAL CHEMISTRY, 1993, 65 (11) :1564-1569
[3]   ELECTROCHEMICAL OXIDATION OF AMINE-CONTAINING COMPOUNDS - A ROUTE TO THE SURFACE MODIFICATION OF GLASSY-CARBON ELECTRODES [J].
DEINHAMMER, RS ;
HO, M ;
ANDEREGG, JW ;
PORTER, MD .
LANGMUIR, 1994, 10 (04) :1306-1313
[4]   MEDIUM EFFECTS IN THE STRIPPING VOLTAMMETRY OF MERCURY [J].
GUNASINGHAM, H ;
ANG, KP ;
NGO, CC .
ANALYST, 1988, 113 (10) :1533-1536
[5]   DETERMINATION OF SILVER(I) BY CYCLIC VOLTAMMETRY AT IODINE-COATED ELECTRODES [J].
HOURANI, MK .
ANALYST, 1994, 119 (09) :1975-1978
[6]  
JINRUI X, 1994, ANALYST, V119, P1599
[7]  
KOVALENKO AA, 1977, ZH ANAL KHIM, V32, P273
[8]   DETERMINATION OF MERCURY IONS ON A DIPHENYLCARBAZONE BULK MODIFIED GRAPHITE ELECTRODE [J].
LABUDA, J ;
PLASKON, V .
ANALYTICA CHIMICA ACTA, 1990, 228 (02) :259-263
[9]   2-MERCAPTO-BENZIMIDAZOLE AS AN ANALYTICAL REAGENT - ESTIMATION OF PALLADIUM [J].
MAJUMDAR, AK ;
CHAKRABARTTY, MM .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1958, 162 (02) :101-105
[10]   DIFFERENTIAL-PULSE ANODIC-STRIPPING VOLTAMMETRY OF COPPER WITH A CHEMICALLY-MODIFIED GLASSY-CARBON ELECTRODE [J].
MIWA, T ;
JIN, LT ;
MIZUIKE, A .