Transapical aortic valve implantation using a new self-expandable bioprosthesis (ACURATE TA™): 6-month outcomes†

被引:55
作者
Kempfert, Joerg [1 ]
Treede, Hendrik [2 ]
Rastan, Ardawan J. [3 ]
Schoenburg, Markus [1 ]
Thielmann, Matthias [4 ]
Sorg, Stefan [5 ]
Mohr, Friedrich-W. [3 ]
Walther, Thomas [1 ]
机构
[1] Kerckhoff Clin Bad Nauheim, Dept Cardiac Surg, D-61231 Bad Nauheim, Germany
[2] Univ Hamburg, Dept Cardiac Surg, Heartctr, Hamburg, Germany
[3] Univ Leipzig, Dept Cardiac Surg, Heartctr, Leipzig, Germany
[4] Univ Duisburg Essen, West German Heart Ctr Essen, Dept Cardiac Surg, Essen, Germany
[5] Univ Freiburg, Dept Cardiac Surg, Clin Cardiac Surg, D-79106 Freiburg, Germany
关键词
Aortic valve implantation; Minimally invasive; Transapical; HIGH-RISK PATIENTS; OUTCOME SOURCE REGISTRY; EDWARDS SAPIEN VALVE; EUROPEAN REGISTRY; TRANSCATHETER; REPLACEMENT; EXPERIENCE; STENOSIS; PROSTHESIS; PREDICTORS;
D O I
10.1093/ejcts/ezs139
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The ACURATE TA((TM)) Aortic Bioprosthesis and Delivery System (Symetis S.A., Ecublens, Switzerland) is a new transcatheter aortic valve designed for transapical implantation. The six-month results from the completed first-in-man study are reported. The Symetis ACURATE TA((TM)) is composed of a porcine biological tissue valve attached to a self-expandable nitinol stent. It allows for anatomical orientation and facilitates intuitive implantation providing tactile feedback. Since November 2009, a total of 40 high-risk elderly patients have been treated. The mean age of enrolled patients was 83.2 +/- 4.0; 60.0% were female, with a mean logistic EuroSCORE of 21.2 +/- 10.8% and a mean Society of Thoracic Surgeons (STS) score of 9.0 +/- 4.7%. All implants were delivered successfully in the intra-annular and subcoronary position. One patient was converted to conventional surgery due to coronary impingement (after valve-in-valve implantation). One additional patient received valve-in-valve treatment (SAPIEN THV TA((TM))). Five patients expired within 30 days and two additional patients expired during the 6-month follow-up due to non-valve-related causes resulting in a mid-term survival rate of 82.5%. Two patients suffered a stroke and another three required new onset pacemaker implantation. The mean aortic gradient significantly improved and remained stable throughout the follow-up (baseline: 51.9 +/- 14.3 mmHg, 30 days: 12.3 +/- 5.1 mmHg, 6 months: 11.9 +/- 5.8 mmHg). At the 6-month follow-up, 96.7% of patients demonstrated either none/trace or mild (1+/4) paravalvular leakage only. According to the Valve Academic Research Council the device's success rate was 92.5%, with a 30-day safety profile of 25%. At the 6-month follow-up, the ACURATE TA((TM)) device showed stable valve function with low rates of paravalvular leakages. The cohort of high-risk patients demonstrated good clinical outcomes and 6-month survival.
引用
收藏
页码:52 / 57
页数:6
相关论文
共 50 条
  • [31] Outcomes after transcatheter valve-in-valve implantation using a balloon-expandable Edwards Sapien valve in patients with degenerated Freestyle aortic bioprosthesis
    Burgdorf, Christof
    Momchilovska, Andrijana Vukadinovikj
    Remppis, Bjoern Andrew
    ANNALS OF CARDIOTHORACIC SURGERY, 2021, 10 (05) : 667 - 673
  • [32] Limited Expansion of the New Self-Expandable Transcathether Aortic Valve Prosthesis (CoreValve Evolut R)
    Serio, Daniela
    Doss, Mirko
    Kim, Won-Keun
    Moellmann, Helge
    Walther, Thomas
    ANNALS OF THORACIC SURGERY, 2016, 101 (05) : E137 - E138
  • [33] Meta-Analysis of Comparison Between Self-Expandable and Balloon-Expandable Valves for Patients Having Transcatheter Aortic Valve Implantation
    Moretti, Claudio
    D'Ascenzo, Fabrizio
    Mennuni, Marco
    Taha, Salma
    Brambilla, Nedy
    Nijhoff, Freek
    Fraccaro, Chiara
    Barbanti, Marco
    Tamburino, Corrado
    Tarantini, Giuseppe
    Rossi, Marco L.
    Presbitero, Patrizia
    Napodanno, Massimo
    Stella, Pieter
    Bedogni, Francesco
    Omede, Pierluigi
    Conrotto, Federico
    Montefusco, Antonio
    Giordana, Francesca
    Zoccai, Giuseppe Biondi
    Agostoni, Piefrancesco
    D'Amico, Maurizio
    Rinaldi, Mauro
    Marra, Sebastiano
    Gaita, Fiorenzo
    AMERICAN JOURNAL OF CARDIOLOGY, 2015, 115 (12) : 1720 - 1725
  • [34] A validated methodology for patient specific computational modeling of self-expandable transcatheter aortic valve implantation
    Bosmans, Bart
    Famaey, Nele
    Verhoelst, Eva
    Bosmans, Johan
    Vander Sloten, Jos
    JOURNAL OF BIOMECHANICS, 2016, 49 (13) : 2824 - 2830
  • [35] Successful reposition and treatment of severe aortic paraprosthetic regurgitation after percutaneous implantation of a self-expandable biological valve
    Frerker, Christian
    Bergmann, Martin
    Schewel, Dimitry
    Busse, Cord
    Jensen, Folke
    Schmoeckel, Michael
    Schneider, Carsten
    Kuck, Karl-Heinz
    Schaefer, Ulrich
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2012, 13 (05) : 346 - 350
  • [36] Impact of Repositioning on Outcomes Following Transcatheter Aortic Valve Replacement With a Self-Expandable Valve
    Attizzani, Guilherme F.
    Dallan, Luis Augusto P.
    Markowitz, Alan
    Yakubov, Steven J.
    Deeb, G. Michael
    Reardon, Michael J.
    Forrest, John K.
    Mangi, Abeel A.
    Huang, Jian
    Popma, Jeffrey J.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (15) : 1816 - 1824
  • [37] Transcatheter Self-Expandable Valve Implantation for Aortic Stenosis in Small Aortic Annuli The TAVI-SMALL Registry
    Regazzoli, Damiano
    Chiarito, Mauro
    Cannata, Francesco
    Pagnesi, Matteo
    Miura, Mizuki
    Ziviello, Francesca
    Picci, Andrea
    Reifart, Joerg
    De Marco, Federico
    Bedogni, Francesco
    Adamo, Marianna
    Curello, Salvatore
    Teles, Rui
    Taramasso, Maurizio
    Barbanti, Marco
    Tamburino, Corrado
    Stefanini, Giulio G.
    Mangieri, Antonio
    Giannini, Francesco
    Pagnotta, Paolo A.
    Maisano, Francesco
    Kim, Won-Keun
    Van Mieghem, Nicolas M.
    Colombo, Antonio
    Reimers, Bernhard
    Latib, Azeem
    JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (02) : 196 - 206
  • [38] Prosthesis Infolding Incidence and Short-Term Outcomes in Transcatheter Aortic Valve Implantation Using Evolut Self-Expandable Device: A Multicenter Study
    Bezzeccheri, Andrea
    Poletti, Enrico
    Vermeersch, Paul
    Veulemans, Verena
    Piuhola, Jarkko
    Kerkela, Seija
    Lehtola, Heidi
    De Backer, Ole
    Quagliana, Angelo
    Mammone, Concetta
    Ribichini, Flavio
    Prihadi, Edgard
    Scott, Benjamin
    Zivelonghi, Carlo
    Verheye, Stefan
    Agostoni, Pierfrancesco
    AMERICAN JOURNAL OF CARDIOLOGY, 2024, 221 : 102 - 109
  • [39] Transfermoral transcatheter aortic valve implantation using self-expanding Allegra bioprosthesis: One-year single-center outcomes
    Milan, Joanna
    Gozdek, Miroslaw
    Targonski, Radoslaw
    Kowalewski, Mariusz
    Stanska, Aleksandra
    Fijalkowski, Marcin
    Lango, Romuald
    Jaguszewski, Milosz
    Jagielak, Dariusz
    CARDIOLOGY JOURNAL, 2021, 28 (06) : 825 - 830
  • [40] Transfermoral aortic valve implantation using self-expanding New Valve Technology (NVT) Allegra bioprosthesis: A pilot prospective study
    Jagielak, Dariusz
    Stanska, Aleksandra
    Klapkowski, Andrzej
    Brzezinski, Maciej
    Kowalik, Maciej
    Ciecwierz, Dariusz
    Jaguszewski, Milosz
    Fijalkowski, Marcin
    CARDIOLOGY JOURNAL, 2021, 28 (03) : 384 - 390