Aim: Reliable post-implant evaluation of prostate seed implants requires optimal seed identification and accurate delineation of anatomical structures. In this study the GEC-ESTRO groups BRAPHYQS and PRO-BATE investigated the interobserver variability in post-implant prostate contouring, seed reconstruction and image fusion and its impact on the dose-volume parameters. Materials: Post-implant T2-TSE, T1-GE and CT images were acquired for three patients, in order to evaluate four post-plan techniques: (a) CT, (b) T1 + T2, (c) CT + T2, (d) cT + T1(int) + T2. Three interobserver studies were set up. (1) Contouring: the CTV-prostate was delineated on CT and T2 by eight physicians. Additionally one reference contour was defined on both image modalities for each patient. (2) Seed reconstruction: seven physicists localised the seeds on Ti and CT, manually and with CT seed finder tools. A reference seed geometry was defined on CT and T1. (3) Fusion: six physicists registered the image sets for technique (b)-(d), using seeds (if visible) and anatomical landmarks. A reference fusion was determined for each combined technique. Results: (1) The SDref for contouring (1 SD with respect to the reference volume) was largest for CT (23%), but also surprisingly large for MRI (17%). This resulted in large SDref values for 090 for all techniques (17-23%). The surprisingly large SDref for MRI was partly due to variations in interpretation of what to include in the prostate contour. (2) The SDref in D90 for seed reconstruction was small (2%) for all techniques, except for T1 + T2 (7%). (3) The SDref in D90 due to image fusion was quite large, especially for direct fusion of CT + T2 (16%) where clearly corresponding landmarks were missing (seeds hardly visible on T2). In general, we observed large differences in D90 depending on the technique used. Conclusions: The dosimetric parameters for prostate post-implant evaluation showed large techniquedependent interobserver variabilities. Contouring and image fusion are the 'weak links' in the procedure. Guidelines and training in contouring together with incorporation of automated fusion software need to be implemented. (c) 2012 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 104 (2012) 192-198
引用
收藏
页码:192 / 198
页数:7
相关论文
共 28 条
  • [1] Impact of prostate volume evaluation by different observers on CT-based post-implant dosimetry
    Al-Qaisieh, B
    Ash, D
    Bottomley, DM
    Carey, BM
    [J]. RADIOTHERAPY AND ONCOLOGY, 2002, 62 (03) : 267 - 273
  • [2] ESTRO/EAU/EORTC recommendations on permanent seed implantation for localized prostate cancer
    Ash, D
    Flynn, A
    Battermann, J
    de Reijke, T
    Lavagnini, P
    Blank, L
    [J]. RADIOTHERAPY AND ONCOLOGY, 2000, 57 (03) : 315 - 321
  • [3] The correlation between D90 and outcome for l-125 seed implant monotherapy for localised prostate cancer
    Ash, Dan
    Al-Qaisieh, Bashar
    Bottomley, David
    Carey, Brendan
    Joseph, Joji
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 79 (02) : 185 - 189
  • [4] Computed tomography determination of prostate volume and maximum dimensions: a study of interobserver variability
    Berthelet, E
    Liu, MCC
    Agranovich, A
    Patterson, K
    Currie, T
    [J]. RADIOTHERAPY AND ONCOLOGY, 2002, 63 (01) : 37 - 40
  • [5] Evaluation of radiobiologic biochemical control in a large permanent prostate brachytherapy population from a single institution using AAPM TG-137 parameters
    Butler, Wayne M.
    Stewart, Renee R.
    Merrick, Gregory S.
    [J]. BRACHYTHERAPY, 2011, 10 (01) : 16 - 28
  • [6] Crook Juanita, 2004, Brachytherapy, V3, P55, DOI 10.1016/j.brachy.2004.05.001
  • [7] Accuracy of seed reconstruction in prostate postplanning studied with a CT- and MRI-compatible phantom
    De Brabandere, Marisol
    Kirisits, Christian
    Peeters, Ronald
    Haustermans, Karin
    Van den Heuvel, Frank
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 79 (02) : 190 - 197
  • [8] The contribution of magnetic resonance imaging to the three-dimensional treatment planning of localized prostate cancer
    Debois, M
    Oyen, R
    Maes, F
    Verswijvel, G
    Gatti, G
    Bosmans, H
    Feron, M
    Bellon, E
    Kutcher, G
    Van Poppel, H
    Vanuytsel, L
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (04): : 857 - 865
  • [9] Intraobserver and interobserver variability of MR imaging- and CT-derived prostate volumes after transperineal interstitial permanent prostate brachytherapy
    Dubois, DF
    Prestidge, BR
    Hotchkiss, LA
    Prete, JJ
    Bice, WS
    [J]. RADIOLOGY, 1998, 207 (03) : 785 - 789
  • [10] Gastaldi Emilio, 2009, Arch Ital Urol Androl, V81, P215