Cobalt(II) ion-selective electrode with solid contact

被引:9
|
作者
Wardak, Cecylia [1 ]
机构
[1] Marie Curie Sklodowska Univ, Fac Chem, Dept Analyt Chem & Instrumental Anal, PL-20031 Lublin, Poland
来源
CENTRAL EUROPEAN JOURNAL OF CHEMISTRY | 2008年 / 6卷 / 04期
关键词
Cobalt-selective electrode; Potentiometry; Chelating agent; Flow injection analysis;
D O I
10.2478/s11532-008-0060-1
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A new all plastic sensor for Co2+ ions based on 2-amino-5 (hydroxynaphtyloazo-1')-1,3,4 thiadiazole (ATIDAN) as ionophore was prepared. The electrode exhibits a low detection limit of 1.5 x 10(-6) mol L-1 and almost theoretical Nernstian slope in the activity range 4.0 x 10(-6)- 1 x 10(-1) mol L-1 of cobalt ions. The response time of the sensor is less than 10 s and it can be used over a period of 6 months without any measurable divergence in potential. The proposed sensor shows a fairly good selectivity for Co(II) over other metal ions. The electrode was successfully applied for determination of Co2+ in real samples and as an indicator electrode in potentiometric titration of Co2+ ions with EDTA.
引用
收藏
页码:607 / 612
页数:6
相关论文
共 50 条
  • [1] Uranyl ion-selective electrode with solid contact
    Pietrzak, Karolina
    Wardak, Cecylia
    PROCEEDINGS OF THE 16TH INTERNATIONAL STUDENTS CONFERENCE MODERN ANALYTICAL CHEMISTRY, 2020, : 45 - 50
  • [2] Solid-contact bromide ion-selective electrode
    Stefanova, OK
    Alagova, ZS
    Khripun, GA
    RUSSIAN JOURNAL OF APPLIED CHEMISTRY, 1995, 68 (08) : 1212 - 1213
  • [3] Solid-contact ion-selective electrode with stable internal electrode
    Vamvakaki, M
    Chaniotakis, NA
    ANALYTICA CHIMICA ACTA, 1996, 320 (01) : 53 - 61
  • [4] Application of ionic liquid to the construction of Cu(II) ion-selective electrode with solid contact
    Wardak, Cecylia
    Lenik, Joanna
    SENSORS AND ACTUATORS B-CHEMICAL, 2013, 189 : 52 - 59
  • [5] Application of ionic liquid to the construction of Cu(II) ion-selective electrode with solid contact
    Lenik, Joanna
    Wardak, Cecylia
    Grabarczyk, Malgorzata
    26TH EUROPEAN CONFERENCE ON SOLID-STATE TRANSDUCERS, EUROSENSOR 2012, 2012, 47 : 152 - 155
  • [6] ION-SELECTIVE ELECTRODE FOR DETERMINATION OF COBALT
    OBMETKO, AA
    RAKHMANKO, EM
    LOMAKO, VL
    JOURNAL OF ANALYTICAL CHEMISTRY OF THE USSR, 1990, 45 (08): : 1148 - 1152
  • [7] Solid Contact Nitrate Ion-selective Electrode Based on Cobalt(II) Complex with 4,7-diphenyl-1,10-phenanthroline
    Pietrzak, Karolina
    Wardak, Cecylia
    Lyszczek, Renata
    ELECTROANALYSIS, 2020, 32 (04) : 724 - 731
  • [8] DESIGN OF A FLUORIDE-SELECTIVE ELECTRODE WITH INTERNAL CONTACT BASED ON A CU(II) ION-SELECTIVE ELECTRODE
    RADIC, N
    BRALIC, M
    MICROCHEMICAL JOURNAL, 1993, 48 (03) : 377 - 382
  • [9] Electrosynthesized polypyrrole/zeolite composites as solid contact in potassium ion-selective electrode
    Yu, Kai
    He, Ning
    Kumar, Narendra
    Wang, NianXing
    Bobacka, Johan
    Ivaska, Ari
    ELECTROCHIMICA ACTA, 2017, 228 : 66 - 75
  • [10] POTASSIUM ION-SELECTIVE ELECTRODE BASED ON A COBALT(II) HEXACYANOFERRATE FILM-MODIFIED ELECTRODE
    GAO, ZQ
    ZHOU, XY
    WANG, GQ
    LI, PB
    ZHAO, ZF
    ANALYTICA CHIMICA ACTA, 1991, 244 (01) : 39 - 48