Accuracy of patient dose estimation in cone beam computed tomography in breast irradiation by size-specific dose estimates with position correction

被引:1
作者
Ueno, Hiroyuki [1 ,2 ]
Matsubara, Kosuke [1 ]
Bou, Sayuri [3 ,4 ]
Hizume, Masato [2 ]
机构
[1] Kanazawa Univ, Grad Sch Med Sci, Div Hlth Sci, Kanazawa, Japan
[2] Takaoka City Hosp, Dept Radiol, Takaoka, Japan
[3] Takaoka City Hosp, Dept Radiotherapy, Takaoka, Japan
[4] Kanazawa Univ, Grad Sch Med Sci, Dept Radiol, Kanazawa, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2022年 / 23卷 / 12期
关键词
cone-beam computed tomography; image-guided radiation therapy; size-specific dose estimates; GUIDED RADIATION-THERAPY; CT; RADIOTHERAPY;
D O I
10.1002/acm2.13851
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study aims to investigate the effects of the position correction of size-specific dose estimates (SSDE) on patient dose estimation in cone beam computed tomography (CBCT). The relationship between the phantom position and absorbed dose in the right breast was studied using optically stimulated luminescence dosimeters and a simulated human body phantom. The effect of position correction for CT dose index (CTDI) on SSDE was investigated in 51 patients who underwent right breast irradiation by comparing the SSDE with position correction and SSDE without position correction. The absorbed dose in the right breast tended to decrease by 10.2% as the phantom was placed away from the center of CBCT. The mean and standard deviation of SSDE were 2.54 +/- 0.29 and 2.92 +/- 0.30 mGy with and without position correction, respectively. The SSDE with position correction was 13.1% lower than that without position correction (p < 0.05). SSDE was different when the patient's torso center was located at the isocenter of CBCT, and when it was not. The same tendency was seen in the case of the breast. Therefore, if the center of the patient is not at the acquisition center of the CT scanner, position correction is required when estimating SSDE.
引用
收藏
页数:7
相关论文
共 18 条
  • [1] An investigation of image guidance dose for breast radiotherapy
    Alvarado, Rosemerie
    Booth, Jeremy T.
    Bromley, Regina M.
    Gustafsson, Helen B.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (03): : 25 - 38
  • [2] Imaging doses from the Elekta Synergy X-ray cone beam CT system
    Amer, A.
    Marchant, T.
    Sykes, J.
    Czajka, J.
    Moore, C.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (954) : 476 - 482
  • [3] Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies
    Buckley, Jarryd G.
    Wilkinson, Dean
    Malaroda, Alessandra
    Metcalfe, Peter
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 174 - 183
  • [4] Dixon R.L., 2010, Comprehensive methodology for the evaluation of radiation dose in x-ray computed tomography: A New Measurement Paradigm Based on a Unified Theory for Axial, Helical, Fan-Beam
  • [5] Deep-inspiration breath-hold kilovoltage cone-beam CT for setup of stereotactic body radiation therapy for lung tumors: Initial experience
    Duggan, Dennis M.
    Ding, George X.
    Coffey, Charles W., II
    Kirby, Wyndee
    Hallahan, Dennis E.
    Malcolm, Arnold
    Lu, Bo
    [J]. LUNG CANCER, 2007, 56 (01) : 77 - 88
  • [6] American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning
    Fraass, B
    Doppke, K
    Hunt, M
    Kutcher, G
    Starkschall, G
    Stern, R
    Van Dyke, J
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 1773 - 1829
  • [7] ICRU, 1992, 46 ICRU
  • [8] International Electrotechnical Commission (IEC), 2012, MED EL EQ 2 44
  • [9] Japan Society of Medical Physics Task Group 01, 2008, JPN J MED PHYS, V27
  • [10] John MB., 2011, SIZE SPECIFIC DOSE E