Tracking moving objects with megavoltage portal imaging: A feasibility study

被引:37
作者
Meyer, Juergen [1 ]
Richter, Anne [1 ]
Baier, Kurt [1 ]
Wilbert, Juergen [1 ]
Guckenberger, Matthias [1 ]
Flentje, Michael [1 ]
机构
[1] Univ Wurzburg, Dept Radiat Oncol, D-97080 Wurzburg, Germany
关键词
tumor tracking; megavoltage (MV) portal imaging; the mean of the sum of squared differences (MSSD); mutual information; similarity measures;
D O I
10.1118/1.2191009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Four different algorithms were investigated with the aim to determine their suitability to track an object in conventional megavoltage portal images. The algorithms considered were the mean of the sum of squared differences (MSSD), mutual information (MI), the correlation ratio (CR), and the correlation coefficient (CC). Simulation studies were carried out with various image series containing a rigid object of interest that was moved along a predefined trajectory. For each of the series the signal-to-noise ratio (SNR) was varied to compare the performance of the algorithms under noisy conditions. For a poor SNR of -6 dB the mean tracking error was 2.4, 6.5, 39.0, and 17.2 pixels for MSSD, CC, CR and MI, respectively, with a standard deviation of 1.9, 12.9, 19.5, and 7.5 pixels, respectively. The size of a pixel was 0.5 mm. These results improved to 1.1, 1.3, 1.3, and 2.0 pixels, respectively, with a standard deviation of 0.6, 0.8, 0.8, and 2.1 pixels, respectively, when a mean filter was applied to the images prior to tracking. The implementation of MSSD into existing in-house software demonstrated that, depending on the search range, it was possible to process between 2 and 15 images/s, making this approach capable of real-time applications. In conclusion, the best geometric tracking accuracy overall was obtained with MSSD, followed by CC, CR, and MI. The simplest and best algorithm, both in terms of geometric accuracy as well as computational cost, was the MSSD algorithm and was therefore the method of choice. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1275 / 1280
页数:6
相关论文
共 39 条
  • [21] Tracking moving organs in real time
    Murphy, MJ
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) : 91 - 100
  • [22] Synchronized moving aperture radiation therapy (SMART): average tumour trajectory for lung patients
    Neicu, T
    Shirato, H
    Seppenwoolde, Y
    Jiang, SB
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (05) : 587 - 598
  • [23] A comparison of similarity measures for use in 2-D-3-D medical image registration
    Penney, GP
    Weese, J
    Little, JA
    Desmedt, P
    Hill, DLG
    Hawkes, DJ
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (04) : 586 - 595
  • [24] Moving targets: detection and tracking of internal organ motion for treatment planning and patient set-up
    Rietzel, E
    Rosenthal, SJ
    Gierga, DP
    Willett, CG
    Chen, GTY
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S68 - S72
  • [25] ROCHE A, 1998, MICCAI1998
  • [26] Image-guided hypo-fractionated stereotactic radiosurgery to spinal lesions
    Ryu, SI
    Chang, SD
    Kim, DH
    Murphy, MJ
    Le, QT
    Martin, DP
    Adler, JR
    [J]. NEUROSURGERY, 2001, 49 (04) : 838 - 846
  • [27] A novel tracking technique for the continuous precise measurement of tumour positions in conformal radiotherapy
    Seiler, PG
    Blattmann, H
    Kirsch, S
    Muench, RK
    Schilling, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (09) : N103 - N110
  • [28] Initial experience with ultrasound localization for positioning prostate cancer patients for external beam radiotherapy
    Serago, CF
    Chungbin, SJ
    Buskirk, SJ
    Ezzell, GA
    Collie, AC
    Vora, SA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1130 - 1138
  • [29] Prediction of respiratory tumour motion for real-time image-guided radiotherapy
    Sharp, GC
    Jiang, SB
    Shimizu, S
    Shirato, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (03) : 425 - 440
  • [30] Feasibility of insertion/implantation of 2.0-mm-diameter gold internal fiducial markers for precise setup and real-time tumor tracking in radiotherapy
    Shirato, H
    Harad, T
    Harabayashi, T
    Hida, K
    Endo, H
    Kitamura, K
    Onimaru, R
    Yamazaki, K
    Kurauchi, N
    Shimizu, T
    Shinohara, N
    Matsushita, M
    Dosaka-Akita, H
    Miyasaka, K