Tracking moving objects with megavoltage portal imaging: A feasibility study

被引:37
作者
Meyer, Juergen [1 ]
Richter, Anne [1 ]
Baier, Kurt [1 ]
Wilbert, Juergen [1 ]
Guckenberger, Matthias [1 ]
Flentje, Michael [1 ]
机构
[1] Univ Wurzburg, Dept Radiat Oncol, D-97080 Wurzburg, Germany
关键词
tumor tracking; megavoltage (MV) portal imaging; the mean of the sum of squared differences (MSSD); mutual information; similarity measures;
D O I
10.1118/1.2191009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Four different algorithms were investigated with the aim to determine their suitability to track an object in conventional megavoltage portal images. The algorithms considered were the mean of the sum of squared differences (MSSD), mutual information (MI), the correlation ratio (CR), and the correlation coefficient (CC). Simulation studies were carried out with various image series containing a rigid object of interest that was moved along a predefined trajectory. For each of the series the signal-to-noise ratio (SNR) was varied to compare the performance of the algorithms under noisy conditions. For a poor SNR of -6 dB the mean tracking error was 2.4, 6.5, 39.0, and 17.2 pixels for MSSD, CC, CR and MI, respectively, with a standard deviation of 1.9, 12.9, 19.5, and 7.5 pixels, respectively. The size of a pixel was 0.5 mm. These results improved to 1.1, 1.3, 1.3, and 2.0 pixels, respectively, with a standard deviation of 0.6, 0.8, 0.8, and 2.1 pixels, respectively, when a mean filter was applied to the images prior to tracking. The implementation of MSSD into existing in-house software demonstrated that, depending on the search range, it was possible to process between 2 and 15 images/s, making this approach capable of real-time applications. In conclusion, the best geometric tracking accuracy overall was obtained with MSSD, followed by CC, CR, and MI. The simplest and best algorithm, both in terms of geometric accuracy as well as computational cost, was the MSSD algorithm and was therefore the method of choice. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1275 / 1280
页数:6
相关论文
共 39 条
  • [1] Image-guided robotic radiosurgery
    Adler, JR
    Murphy, MJ
    Chang, SD
    Hancock, SL
    [J]. NEUROSURGERY, 1999, 44 (06) : 1299 - 1306
  • [2] BAIER K, 2005, Z MED PHYS, V15, P1
  • [3] BAIER K, 2005, 1J INT C MED PHYS NU
  • [4] Accuracy of a wireless localization system for radiotherapy
    Balter, JM
    Wright, JN
    Newell, LJ
    Friemel, B
    Dimmer, S
    Cheng, Y
    Wong, J
    Vertatschitsch, E
    Mate, TP
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (03): : 933 - 937
  • [5] A computerized remote table control for fast on-line patient repositioning: Implementation and clinical feasibility
    Bel, A
    Petrascu, O
    Van de Vondel, I
    Coppens, L
    Linthout, N
    Verellen, D
    Storme, G
    [J]. MEDICAL PHYSICS, 2000, 27 (02) : 354 - 358
  • [6] Patient setup error measurement using 3D intensity-based image registration techniques
    Clippe, S
    Sarrut, D
    Malet, C
    Miguet, S
    Ginestet, C
    Carrie, C
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01): : 259 - 265
  • [7] Real-time intra-fraction-motion tracking using the treatment couch: a feasibility study
    D'Souza, WD
    Naqvi, SA
    Yu, CX
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) : 4021 - 4033
  • [8] WHEN AND HOW CAN WE IMPROVE PRECISION IN RADIOTHERAPY
    DUTREIX, A
    [J]. RADIOTHERAPY AND ONCOLOGY, 1984, 2 (04) : 275 - 292
  • [9] How much margin reduction is possible through gating or breath hold?
    Engelsman, M
    Sharp, GC
    Bortfeld, T
    Onimaru, R
    Shirato, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (03) : 477 - 490
  • [10] A comparison of image registration techniques for the correlation of radiolabelled antibody distribution with tumour morphology
    Flynn, AA
    Green, AJ
    Boxer, G
    Pedley, RB
    Begent, RHJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) : N151 - N159