Radiological properties of the PRESAGE and PAGAT polymer dosimeters

被引:99
作者
Brown, S. [1 ,2 ]
Venning, A. [3 ]
De Deene, Y. [4 ]
Vial, P. [1 ,2 ]
Oliver, L. [1 ,2 ]
Adamovics, J. [5 ]
Baldock, C. [1 ]
机构
[1] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
[2] Royal N Shore Hosp, No Sydney Canc Ctr, Dept Radiat Oncol, St Leonards, NSW 2065, Australia
[3] Wellington Hosp, Wellington Canc Ctr, Dept Radiat Oncol, Wellington, New Zealand
[4] Ghent Univ Hosp, MR Dept, B-9000 Ghent, Belgium
[5] Rider Univ, Dept Biol & Chem, Lawrenceville, NJ 08648 USA
关键词
Polymer dosimetry; Gel dosimetry; PRESAGE; PAGAT; Radiological properties;
D O I
10.1016/j.apradiso.2008.06.005
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The radiological properties of the PRESAGE and PAGAT polymer dosimeters have been investigated and their water equivalence determined for use in radiotherapy dosimetry. The radiological water equivalence of each of the polymer dosimeters was determined by comparing the photon and electron interaction cross-sections over the 10 keV-20 MeV energy range and by Monte Carlo modelling the depth dose from a linear accelerator using the BEAMnrc software package. PRESAGE was found to have an effective Z-value and mass density (kgm(-3)) approximately 17% and 10% higher than water, respectively. A maximum difference of 85% was discovered in the photoelectric interaction probability curve of PRESAGE when compared to water over the energy range 10-100keV, partially due to the Z(3) dependence of the photoelectric effect. The mass radiative stopping power ratios and mass scattering power ratios were both found to have less than 9% difference from water. The depth dose for PRESAGE from a 6 MV photon beam had an absolute percentage difference to water of less than 2% and a relative percentage difference of less than 8%. The mass density of PAGAT was found to be 2.6% higher than Water due to its high gelatine and monomer concentration. The cross-sectional attenuation and absorption coefficient ratios were found to be within 5% for energies between 10 and 100keV and within 1% for energies between 100keV and 20 MeV. The mass collisional stopping power, mass radiative stopping power and mass scattering power ratios were all less than 1% over the energy range studied. The depth dose had an absolute percentage difference to water of less than 1% and a relative percentage difference of less than 2.5%. These results indicate that the PAGAT polymer gel formulation is more radiological water equivalent than the PRESAGE formulation. However, the PRESAGE dosimeter offers some advantages in terms of ease of use and its lack of water equivalence may be overcome with dosimetric correction factors. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1970 / 1974
页数:5
相关论文
共 16 条
[1]   Experimental procedure for the manufacture and calibration of polyacrylamide gel (PAG) for magnetic resonance imaging (MRI) radiation dosimetry [J].
Baldock, C ;
Burford, RP ;
Billingham, N ;
Wagner, GS ;
Patval, S ;
Badawi, RD ;
Keevil, SF .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (03) :695-702
[2]  
Berger M.J., 1995, STOPPING POWER RANGE
[3]  
De Deene Y, 2000, MAGN RESON MED, V43, P116, DOI 10.1002/(SICI)1522-2594(200001)43:1<116::AID-MRM14>3.0.CO
[4]  
2-5
[5]   A practical three-dimensional dosimetry system for radiation therapy [J].
Guo, Pengyi ;
Adamovics, John ;
Oldham, Mark .
MEDICAL PHYSICS, 2006, 33 (10) :3962-3972
[6]  
Hubbell J. H., 2004, Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients, DOI DOI 10.18434/T4D01F
[7]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[8]  
JOHNS JE, 1983, PHYS RADIOLOGY
[9]  
Keall P, 1999, Australas Phys Eng Sci Med, V22, P85
[10]  
KHAN FM, 1994, PHYS RADIOTHERAPY