Vibration characteristics of piezoelectric lead zirconate titanate by fluid flow in intravascular oxygenator

被引:18
作者
Kim, Gi-Beum [1 ]
Hong, Chul-Un [1 ]
Kwon, Tae-Kyu [1 ]
机构
[1] Chonbuk Natl Univ, Coll Engn, Div Bion & Bioinformat, Chonju 561756, South Korea
来源
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS | 2006年 / 45卷 / 4B期
关键词
hollow fiber membrane; PVDF sensor; PZT actuator; oxygen transfer; membrane oxygenator; artificial lung; vibrating intravascular lung assist device (VIVLAD);
D O I
10.1143/JJAP.45.3811
中图分类号
O59 [应用物理学];
学科分类号
摘要
In this paper, we describe the enhancement in oxygen transfer rate in hollow-fiber-membrane (HFM) modules using a plumbum piezoelectric lead zirconate titanate (PZT) actuator and a piezoelectric poly vinylidene fluoride (PVDF) sensor. We developed a vibrating intravascular lung assist device (VIVLAD) for patients having chronic respiratory problems and performed experiments on hydrodynamic flow through a bundle of sinusoidal vibrating hollow fibers. These modules were used to provide some insight into how wall vibrations might enhance the performance of an intravascular lung assist device. The experimental design and procedure are then applied to the fabrication of a device used to assess the effectiveness of membrane vibrations. The test section was a cylindrical duct with an inner diameter of 30 mm. The flow rate was controlled by a pump and monitored by a built-in flowmeter. The vibration apparatus was composed of a piezovibrator, a function generator, and a power amplifier. The direction of vibration was radial to the fluid flow. Gas flow rates of up to 6 L/min through 120-cm-long hollow fibers were achieved by exciting the piezovibrator. The time and frequency responses of PVDF sensors were investigated through various frequencies in VIVLAD. In these devices, the flow of blood and the source of oxygen were separated by it semipermeable membrane that allows oxygen and carbon dioxide to diffuse into and out of the fluid, respectively. Results of the experiments have shown that a vibrating intravascular lung assist device performs effectively.
引用
收藏
页码:3811 / 3817
页数:7
相关论文
共 7 条
[1]   Gas flow dynamics in hollow-fiber membranes [J].
Federspiel, WJ ;
Williams, JL ;
Hattler, BG .
AICHE JOURNAL, 1996, 42 (07) :2094-2099
[2]   Results of an artificial-lung survey to lung transplant program directors [J].
Haft, JW ;
Griffith, BP ;
Hirschl, RB ;
Bartlett, RH .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2002, 21 (04) :467-473
[3]  
Hattler B G, 1992, ASAIO J, V38, pM322, DOI 10.1097/00002480-199207000-00046
[4]   DEVELOPMENT OF AN INTRAVENOUS MEMBRANE-OXYGENATOR - ENHANCED INTRAVENOUS GAS-EXCHANGE THROUGH CONVECTIVE MIXING OF BLOOD AROUND HOLLOW-FIBER MEMBRANES [J].
HATTLER, BG ;
REEDER, GD ;
SAWZIK, PJ ;
LUND, LW ;
WALTERS, FR ;
SHAH, AS ;
RAWLEIGH, J ;
GOODE, JS ;
KLAIN, M ;
BOROVETZ, HS .
ARTIFICIAL ORGANS, 1994, 18 (11) :806-812
[5]  
KIM GB, 2004, CHEM ENG, V44, P151
[6]  
KIM KB, 2000, J BIOMED ENG RES, V21, P311
[7]   Development of an implantable artificial lung: Challenges and progress [J].
Zwischenberger, JB ;
Anderson, CM ;
Cook, KE ;
Lick, SD ;
Mockros, LF ;
Bartlett, RH .
ASAIO JOURNAL, 2001, 47 (04) :316-320