Benchmarking of a full-body inertial motion capture system for clinical gait analysis

被引:78
作者
Cloete, Teunis [1 ]
Scheffer, Cornie [1 ]
机构
[1] Univ Stellenbosch, Biomed Engn Res Grp, Dept Mech & Mechatron Engn, ZA-7600 Stellenbosch, South Africa
来源
2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vols 1-8 | 2008年
关键词
D O I
10.1109/IEMBS.2008.4650232
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order for gait analysis to be established as part of routine clinical diagnoses, an accurate, flexible and user-friendly motion capture system is required. Commonly used optical, mechanical and acoustic systems offer acceptable accuracy and repeatability, but are often expensive and restricted to laboratory use. Inertial motion capture has seen great innovation in the last few years, but the technology is not yet considered mature enough for clinical gait analysis. In this paper we compare the kinematic reliability of inertial motion capture with optical motion capture during routine gait studies of eight able-bodied subjects. The root mean squared, RMS, and coefficient of correlation, R, was used to compare data sets. Saggital plane joint angles in the knee and hip compared very well. Corresponding transverse and frontal plane values were moderately accurate. The ankle joint angles calculated from the two systems were less accurate. This was believed to be due to the use of different rotation axis orientations used for calculation of angular rotations.
引用
收藏
页码:4579 / 4582
页数:4
相关论文
共 14 条
[1]   Repeatability of gait data using a functional hip joint centre and a mean helical knee axis [J].
Besier, TF ;
Sturnieks, DL ;
Alderson, JA ;
Lloyd, DG .
JOURNAL OF BIOMECHANICS, 2003, 36 (08) :1159-1168
[2]   A new approach to accurate measurement of uniaxial joint angles based on a combination of accelerometers and gyroscopes [J].
Dejnabadi, H ;
Jolles, BM ;
Aminian, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (08) :1478-1484
[3]   Estimation and visualization of sagittal kinematics of lower limbs orientation using body-fixed sensors [J].
Dejnabadi, Hooman ;
Jolles, Brigitte M. ;
Casanova, Emilio ;
Fua, Pascal ;
Aminian, Kamiar .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (07) :1385-1393
[4]  
DIEBEL J, 2006, 943019010 STANF U, P51903
[5]  
FAVRE J, 2006, P ISB 3D 3 D AN HUM, P51903
[6]   A JOINT COORDINATE SYSTEM FOR THE CLINICAL DESCRIPTION OF 3-DIMENSIONAL MOTIONS - APPLICATION TO THE KNEE [J].
GROOD, ES ;
SUNTAY, WJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (02) :136-144
[7]   REPEATABILITY OF KINEMATIC, KINETIC, AND ELECTROMYOGRAPHIC DATA IN NORMAL ADULT GAIT [J].
KADABA, MP ;
RAMAKRISHNAN, HK ;
WOOTTEN, ME ;
GAINEY, J ;
GORTON, G ;
COCHRAN, GVB .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1989, 7 (06) :849-860
[8]   Skin movement artefact assessment and compensation in the estimation of knee-joint kinematics [J].
Lucchetti, L ;
Cappozzo, A ;
Cappello, A ;
Della Croce, U .
JOURNAL OF BIOMECHANICS, 1998, 31 (11) :977-984
[9]   Measuring orientation of human body segments using miniature gyroscopes and accelerometers [J].
Luinge, HJ ;
Veltink, PH .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (02) :273-282
[10]  
ROETENBERG D, 2007, MOVEN FULL 6DO UNPUB, P51903