Monte Carlo calculations of AAPM Task Group Report No. 43 dosimetry parameters for the 125I I-Seed AgX100 source model

被引:14
作者
Mourtada, Firas [1 ,2 ]
Mikell, Justin [3 ]
Ibbott, Geoffrey [4 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 1210, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Expt Diagnost Imaging, Houston, TX 77030 USA
[3] Univ Texas Houston, Grad Sch Biomed Sci, Houston, TX USA
[4] Univ Texas MD Anderson Canc Ctr, Radiol Phys Ctr, Houston, TX 77030 USA
关键词
I-125; seed; Brachytherapy; Dosimetry; Monte Carlo; BRACHYTHERAPY SEED;
D O I
10.1016/j.brachy.2011.06.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of this study was to determine the dosimetric parameters of the AgX100, a new I-125 brachytherapy seed model, using Monte Carlo (MC) simulations according to the protocol specified by the updated American Association of Physicists in Medicine Task Group No. 43 Report (TG-43U1) and compare these parameters with those of the established brachytherapy I-125 seed models 6711 and I25.S06. METHODS AND MATERIALS: Independent verification of the new seed geometry was performed using high-resolution digital radiography and scanning electron microscopy. MCNPX v.2.5 MC simulations of the AgX100 seed were performed to derive its TG-43U1 parameters, the dose rate constant, the radial dose function, and the two- and one-dimensional anisotropy functions in liquid water. A dosimetric error propagation analysis was also performed to include uncertainty because of seed manufacturing tolerances and physics parameters. RESULTS: The MC-calculated dose rate constant for the AgX100 seed was 0.943 cGy.h(-1).U-1 +/- 2.6% (k = 1) based on the air kerma strength for a simulated point detector. Tabulated results of the radial dose function for line and point source approximations and the two-dimensional anisotropy function are also reported. CONCLUSIONS: The MC-predicted dose distribution of the AgX100 seed was found to be comparable with that of the model 6711 seed but much different from the dose distribution of the model 125.S06 seeds. However, at shallow distances, there were some dosimetric differences between the AgX100 and 6711 seed, which warrant separate TG-43U1 parameters for use in clinical treatment planning systems. (C) 2012 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:237 / 244
页数:8
相关论文
共 12 条
  • [1] [Anonymous], 1999, CRC handbook of chemistry and physics
  • [2] Cullen D.E., 1997, EPDL97 EVALUATED PHO, V6
  • [3] An analysis of MCNP cross-sections and tally methods for low-energy photon emitters
    DeMarco, JJ
    Wallace, RE
    Boedeker, K
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) : 1321 - 1332
  • [4] Monte Carlo and experimental dosimetry of an 125I brachytherapy seed
    Dolan, James
    Li, Zuofeng
    Williamson, Jeffrey F.
    [J]. MEDICAL PHYSICS, 2006, 33 (12) : 4675 - 4684
  • [5] American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning
    Fraass, B
    Doppke, K
    Hunt, M
    Kutcher, G
    Starkschall, G
    Stern, R
    Van Dyke, J
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 1773 - 1829
  • [6] Monte Carlo-aided dosimetry of the symmetra model I25,S06 125I, interstitial brachytherapy seed
    Hedtjärn, H
    Carlsson, GA
    Williamson, JF
    [J]. MEDICAL PHYSICS, 2000, 27 (05) : 1076 - 1085
  • [7] Brachytherapy dosimetry parameters calculated for a 131Cs source
    Rivard, Mark J.
    [J]. MEDICAL PHYSICS, 2007, 34 (02) : 754 - 762
  • [8] Monte Carlo radiation dose simulations and dosimetric comparison of the model 6711 and 9011 125I brachytherapy sources
    Rivard, Mark J.
    [J]. MEDICAL PHYSICS, 2009, 36 (02) : 486 - 491
  • [9] Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations
    Rivard, MJ
    Coursey, BM
    DeWerd, LA
    Hanson, WF
    Huq, MS
    Ibbott, GS
    Mitch, MG
    Nath, R
    Williamson, JF
    [J]. MEDICAL PHYSICS, 2004, 31 (03) : 633 - 674
  • [10] Benchmarking BrachyDose: Voxel based EGSnrc Monte Carlo calculations of TG-43 dosimetry parameters
    Taylor, R. E. P.
    Yegin, G.
    Rogers, D. W. O.
    [J]. MEDICAL PHYSICS, 2007, 34 (02) : 445 - 457