Achilles Tension Mitigates Fibular Malalignment Measured in Cadaveric Studies of Syndesmotic Clamping

被引:5
作者
Goetz, Jessica E. [1 ]
Szabo, Nicole [1 ]
Rudert, M. James [1 ]
Karam, Matthew D. [1 ]
Phisitkul, Phinit [2 ]
机构
[1] Univ Iowa, Dept Orthoped & Rehabil, Iowa City, IA USA
[2] Tristate Specialists LLP, Sioux City, IA USA
关键词
syndesmosis; syndesmotic reduction; CT scan; ROTATIONAL ANKLE FRACTURES; SCREW FIXATION; REDUCTION; MALREDUCTION; JOINT; ACCURACY;
D O I
10.1177/1071100718816037
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Fibular malreduction is becoming a commonly recognized complication of surgical repair of the syndesmosis when a reduction clamp is used. The goal of this work was to determine the interdependent effects of transsyndesmotic reduction clamp position and applied compression force on fibular alignment in a realistic cadaveric preparation of complete syndesmotic injury. Methods: Six through-the-knee cadaveric specimens were CT scanned intact, with the distal syndesmosis fully destabilized, and with 53, 102, and 160 N clamping forces each applied along an anteriorly, centrally, and posteriorly directed transsyndesmotic axis. Testing was repeated incorporating 178 N of Achilles tendon tension using all 3 clamping forces applied along the centrally directed axis. Fibular reduction was automatically quantified from CT scan-generated bony surfaces as rotation of the fibula around the tibia, rotation of the fibula within the incisura, medial/lateral fibular displacement, and anterior/posterior fibular displacement. Results: Transsyndesmotic clamping along the anteriorly directed axis resulted in the best reduction quality by all 4 quantified measures. Along the centrally and posteriorly directed axes, progressively greater forces caused significantly greater sagittal plane fibular malreduction. Addition of Achilles tension reduced the magnitude of fibular malreduction and overcompression. Conclusion: Placing the medial tine of a transsyndesmotic reduction clamp on the anterior medial tibia resulted in the most accurate syndesmotic reduction and provided some protection against overcompression with large reduction clamp forces. Achilles tension appeared to contribute to reduction, decreasing the magnitude of measured malreduction from clamping.
引用
收藏
页码:465 / 474
页数:10
相关论文
共 30 条
  • [1] Radiographic Change of the Distal Tibiofibular Joint Following Removal of Transfixing Screw Fixation
    Baek, Jong Hun
    Kim, Tae Yong
    Kwon, Yoo Beom
    Jeong, Bi O.
    [J]. FOOT & ANKLE INTERNATIONAL, 2018, 39 (03) : 318 - 325
  • [2] Functional Outcomes of Syndesmotic Injuries Based on Objective Reduction Accuracy at a Minimum 1-Year Follow-Up
    Cherney, Steven M.
    Cosgrove, Christopher T.
    Spraggs-Hughes, Amanda G.
    McAndrew, Christopher M.
    Ricci, William M.
    Gardner, Michael J.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (01) : 43 - 51
  • [3] In Vivo Syndesmotic Overcompression After Fixation of Ankle Fractures With a Syndesmotic Injury
    Cherney, Steven M.
    Haynes, Jacob A.
    Spraggs-Hughes, Amanda G.
    McAndrew, Christopher M.
    Ricci, William M.
    Gardner, Michael J.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (09) : 414 - 419
  • [4] CHISSELL HR, 1995, J BONE JOINT SURG BR, V77B, P435
  • [5] CIBC, 2016, VOL IM SEGM VIS
  • [6] Medial Clamp Tine Positioning Affects Ankle Syndesmosis Malreduction
    Cosgrove, Christopher T.
    Putnam, Sara M.
    Cherney, Steven M.
    Ricci, William M.
    Spraggs-Hughes, Amanda
    McAndrew, Christopher M.
    Gardner, Michael J.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (08) : 440 - 446
  • [7] Intraoperative Syndesmotic Reduction: Three-Dimensional Versus Standard Fluoroscopic Imaging
    Davidovitch, Roy I.
    Weil, Yoram
    Karia, Raj
    Forman, Jordanna
    Looze, Christopher
    Liebergall, Meir
    Egol, Kenneth
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (20) : 1838 - 1843
  • [8] Ebinger Thomas, 2013, Iowa Orthop J, V33, P40
  • [9] Use of a Pointed Reduction Clamp Placed on the Distal Fibula to Ensure Proper Restoration of Fibular Length and Rotation and Anatomic Reduction of the Syndesmosis: A Technique Tip
    Fitzpatrick, Ellen P.
    Kwon, John Y.
    [J]. FOOT & ANKLE INTERNATIONAL, 2014, 35 (09) : 943 - 948
  • [10] Intraoperative Three-Dimensional Imaging in the Treatment of Acute Unstable Syndesmotic Injuries
    Franke, Jochen
    von Recum, Jan
    Suda, Arnold J.
    Gruetzner, Paul Alfred
    Wendl, Klaus
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (15) : 1386 - 1390