Comparison of cascade impaction and laser diffraction for particle size distribution measurements

被引:40
|
作者
Ziegler, J [1 ]
Wachtel, H [1 ]
机构
[1] Boehringer Ingelheim Pharma GmbH & Co KG, D-55218 Ingelheim, Germany
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 2005年 / 18卷 / 03期
关键词
soft mist inhaler; particle size distribution; impactor; laser diffraction;
D O I
10.1089/jam.2005.18.311
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
The Andersen cascade impactor (ACI) and laser diffraction (LD) can be correlated at ambient temperature for aqueous drug formulations atomized by Soft MiSt (TM) inhalers. A comparison of the two particle size determination methods at different conditions (flow rate, relative humidity) was performed. Under well-defined conditions, the faster LD can substitute the time-consuming ACI at least for routine tests. The measurements were performed with three different drug formulations. The aerosol was generated by Soft Mist (TM) inhalers, and the droplet distributions were measured simultaneously using a laser diffraction analyzer together with the eight-stage Andersen cascade impactor. The simultaneous measurements ensure that aerosol and air conditions are identical for both LD and ACI. In order to measure the scattered laser light intensity of the aerosol passing the induction port, glass windows were fitted to the induction port. The evaporation effect of the aqueous aerosols on the PSD was investigated at ambient humidity and high humidity (RH > 90%). The simultaneous determination of the droplet size distribution leads to a good correlation between the ACI and LD method only if the measurements were performed at RH of > 90%. The humidity of the ambient air had the strongest influence on PSD not only for ACI, but also for LD. In our set-up, the almost saturated air prevents aqueous droplets from drying. The influence of the flow rate on LD was negligible, whereas for ACI, a flow rate dependence is expected. The advantages of LD and the demonstrated compatibility to established EP/USP methods motivate the substitution of the ACI and the use of LD for routine measurements.
引用
收藏
页码:311 / 324
页数:14
相关论文
共 50 条
  • [1] Inter-laboratory Study of Particle Size Distribution Measurements by Laser Diffraction
    Kuchenbecker, Petra
    Gemeinert, Marion
    Rabe, Torsten
    PARTICLE & PARTICLE SYSTEMS CHARACTERIZATION, 2012, 29 (04) : 304 - 310
  • [2] Technical note: Use of laser diffraction for particle size distribution measurements in duodenal digesta
    Olaisen, V
    Nesse, N
    Volden, H
    JOURNAL OF ANIMAL SCIENCE, 2001, 79 (03) : 761 - 765
  • [3] Using laser diffraction to measure particle size and distribution
    Levoguer, Carl
    Metal Powder Report, 2013, 68 (03) : 15 - 18
  • [4] Particle size distribution of the sludge by laser diffraction method
    Fryzlewicz-Kozak, Beata
    PRZEMYSL CHEMICZNY, 2018, 97 (12): : 2170 - 2172
  • [5] Comparison of optical particle sizing and cascade impaction for measuring the particle size of a suspension metered dose inhaler
    Pu, Yu
    Kline, Lukeysha C.
    Khawaja, Nazia
    Van Liew, Melissa
    Berry, Julianne
    DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 2015, 41 (05) : 737 - 743
  • [6] Comparison of nebulized particle size distribution with Malvern laser diffraction analyzer versus Andersen Cascade Impactor and low-flow marple personal cascade impactor
    Kwong, WTJ
    Ho, SL
    Coates, AL
    JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2000, 13 (04): : 303 - 314
  • [7] COMPARISON OF SIEVING AND LASER DIFFRACTION FOR THE PARTICLE-SIZE MEASUREMENTS OF RAW-MATERIALS USED IN FOODSTUFF
    NATHIERDUFOUR, N
    BOUGEARD, L
    DEVAUX, MF
    BERTRAND, D
    DEMONREDON, FL
    POWDER TECHNOLOGY, 1993, 76 (02) : 191 - 200
  • [8] Particle characterization for oil sand processing - 2: Particle size measurements by laser diffraction
    Cowles, RJH
    PETROLEUM SCIENCE AND TECHNOLOGY, 2000, 18 (1-2) : 203 - 220
  • [9] Determination of Particle Size Distribution: Comparison of Standard Hydrometer Method and Laser Diffraction Analysis for Use in Forestry
    Sedlackova, Katerina
    Sevelova, Lenka
    Igaz, Dusan
    Aydin, Elena
    FORESTS, 2024, 15 (02):
  • [10] Comparison of the particle size distribution of fluticasone propionate/formoterol and beclometasone/formoterol MDI using laser diffraction
    Haidl, Peter
    Scheuch, Gerhard
    Kietzig, Claudius
    Kappeler, Dominik
    Cloes, Rolf
    EUROPEAN RESPIRATORY JOURNAL, 2014, 44