T2 Measurement in Articular Cartilage: Impact of the Fitting Method on Accuracy and Precision at Low SNR

被引:138
作者
Raya, Jose G. [1 ]
Dietrich, Olaf [1 ]
Horng, Annie
Weber, Juergen [1 ]
Reiser, Maximilian F. [1 ]
Glaser, Christian
机构
[1] Univ Munich, Josef Lissner Lab Biomed Imaging, Dept Clin Radiol Grosshadern, D-81377 Munich, Germany
关键词
T(2) relaxation time; cartilage; SNR; T(2) fitting; accuracy; precision; T2; RELAXATION-TIME; PATELLAR CARTILAGE; SPATIAL VARIATION; MAGNITUDE IMAGES; MR-IMAGES; NOISE; REPRODUCIBILITY; PROPAGATION; VALUES; NMR;
D O I
10.1002/mrm.22178
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
T(2) relaxation time is a promising MRI parameter for the detection of cartilage degeneration in osteoarthritis. However, the accuracy and precision of the measured T(2) may be substantially impaired by the low signal-to-noise ratio of images available from clinical examinations. The purpose of this work was to assess the accuracy and precision of the traditional fit methods (linear least-squares regression and nonlinear fit to an exponential) and two new noise-corrected fit methods: fit to a noise-corrected exponential and fit of the noise-corrected squared signal intensity to an exponential. Accuracy and precision have been analyzed in simulations, in phantom measurements, and in seven repetitive acquisitions of the patellar cartilage in six healthy volunteers. Traditional fit methods lead to a poor accuracy for low T(2), with overestimations of the exact T(2) up to 500%. The noise-corrected fit methods demonstrate a very good accuracy for all T(2) values and signal-to-noise ratio. Even more, the fit to a noise-corrected exponential results in precisions comparable to the best achievable precisions (Cramer-Rao lower bound). For in vivo images, the traditional fit methods considerably overestimate T(2) near the bone-cartilage interface. Therefore, using an adequate fit method may substantially improve the sensitivity of T(2) to detect pathology in cartilage and change in T2 follow-up examinations. Magn Reson Med 63:181-193,2010. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:181 / 193
页数:13
相关论文
共 43 条
[1]  
Abragam A., 2004, PRINCIPLES NUCL MAGN
[2]   A theoretical study of the effect of experimental noise on the measurement of anisotropy in diffusion imaging [J].
Bastin, ME ;
Armitage, PA ;
Marshall, I .
MAGNETIC RESONANCE IMAGING, 1998, 16 (07) :773-785
[3]  
Blumenkrantz G, 2007, EUR CELLS MATER, V13, P75
[4]   T-2 maximum likelihood estimation from multiple spin-echo magnitude images [J].
Bonny, JM ;
Zanca, M ;
Boire, JY ;
Veyre, A .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (02) :287-293
[5]  
BRANDT S, 1989, STAT COMPUTATIONAL M, P138
[6]   Signal-to-noise measurements in magnitude images from NMR phased arrays [J].
Constantinides, CD ;
Atalar, E ;
McVeigh, ER .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (05) :852-857
[7]   Spatial variation of T2 in human articular cartilage [J].
Dardzinski, BJ ;
Mosher, TJ ;
Li, SZ ;
VanSlyke, MA ;
Smith, MB .
RADIOLOGY, 1997, 205 (02) :546-550
[8]  
Dietrich O, 2001, MAGNET RESON MED, V45, P448, DOI 10.1002/1522-2594(200103)45:3<448::AID-MRM1059>3.0.CO
[9]  
2-W
[10]   Influence of multichannel combination, parallel imaging and other reconstruction techniques on MRI noise characteristics [J].
Dietrich, Olaf ;
Raya, Jose G. ;
Reeder, Scott B. ;
Ingrisch, Michael ;
Reiser, Maximilian F. ;
Schoenberg, Stefan O. .
MAGNETIC RESONANCE IMAGING, 2008, 26 (06) :754-762