Effect of Decentration on the Optical Quality of Monofocal, Extended Depth of Focus, and Bifocal Intraocular Lenses

被引:35
作者
Xu, Jie [1 ,2 ,3 ,4 ]
Zheng, Tianyu [1 ,2 ,3 ,4 ]
Lu, Yi [1 ,2 ,3 ,4 ]
机构
[1] Fudan Univ, Eye & Ear Nose & Throat Hosp, Dept Ophthalmol, Shanghai, Peoples R China
[2] Fudan Univ, Eye & Ear Nose & Throat Hosp, Eye Inst, Shanghai, Peoples R China
[3] Minist Hlth, Key Lab Myopia, Shanghai, Peoples R China
[4] Shanghai Key Lab Visual Impairment & Restorat, Shanghai, Peoples R China
基金
中国国家自然科学基金;
关键词
TILT; PERFORMANCE; RANGE;
D O I
10.3928/1081597X-20190708-02
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the effect of intraocular lens (IOL) decentration on the optical quality in patients with implantation of three IOLs: monofocal, extended depth of focus (EDOF), and bifocal. METHODS: Patients had cataract surgery with implantation of one of the three above-mentioned IOLs. Higher order aberrations (HOAs), modulation transfer function (MTF), point spread function (PSF), retinal straylight, and dysphotopsia phenomena were evaluated 3 months after surgery. IOL decentration was quantified as the distance between the visual axis center and the IOL center using the OPD-Scan III aberrometer (Nidek Co., Ltd., Gamagori, Japan). The patients who received each IOL type were then divided into two subgroups (decentration of <= 0.25 or > 0.25 mm) to analyze the effect of IOL decentration on these optical qualities. RESULTS: The study included 54 eyes (54 patients), with 18 eyes in each IOL group. The distance of IOL decentration did not differ significantly among the three groups. With a decentration of more than 0.25 mm, MTF, PSF, and coma were only significantly deteriorated in the bifocal IOL (ZMB00; Abbott Medical Optics, Santa Ana, CA). HOAs, coma, PSF, and glare perception were better in the monofocal and EDOF IOLs than those in the ZMB00 IOL when decentration was more than 0.25 mm. Furthermore, IOL decentration was significantly correlated with HOAs, coma, MTF, and PSF in the ZMB00 IOL. CONCLUSIONS: The monofocal and EDOF IOLs are more immune to optical quality degradation caused by IOL decentration than the ZMB00 IOL.
引用
收藏
页码:484 / +
页数:10
相关论文
共 20 条
[1]   Extended depth of focus intraocular lenses for presbyopia [J].
Akella, Sruti S. ;
Juthani, Viral V. .
CURRENT OPINION IN OPHTHALMOLOGY, 2018, 29 (04) :319-323
[2]   Multifocal intraocular lenses: An overview [J].
Alio, Jorge L. ;
Plaza-Puche, Ana B. ;
Fernandez-Buenaga, Roberto ;
Pikkel, Joseph ;
Maldonado, Miguel .
SURVEY OF OPHTHALMOLOGY, 2017, 62 (05) :611-634
[3]   Extended depth of focus contact lenses vs. two commercial multifocals: Part 1. Optical performance evaluation via computed through-focus retinal image quality metrics [J].
Bakaraju, Ravi C. ;
Ehrmann, Klaus ;
Ho, Arthur .
JOURNAL OF OPTOMETRY, 2018, 11 (01) :10-20
[4]   Tilt and decentration of spherical and aspheric intraocular lenses: Effect on higher-order aberrations [J].
Baumeister, Martin ;
Buehren, Jens ;
Kohnen, Thomas .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (06) :1006-1012
[5]   A New Extended Depth of Focus Intraocular Lens Based on Spherical Aberration [J].
Bellucci, Roberto ;
Curatolo, Maria C. .
JOURNAL OF REFRACTIVE SURGERY, 2017, 33 (06) :389-+
[6]   A Comparative Evaluation of a New Generation of Diffractive Trifocal and Extended Depth of Focus Intraocular Lenses [J].
Cochener, Beatrice ;
Boutillier, Guillaume ;
Lamard, Mathieu ;
Auberger-Zagnoli, Claire .
JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (08) :507-+
[7]   Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye [J].
Eppig, Timo ;
Scholz, Katja ;
Loeffler, Andre ;
Messner, Arthur ;
Langenbucher, Achim .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (06) :1091-1100
[8]   Variability in angle κ and its influence on higher-order aberrations in pseudophakic eyes [J].
Harrer, Annette ;
Hirnschall, Nino ;
Tabernero, Juan ;
Artal, Pablo ;
Draschl, Petra ;
Maedel, Sophie ;
Findl, Oliver .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (08) :1015-1019
[9]   Comparison of long-term decentration and tilt in two types of multifocal intraocular lenses with OPD-Scan III aberrometer [J].
He, Wenwen ;
Qiu, Xiaodi ;
Zhang, Shaohua ;
Du, Yu ;
Zhang, Yinglei ;
Lu, Yi ;
Zhu, Xiangjia .
EYE, 2018, 32 (07) :1237-1243
[10]   Comparing visual acuity, range of vision and spectacle independence in the extended range of vision and monofocal intraocular lens [J].
Hogarty, Daniel T. ;
Russell, Deborah J. ;
Ward, Bernadette M. ;
Dewhurst, Nicholas ;
Burt, Peter .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2018, 46 (08) :854-860