Fiber-Optic Probes for Small-Scale Measurements of Scalar Irradiance

被引:27
|
作者
Rickelt, Lars Fledelius [1 ]
Lichtenberg, Mads [1 ]
Trampe, Erik Christian Lovbjerg [1 ]
Kuehl, Michael [1 ,2 ]
机构
[1] Univ Copenhagen, Marine Biol Sect, Dept Biol, Helsingor, Denmark
[2] Univ Technol Sydney, Plant Funct Biol & Climate Change Cluster, Ultimo Sydney, NSW, Australia
关键词
SPECTRAL LIGHT MEASUREMENTS; SCATTERING BULBS; MICROBIAL MAT; PHOTOSYNTHESIS; DOSIMETRY; MICROPROBES; COMMUNITIES; CALIBRATION; RESOLUTION; GRADIENTS;
D O I
10.1111/php.12560
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
A new method for producing fiber-optic microprobes for scalar irradiance (=fluence rate) measurements is described. Such fine-scale measurements are important in many photobiological disciplines. With the new method, it is possible to cast spherical 30-600 m wide light integrating sensor tips onto tapered or untapered optical fibers. The sensor tip is constructed by first casting a clear polymethyl methacrylate (PMMA) sphere (similar to 80% of the size of the final probe tip diameter) onto the optical fiber via dip-coating. Subsequently, the clear sphere is covered with light diffusing layers of PMMA mixed with TiO2 until the fiber probe exhibits a satisfactory isotropic response (typically +/- 5-10%). We also present an experimental setup for measuring the isotropic response of fiber-optic scalar irradiance probes in air and water. The fiber probes can be mounted in a syringe equipped with a needle, facilitating retraction of the spherical fiber tip. This makes it, e.g. possible to cut a hole in cohesive tissue with the needle before inserting the probe. The light-collecting properties of differently sized scalar irradiance probes (30, 40, 100, 300 and 470 m) produced by this new method were compared to probes produced with previously published methods. The new scalar irradiance probes showed both higher throughput of light, especially for blue light, as well as a better isotropic light collection over a wide spectral range. The new method also allowed manufacturing of significantly smaller scalar irradiance microprobes (down to 30 m tip diameter) than hitherto possible, and such sensors allow minimally invasive high-resolution scalar irradiance measurements in thin biofilms, leaves and animal tissues.
引用
收藏
页码:331 / 342
页数:12
相关论文
共 50 条
  • [21] Characterization of laser-induced plasmas for fiber-optic probes
    Stratis, DN
    Eland, KL
    Angel, SM
    ENVIRONMENTAL MONITORING AND REMEDIATION TECHNOLOGIES, 1999, 3534 : 592 - 600
  • [22] Fiber-optic probes for biomedical optoacoustic imaging, monitoring, and sensing
    Esenaliev, Rinat
    SEVENTH EUROPEAN WORKSHOP ON OPTICAL FIBRE SENSORS (EWOFS 2019), 2019, 11199
  • [23] Fiber-Optic Probes for Ring-Shaped Multiparticle Capture
    Wang, Tao
    Yao, Linzhi
    Gao, Bingkun
    Jiang, Chunlei
    Wang, Yunkai
    Qan, Zhaoi
    Sun, Jiayao
    Dong, Taiji
    Sun, Yu
    Dai, Penghui
    IEEE SENSORS JOURNAL, 2024, 24 (05) : 6129 - 6135
  • [24] Clinical applications of fiber-optic probes in optical coherence tomography
    McLaughlin, Robert A.
    Sampson, David D.
    OPTICAL FIBER TECHNOLOGY, 2010, 16 (06) : 467 - 475
  • [25] Small-scale measurement of soil water content using a fiber optic sensor
    Garrido, F
    Ghodrati, M
    Chendorain, M
    SOIL SCIENCE SOCIETY OF AMERICA JOURNAL, 1999, 63 (06) : 1505 - 1512
  • [26] RESPIRATION MEASUREMENTS IN SMALL-SCALE
    RASMUSSEN, HN
    RASMUSSEN, UF
    ANALYTICAL BIOCHEMISTRY, 1993, 208 (02) : 244 - 248
  • [27] Geomechanical modeling of distributed fiber-optic sensor measurements
    Sherman, Christopher
    Mellors, Robert
    Morris, Joseph
    Ryerson, Frederick
    INTERPRETATION-A JOURNAL OF SUBSURFACE CHARACTERIZATION, 2019, 7 (01): : SA21 - SA27
  • [28] Hollow fiber-optic probe for in vivo Raman measurements
    Katagiri, Takashi
    Hattori, Yusuke
    Komachi, Yuichi
    Matsuura, Yuji
    Tashiro, Hideo
    Sato, Hidetoshi
    SMART MEDICAL AND BIOMEDICAL SENSOR TECHNOLOGY IV, 2006, 6380
  • [29] LIGHT REFLECTION MEASUREMENTS ON BLOOD BY FIBER-OPTIC CATHETER
    MOOK, GA
    OSYPKA, P
    STURM, RE
    WOOD, EH
    ACTA PHYSIOLOGICA ET PHARMACOLOGICA NEERLANDICA, 1966, 14 (01): : 67 - &
  • [30] Fiber-optic infrared radiometer for accurate temperature measurements
    Sade, Sharon
    Eyal, Ophir
    Scharf, Vered
    Katzir, Abraham
    Applied Optics, 2002, 41 (10): : 1908 - 1914