Refinement of MLC modeling improves commercial QA dosimetry system for SRS and SBRT patient-specific QA

被引:21
作者
Hillman, Yair [1 ]
Kim, Josh [2 ]
Chetty, Indrin [2 ]
Wen, Ning [2 ]
机构
[1] Karmanos Canc Inst McLaren Macomb, Dept Radiat Oncol, Mt Clemens, MI USA
[2] Henry Ford Hosp, Dept Radiat Oncol, Detroit, MI 48202 USA
关键词
Patient-specific QA; quality assurance; small fields; SRS and SBRT; stereotactic radiosurgery; QUALITY-ASSURANCE; SCINTILLATOR DETECTOR; MONTE-CARLO; LOG FILES; RADIOTHERAPY; ERRORS; IMRT; FIELDS;
D O I
10.1002/mp.12808
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeMobius 3D (M3D) provides a volumetric dose verification of the treatment planning system's calculated dose using an independent beam model and a collapsed cone convolution superposition algorithm. However, there is a lack of investigation into M3D's accuracy and effectiveness for stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT) quality assurance (QA). Here, we collaborated with the vendor to develop a revised M3D beam model for SRS/SBRT cases treated with a 6X flattening filter-free (FFF) beam and high-definition multiple leaf collimator (HDMLC) on an Edge linear accelerator. MethodsEighty SRS/SBRT cases, planned with AAA dose algorithm and validated with Gafchromic film, were compared to M3D dose calculations using 3D gamma analysis with 2%/2 mm gamma criteria and a 10% threshold. A revised beam model was developed by refining the HD-MLC model in M3D to improve small field dose calculation accuracy and beam profile agreement. All cases were reanalyzed using the revised beam model. The impact of heterogeneity corrections for lung cases was investigated by applying lung density overrides to five cases. ResultsFor the standard and revised beam models, respectively, the mean gamma passing rates were 94.6% [standard deviation (SD): 6.1%] and 98.0% [SD: 1.7%] (for the overall patient), 88.2% [SD: 17.3%] and 93.8% [SD: 6.8%] (for the brain PTV), 71.4% [SD: 18.4%] and 81.5% [SD: 14.3%] (for the lung PTV), 83.3% [SD: 16.7%] and 67.9% [SD: 23.0%] (for the spine PTV), and 78.6% [SD: 14.0%] and 86.8% [SD: 12.5%] (for the PTV of all other sites). The lung PTV mean gamma passing rates improved from 74.1% [SD: 7.5%] to 89.3% [SD: 7.2%] with the lung density overridden. The revised beam model achieved an output factor within 3% of plastic scintillator measurements for 2 x 2 cm(2) MLC field size, but larger discrepancies are still seen for smaller field sizes which necessitate further improvement of the beam model. ConclusionSpecial attention needs to be paid to small field dosimetry, MLC modeling, and inhomogeneity corrections in the beam model for SRS/SBRT QA. The improvements noted in this study, and further collaborations between clinical physicists and the vendor to refine the M3D beam model could enable M3D to become a premier SRS/SBRT QA tool.
引用
收藏
页码:1351 / 1359
页数:9
相关论文
共 27 条
  • [1] Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries
    Agnew, A.
    Agnew, C. E.
    Grattan, M. W. D.
    Hounsell, A. R.
    McGarry, C. K.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09) : N49 - N63
  • [2] Childress N., 2012, Mobius3D White Paper: Dose Calculation Algorithm
  • [3] Parallel/Opposed: IMRT QA using treatment log files is superior to conventional measurement-based method
    Childress, Nathan
    Chen, Quan
    Rong, Yi
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01): : 4 - 7
  • [4] PTV dose prescription strategies for SBRT of metastatic liver tumours
    de Pooter, Jacco A.
    Wunderink, Wouter
    Romero, Alejandra Mendez
    Storchi, Pascal R. M.
    Heijmen, Ben J. M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 85 (02) : 260 - 266
  • [5] IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119
    Ezzell, Gary A.
    Burmeister, Jay W.
    Dogan, Nesrin
    LoSasso, Thomas J.
    Mechalakos, James G.
    Mihailidis, Dimitris
    Molineu, Andrea
    Palta, Jatinder R.
    Ramsey, Chester R.
    Salter, Bill J.
    Shi, Jie
    Xia, Ping
    Yue, Ning J.
    Xiao, Ying
    [J]. MEDICAL PHYSICS, 2009, 36 (11) : 5359 - 5373
  • [6] American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning
    Fraass, B
    Doppke, K
    Hunt, M
    Kutcher, G
    Starkschall, G
    Stern, R
    Van Dyke, J
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 1773 - 1829
  • [7] Variation of kQclin, Qmsrfclin, fmsr for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio
    Francescon, Paolo
    Beddar, Sam
    Satariano, Ninfa
    Das, Indra J.
    [J]. MEDICAL PHYSICS, 2014, 41 (10)
  • [8] AAPM-RSS Medical Physics Practice Guideline 9.a. for SRS-SBRT
    Halvorsen, Per H.
    Cirino, Eileen
    Das, Indra J.
    Garrett, Jeffrey A.
    Yang, Jun
    Yin, Fang-Fang
    Fairobent, Lynne A.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05): : 10 - 21
  • [9] Dosimetric comparison of Acuros XB deterministic radiation transport method with Monte Carlo and model-based convolution methods in heterogeneous media
    Han, Tao
    Mikell, Justin K.
    Salehpour, Mohammad
    Mourtada, Firas
    [J]. MEDICAL PHYSICS, 2011, 38 (05) : 2651 - 2664
  • [10] Stereotactic Ablative Radiotherapy for Non-Small Cell Lung Cancer: Rationale and Outcomes
    Iyengar, Puneeth
    Timmerman, Robert D.
    [J]. JOURNAL OF THE NATIONAL COMPREHENSIVE CANCER NETWORK, 2012, 10 (12): : 1514 - 1520