Fast Accurate MR Thermometry Using Phase Referenced Asymmetric Spin-Echo EPI at High Field

被引:15
作者
Streicher, Markus N. [1 ]
Schaefer, Andreas [1 ]
Ivanov, Dimo [1 ]
Mueller, Dirk K. [1 ]
Amadon, Alexis [1 ]
Reimer, Enrico [1 ]
Huber, Laurentius [1 ]
Dhital, Bibek [1 ]
Rivera, Debra [1 ]
Koegler, Carsten [1 ]
Trampel, Robert [1 ]
Pampel, Andre [1 ]
Turner, Robert [1 ]
机构
[1] Max Planck Inst Human Cognit & Brain Sci, D-04103 Leipzig, Germany
关键词
referenced; MR thermometry; fat; spin-echo; asymmetric; proton resonance frequency shift; PRF SHIFT THERMOMETRY; CHEMICAL-SHIFT; MAGNETIC-FIELD; STEADY-STATE; TEMPERATURE; SUSCEPTIBILITY; NMR; FAT; RESPIRATION; TISSUE;
D O I
10.1002/mrm.24681
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeA novel highly accurate method for MR thermometry, effective at high field, is introduced and validated, which corrects for slow and fast field fluctuations by means of reference images. MethodsAn asymmetric spin-echo echo planar imaging sequence was made frequency-selective to water or a reference substance by controlling the slice-select gradient polarity and the duration of the excitation and refocusing radiofrequency pulses. Images were acquired pairwise, and the temperature-sensitive water images were corrected for field fluctuations using the reference images. In a phantom radiofrequency heating experiment, dissolved dimethyl sulfoxide was used as a reference substance. Temperature stability was tested in vivo on the human brain, referenced using subcutaneous scalp fat. Water and fat phase images were acquired only 50 ms apart. Bloch simulations validated the frequency selection accuracy. ResultsAsymmetric spin-echo imaging using a simple frequency selection method provides highly accurate referenced MR thermometry in phantoms and in vivo at 7 T. Effects of field fluctuations caused by field drift, breathing, and heart beat were corrected. The technique is highly robust against B-1 inhomogeneities. ConclusionFrequency selection using gradient-reversal can enable fast accurate referenced in vivo MR thermometry, assisting thermal characterization of radiofrequency coils and possibly in vivo SAR monitoring. Magn Reson Med 71:524-533, 2014. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:524 / 533
页数:10
相关论文
共 50 条
  • [31] 2-V
  • [32] A model for susceptibility artefacts from respiration in functional echo-planar magnetic resonance imaging
    Raj, D
    Paley, DP
    Anderson, AW
    Kennan, RP
    Gore, JC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12) : 3809 - 3820
  • [33] Composition of adipose tissue and marrow fat in humans by 1H NMR at 7 Tesla
    Ren, Jimin
    Dimitrov, Ivan
    Sherry, A. Dean
    Malloy, Craig R.
    [J]. JOURNAL OF LIPID RESEARCH, 2008, 49 (09) : 2055 - 2062
  • [34] Referenceless PRF shift thermometry
    Rieke, V
    Vigen, KK
    Sommer, G
    Daniel, BL
    Pauly, JM
    Butts, K
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (06) : 1223 - 1231
  • [35] Echo combination to reduce proton resonance frequency (PRF) thermometry errors from fat
    Rieke, Viola
    Pauly, Kim Butts
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (03) : 673 - 677
  • [36] MR thermometry
    Rieke, Viola
    Pauly, Kim Butts
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (02) : 376 - 390
  • [37] Combining Phase Images From Multi-Channel RF Coils Using 3D Phase Offset Maps Derived From a Dual-Echo Scan
    Robinson, Simon
    Grabner, Guenther
    Witoszynskyj, Stephan
    Trattnig, Siegfried
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (06) : 1638 - 1648
  • [38] Reference-Free PRFS MR-Thermometry Using Near-Harmonic 2-D Reconstruction of the Background Phase
    Salomir, Rares
    Viallon, Magalie
    Kickhefel, Antje
    Roland, Joerg
    Morel, Denis R.
    Petrusca, Lorena
    Auboiroux, Vincent
    Goget, Thomas
    Terraz, Sylvain
    Becker, Christoph D.
    Gross, Patrick
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (02) : 287 - 301
  • [39] Patient safety concept for multichannel transmit coils
    Seifert, Frank
    Wuebbeler, Gerd
    Junge, Sven
    Ittermann, Bernd
    Rinneberg, Herbert
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 26 (05) : 1315 - 1321
  • [40] Correction of proton resonance frequency shift temperature maps for magnetic field disturbances using fat signal
    Shmatukha, Andriy V.
    Harvey, Paul R.
    Bakker, Chris J. G.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (03) : 579 - 587