A large, high performance, curved 2D position-sensitive neutron detector

被引:23
作者
Fried, J [1 ]
Harder, JA [1 ]
Mahler, GJ [1 ]
Makowiecki, DS [1 ]
Mead, JA [1 ]
Radeka, V [1 ]
Schaknowski, NA [1 ]
Smith, GC [1 ]
Yu, B [1 ]
机构
[1] Brookhaven Natl Lab, Instrumentat Div, Upton, NY 11973 USA
关键词
neutron detector; position sensitive; interpolating readout; protein crystallography; MWPC;
D O I
10.1016/S0168-9002(01)01787-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new position-sensitive neutron detector has been designed and constructed for a protein crystallography station at LANL's pulsed neutron source. This station will be one of the most advanced instruments at a major neutron user facility for protein crystallography, fiber and membrane diffraction. The detector, based on neutron absorption in (3)He, has a large sensitive area of 3000 cm 2, angular coverage of 120degrees, timing resolution of 1 mus, rate capability in excess of 10(6) s(-1), position resolution of about 1.5 mm FWHM, and efficiency > 50% for neutrons of interest in the range 1-10 Angstrom. Features that are key to these remarkable specifications are the utilization of eight independently operating segments within a single gas volume, fabrication of the detector vessel and internal segments with a radius of curvature of about 70 cm, optimized position readout based on charge division and signal shaping with gated baseline restoration, and engineering design with high-strength aluminum alloy. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:415 / 419
页数:5
相关论文
共 6 条
[1]   TWO-DIMENSIONAL HIGH-PRECISION THERMAL-NEUTRON DETECTOR [J].
BOIE, RA ;
FISCHER, J ;
INAGAKI, Y ;
MERRITT, FC ;
OKUNO, H ;
RADEKA, V .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 200 (2-3) :533-545
[2]  
Radeka V, 1996, BASIC LIFE SCI, V64, P57
[3]   CENTROID FINDING METHOD FOR POSITION-SENSITIVE DETECTORS [J].
RADEKA, V ;
BOIE, RA .
NUCLEAR INSTRUMENTS & METHODS, 1980, 178 (2-3) :543-554
[4]   The LANL Neutron-Science-Center Time-Of-Flight/Position-Sensitive-Detect module: Status and progress [J].
Rose, CR ;
Hammonds, JP ;
Nelson, RA ;
Weizeorick, JT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (02) :151-153
[5]   THE LINEARITY PERFORMANCE OF A 2-DIMENSIONAL, X-RAY PROPORTIONAL CHAMBER WITH 0.58MM ANODE WIRE SPACING [J].
SMITH, GC ;
YU, B .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :541-547
[6]  
ZOJCESKI Z, 2002, IN PRESS NUCL INSTR