A Tensor/Balancer device (Stryker Howmedica Osteonics, Allendale, NJ) was used to restore optimal stability and alignment during 83 consecutive total knee arthroplasties with a minimum of 6-week clinical and radiographic follow-up. The surgical technique is described. Mean flexion-extension symmetry was restored to within 1degrees. Mean femoral rotation required to achieve symmetry in flexion differed between varus (4.38degrees) and valgus (6.0degrees) knees. Of 83 knees, 7 had >3 mm of subjective laxity during component trialing and were associated with a preoperative deformity of >15degrees; 19 knees required femoral rotation of >6degrees and were associated with greater preoperative malalignment. Postoperative knee alignment, range of motion, Knee Society score, and lateral release rate were similar between the 2 groups. The use of a Tensor/Balancer device in total knee arthroplasty can achieve outcomes comparable to existing techniques with potential for improved technical accuracy.
引用
收藏
页码:2 / 10
页数:9
相关论文
共 25 条
[1]  
ANOUCHI YS, 1993, CLIN ORTHOP RELAT R, V287, P170
[2]   FEMORAL ROTATIONAL ALIGNMENT, BASED ON THE ANTEROPOSTERIOR AXIS, IN TOTAL KNEE ARTHROPLASTY IN A VALGUS KNEE - A TECHNICAL NOTE [J].
ARIMA, J ;
WHITESIDE, LA ;
MCCARTHY, DS ;
WHITE, SE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77 (09) :1331-1334
[3]  
BERGER RA, 1993, CLIN ORTHOP RELAT R, V286, P40
[4]  
DORR LD, 1986, CLIN ORTHOP RELAT R, V205, P5
[5]  
Eckhoff D G, 1994, J Arthroplasty, V9, P73, DOI 10.1016/0883-5403(94)90140-6
[6]   THE INFLUENCE OF TIBIAL-PATELLOFEMORAL LOCATION ON FUNCTION OF THE KNEE IN PATIENTS WITH THE POSTERIOR STABILIZED CONDYLAR KNEE PROSTHESIS [J].
FIGGIE, HE ;
GOLDBERG, VM ;
HEIPLE, KG ;
MOLLER, HS ;
GORDON, NH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1035-1040
[7]   ICLH ARTHROPLASTY OF KNEE - 1968-1977 [J].
FREEMAN, MAR ;
TODD, RC ;
BAMERT, P ;
DAY, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :339-344
[8]   Accuracy of soft tissue balancing in total knee arthroplasty [J].
Griffin, FM ;
Insall, JN ;
Scuderi, GR .
JOURNAL OF ARTHROPLASTY, 2000, 15 (08) :970-973
[9]  
INSALL JN, 1985, CLIN ORTHOP RELAT R, V192, P13
[10]   SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY [J].
KARACHALIOS, T ;
SARANGI, PP ;
NEWMAN, JH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (06) :938-942