Capillary isoelectric focusing of hemoglobin variants in the pediatric clinical laboratory

被引:43
作者
Hempe, JM
Granger, JN
Craver, RD
机构
[1] LOUISIANA STATE UNIV,DEPT PATHOL,SCH MED,NEW ORLEANS,LA 70112
[2] CHILDRENS HOSP,NEW ORLEANS,LA
关键词
hemoglobin; capillary electrophoresis; capillary isoelectric focusing; analytical clinical chemistry; pediatrics;
D O I
10.1002/elps.1150181013
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have used capillary isoelectric focusing (cIEF) to diagnose and monitor hemoglobinopathies in children for over three years. This report describes a major revision of our original method (Clin. Chem. 1994, 40, 2288-2295) that improves the analysis of hemoglobin (Hb) Variants by cIEF, especially capillary performance and quantitation precision for minor variants. The revised method uses mixed ampholytes (2% pH 6-8:3-10; 10:1), lower viscosity methylcellulose (0.375%) solution, between-sample capillary conditioning with methanol, and hemolysates prepared from red blood cells (RBC) instead of whole blood. Collectively, these changes prolonged capillary life by minimizing capillary exposure to NaOH, standardized the sample matrix, and improved the precision of peak autointegration. The between-run quantitation imprecision (% relative standard deviation) of the revised method was 0.1-3.5% for all diagnostically important major and minor Hb variants present at normal or abnormal levels. The results show the use of the revised method for (i) posttranslationally modified Hb present at low concentrations in normal blood, (ii) Hb oxidation products produced by improper sample storage, (iii) differential diagnosis of S/beta(+) thalassemia, G-Philadelphia trait, S/C-Harlem disease, and Hb H disease, (iv) sensitive detection of minor variants like Hb A(2)' as indicators of an alpha globin mutation, and (v) neonatal screening using dried blood collected on filter paper. The results show that high-efficiency separation and precise quantitation of Hb variants over a wide range of concentrations makes cIEF a comprehensive assay that can be used without adjunct analyses for the automated primary evaluation of hemoglobinopathies and thalassemias.
引用
收藏
页码:1785 / 1795
页数:11
相关论文
共 19 条
  • [1] BASSET P, 1978, BLOOD, V51, P971
  • [2] BISSE E, 1995, CLIN CHEM, V41, P658
  • [4] Craver RD, 1997, AM J CLIN PATHOL, V107, P88
  • [5] Fairbanks VF, 1980, HEMOGLOBINOPATHIES T
  • [6] HEMPE JM, 1994, CLIN CHEM, V40, P2288
  • [7] DETECTION OF ABNORMAL HEMOGLOBIN BY CAPILLARY ELECTROPHORESIS AND STRUCTURAL IDENTIFICATION
    ISHIOKA, N
    IYORI, N
    NOJI, J
    KURIOKA, S
    [J]. BIOMEDICAL CHROMATOGRAPHY, 1992, 6 (05) : 224 - 226
  • [8] KOEPKE JA, 1975, CLIN CHEM, V21, P1953
  • [9] LANDERS JP, 1995, CLIN CHEM, V41, P495
  • [10] Lehmann R, 1996, J CAPILLARY ELECTROP, V3, P89