Effects of Different Implant Osteotomy Preparation Sizes on Implant Stability and Bone Response in the Minipig Mandible

被引:27
作者
Eom, Tae-Gwan [1 ,2 ]
Kim, Hyo-won [1 ]
Jeon, Gyeo-Rok [2 ]
Yun, Mi-Jung [3 ]
Huh, Jung-Bo [3 ]
Jeong, Chang-Mo [3 ]
机构
[1] OSSTEM IMPLANT Co Ltd, Implant R&D Ctr, Busan, South Korea
[2] Pusan Natl Univ, Sch Med, Dept Biomed Engn, Busan, South Korea
[3] Pusan Natl Univ, Sch Dent, Dent Res Inst, Dept Prosthodont, Busan, South Korea
关键词
bone response; dental implant; implant stability; osteotomy preparation size; RESONANCE FREQUENCY MEASUREMENTS; DENTAL IMPLANTS; TITANIUM IMPLANTS; INSERTION TORQUE; TISSUE-REACTIONS; HEAT-GENERATION; MINIATURE PIGS; IN-VIVO; PLACEMENT; MAXILLA;
D O I
10.11607/jomi.4165
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study was to investigate the influence of implant and drill diameters on the stability of implant and bone response. Materials and Methods: An implant (GS II, Osstem Implant) with a 3.5-mm diameter and drills with three different diameters, differentiating the volume of bone compacted by the implant, were used in this study. Measurement of the insertion torque and observation of bone compression patterns were done during implant site preparation in minipigs. Also, analysis of resonance frequency, histomorphology, marginal bone resorption, and new bone formation were performed. Results: Microstrains and microcracks occurred in cortical bone around the thread when a drill with a smaller diameter than that of the thread was used. Higher implant stability was shown and maintained when a small-diameter drill was used. When using a drill with a smaller diameter than the root diameter of the implant thread, the bone-to-implant contact/bone area (BIC/BA) was higher than the wider drill during the whole test period. However, the use of a wide drill was the most rapid in increasing the BIC/BA value by new bone formation. In the cancellous bone, the speed of new bone formation was not different in each test group. However, the formation of new bone was faster when a large-diameter drill was used in the cortical bone. There was no significant difference in marginal bone loss (MBL) according to drill diameters, but the speed of MBL was fast when using a drill that was smaller than the root diameter of the implant thread. Conclusion: Implant stability, BIC, and BA were high when using a final drill that was smaller than the root diameter of the implant thread compared with a wide final drill. However, the speed of new bone formation was relatively slow and the speed of MBL was relatively fast if the final drill was excessively smaller.
引用
收藏
页码:997 / 1006
页数:10
相关论文
共 50 条
[31]   Implant stability and bone density: assessment of correlation in fresh cadavers using conventional and osteotome implant sockets [J].
Cehreli, Murat Cavit ;
Kokat, Ali Murat ;
Comert, Ayhan ;
Akkocaoglu, Murat ;
Tekdemir, Ibrahim ;
Akca, Kwanc .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (10) :1163-1169
[32]   Influence of Mucosal Thickness, Implant Dimensions and Stability in Cone Morse Implant Installed at Subcrestal Bone Level on the Peri-Implant Bone: A Prospective Clinical and Radiographic Study [J].
Gehrke, Sergio Alexandre ;
Bercianos, Mauro ;
Aguerrondo, Jorge Gonzalo ;
Calvo-Guirado, Jose Luis ;
Prados-Frutos, Juan Carlos .
SYMMETRY-BASEL, 2019, 11 (09)
[33]   Effect of bone quality, implant design, and surgical technique on primary implant stability [J].
Karl, Matthias ;
Grobecker-Karl, Tanja .
QUINTESSENCE INTERNATIONAL, 2018, 49 (03) :189-198
[34]   Implant Stability in Regenerated Bone [J].
Caggiano, Mario ;
Acerra, Alfonso .
APPLIED SCIENCES-BASEL, 2023, 13 (22)
[35]   Histologic and Biomechanical Evaluation of the Effects of Implant Insertion Torque on Peri-Implant Bone Healing [J].
Consolo, Ugo ;
Travaglini, Domenico ;
Todisco, Marzio ;
Trisi, Paolo ;
Galli, Silvia .
JOURNAL OF CRANIOFACIAL SURGERY, 2013, 24 (03) :860-865
[36]   Modal analysis for implant stability assessment: Sensitivity of this methodology for different implant designs [J].
Zanetti, Elisabetta Maria ;
Ciaramella, Stefano ;
Cali, Michele ;
Pascoletti, Giulia ;
Martorelli, Massimo ;
Asero, Riccardo ;
Watts, David C. .
DENTAL MATERIALS, 2018, 34 (08) :1235-1245
[37]   Effect of Implant Design on Preservation of Marginal Bone in the Mandible [J].
Van de Velde, Tommie ;
Collaert, Bruno ;
Sennerby, Lars ;
De Bruyn, Hugo .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2010, 12 (02) :134-141
[38]   Bone Response to Loaded Implants With Non-Matching Implant-Abutment Diameters in the Canine Mandible [J].
Cochran, David L. ;
Bosshardt, Dieter D. ;
Grize, Leticia ;
Higginbottom, Frank L. ;
Jones, Archie A. ;
Jung, Ronald E. ;
Wieland, Marco ;
Dard, Michel .
JOURNAL OF PERIODONTOLOGY, 2009, 80 (04) :609-617
[39]   Relation between initial implant stability quotient and bone-implant contact percentage: an in vitro model study [J].
Huang, Heng-Li ;
Tsai, Ming-Tzu ;
Su, Kuo-Chih ;
Li, Yu-Fen ;
Hsu, Jui-Ting ;
Chang, Chih-Han ;
Fuh, Lih-Jyh ;
Wu, Aaron Yu-Jen .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2013, 116 (05) :E356-E361
[40]   Implant stability during osseointegration in irradiated and non-irradiated minipig alveolar bone: an experimental study [J].
Verdonck, Henk W. D. ;
Meijer, Gert J. ;
Laurin, Thorsten ;
Nieman, Fred H. ;
Stoll, Christian ;
Riediger, Dieter ;
Stoelinga, Paul J. W. ;
de Baat, Cees .
CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (02) :201-206