Effects of Different Implant Osteotomy Preparation Sizes on Implant Stability and Bone Response in the Minipig Mandible

被引:29
作者
Eom, Tae-Gwan [1 ,2 ]
Kim, Hyo-won [1 ]
Jeon, Gyeo-Rok [2 ]
Yun, Mi-Jung [3 ]
Huh, Jung-Bo [3 ]
Jeong, Chang-Mo [3 ]
机构
[1] OSSTEM IMPLANT Co Ltd, Implant R&D Ctr, Busan, South Korea
[2] Pusan Natl Univ, Sch Med, Dept Biomed Engn, Busan, South Korea
[3] Pusan Natl Univ, Sch Dent, Dent Res Inst, Dept Prosthodont, Busan, South Korea
关键词
bone response; dental implant; implant stability; osteotomy preparation size; RESONANCE FREQUENCY MEASUREMENTS; DENTAL IMPLANTS; TITANIUM IMPLANTS; INSERTION TORQUE; TISSUE-REACTIONS; HEAT-GENERATION; MINIATURE PIGS; IN-VIVO; PLACEMENT; MAXILLA;
D O I
10.11607/jomi.4165
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study was to investigate the influence of implant and drill diameters on the stability of implant and bone response. Materials and Methods: An implant (GS II, Osstem Implant) with a 3.5-mm diameter and drills with three different diameters, differentiating the volume of bone compacted by the implant, were used in this study. Measurement of the insertion torque and observation of bone compression patterns were done during implant site preparation in minipigs. Also, analysis of resonance frequency, histomorphology, marginal bone resorption, and new bone formation were performed. Results: Microstrains and microcracks occurred in cortical bone around the thread when a drill with a smaller diameter than that of the thread was used. Higher implant stability was shown and maintained when a small-diameter drill was used. When using a drill with a smaller diameter than the root diameter of the implant thread, the bone-to-implant contact/bone area (BIC/BA) was higher than the wider drill during the whole test period. However, the use of a wide drill was the most rapid in increasing the BIC/BA value by new bone formation. In the cancellous bone, the speed of new bone formation was not different in each test group. However, the formation of new bone was faster when a large-diameter drill was used in the cortical bone. There was no significant difference in marginal bone loss (MBL) according to drill diameters, but the speed of MBL was fast when using a drill that was smaller than the root diameter of the implant thread. Conclusion: Implant stability, BIC, and BA were high when using a final drill that was smaller than the root diameter of the implant thread compared with a wide final drill. However, the speed of new bone formation was relatively slow and the speed of MBL was relatively fast if the final drill was excessively smaller.
引用
收藏
页码:997 / 1006
页数:10
相关论文
共 50 条
[21]   The Effect of Coronal Implant Design and Drilling Protocol on Bone-to-Implant Contact: A 3-Month Study in the Minipig Calvarium [J].
Cohen, Omer ;
Bosshardt, Dieter D. ;
Weinberg, Evegeny ;
Slutzkey, Gil ;
Moses, Ofer .
MATERIALS, 2021, 14 (10)
[22]   Comparison of implant stability after different implant surface treatments in dog bone [J].
Kim, Sun-Jong ;
Kim, Myung-Rae ;
Rim, Jae-Suk ;
Chung, Sung-Min ;
Shin, Sang-Wan .
JOURNAL OF APPLIED ORAL SCIENCE, 2010, 18 (04) :415-420
[23]   Prospective Clinical Trial Evaluating a New Implant System for Implant Survival, Implant Stability and Radiographic Bone Changes [J].
Becker, William ;
Becker, Burton E. ;
Hujoel, Philippe ;
Abu Ras, Zakaria ;
Goldstein, Moshe ;
Smidt, Ami .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2013, 15 (01) :15-21
[24]   Mechanisms of reduced implant stability in osteoporotic bone [J].
Ruffoni, D. ;
Mueller, R. ;
van Lenthe, G. H. .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2012, 11 (3-4) :313-323
[25]   Primary Stability of Three Different Osteotomy Techniques in Medullary Bone: An in Vitro Study [J].
Attanasio, Ferdinando ;
Antonelli, Alessandro ;
Brancaccio, Ylenia ;
Averta, Fiorella ;
Figliuzzi, Michele Mario ;
Fortunato, Leonzio ;
Giudice, Amerigo .
DENTISTRY JOURNAL, 2020, 8 (01)
[26]   Undersized Implant Site Preparation to Enhance Primary Implant Stability in Poor Bone Density: A Prospective Clinical Study [J].
Alghamdi, Hamdan ;
Anand, Pradeep S. ;
Anil, Sukumaran .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 69 (12) :E506-E512
[27]   Mineralized Bone-Implant Contact and Implant Stability Quotient in 16 Human Implants Retrieved after Early Healing Periods: A Histologic and Histomorphometric Evaluation [J].
Degidi, Marco ;
Perrotti, Vittoria ;
Piattelli, Adriano ;
Iezzi, Giovanna .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2010, 25 (01) :45-48
[28]   Correlation of Bone Vascularity in the Posterior Mandible and Subsequent Implant Stability: A Preliminary Study [J].
Kokovic, Vladimir ;
Krsljak, Elena ;
Andric, Miroslav ;
Brkovic, Bozidar ;
Milicic, Biljana ;
Jurisic, Milan ;
Rahman, Muhammed Mustahsen ;
Haemmerle, Christoph H. F. .
IMPLANT DENTISTRY, 2014, 23 (02) :200-205
[29]   The Effect of Melatonin on Implant Stability and Marginal Bone Level around The Implant [J].
Ravella, Hemanth ;
Rao, Mulpuri Venkata Ramoji ;
Manthena, Sathish ;
Koneru, Suneetha ;
Preethi, Penubolu Lakshmi ;
Anumolu, Venkata Naga Sri Harsha .
JOURNAL OF DENTISTRY INDONESIA, 2022, 29 (02) :105-112
[30]   Effects of Implant Morphology on Rotational Stability During Immediate Implant Placement in the Esthetic Zone [J].
Kan, Joseph Y. K. ;
Roe, Phillip ;
Rungcharassaeng, Kitichai .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (03) :667-670