Development of a Multi-modality Compatible Flow Loop System for the Functional Assessment of Mitral Valve Prostheses

被引:8
作者
Jackson, Matthew S. [1 ]
Igo, Stephen R. [1 ]
Lindsey, Thomas E. [2 ]
Maragiannis, Dimitrios [1 ]
Chin, Karen E. [1 ]
Autry, Kyle [1 ]
Schutt, Robert, III [1 ]
Shah, Dipan J. [1 ]
Valsecchi, Pietro [2 ]
Kline, William B. [2 ]
Little, Stephen H. [1 ]
机构
[1] Houston Methodist DeBakey Heart & Vasc Ctr, Houston, TX 77030 USA
[2] ExxonMobil Upstream Res Co, Houston, TX USA
基金
美国国家科学基金会;
关键词
Prosthetic heart valves; Cardiac MRI; PC-MRI; Echocardiography; Flow loop;
D O I
10.1007/s13239-014-0177-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We sought to build a robust pulsatile flow model permitting comparison of 3D Color Doppler and phase contrast magnetic resonance (PCMR) flow and orifice area measurements of prosthetic valve function that would provide an in vitro gold standard for evaluation of the accuracy of new imaging techniques. A multi-modality compatible, pulsatile linear actuation system, control software, and mock circulation loop was designed and created using non-ferromagnetic components, NI Labview, and rapid prototyping. The developed system's ability to create pulsatile, controlled flow was assessed by examining peak diastolic flow, diastolic period, and diastolic flow volume (DFV) of 12 consecutive cardiac cycles at 3 different DFVs. Reproducibility of flow was tested by examining the systems ability to generate the same target DFV after draining and re-configuring the flow loop components. Furthermore, the system was evaluated by CMR and echocardiography to examine image quality, DFV quantification, and an initial functional assessment of prosthetic mitral valve (MV) orifice area by four different imaging methods. The system was able to reproducibly create prescribed flow volumes across 31 mm bioprosthetic and mechanical MV's under a range of flow conditions. In vitro image quality for the identification of MV sewing ring, struts and leaflets was similar to clinical imaging quality for both echocardiographic and MRI methods. For the quantification of DFV, the PCMR method demonstrated an error range of 1.4-9.2% and a mean difference of 4 +/- 2 mL/beat compared to the flowmeter reference standard. Across the three flow conditions the bioprosthetic MV diastolic flow area was statistically similar by MRI planimetry, phase contrast MRI planimetry, and 3D color Doppler planimetry methods (mean area 3.2 +/- 0.5 cm(,)(2) ANOVA, p = NS), but the effective orifice area by 2D spectral Doppler (continuity) method was smaller (mean area 2.0 +/- 0.3 cm(2), ANOVA, p< 0.001). We describe the development and initial functional testing of a novel MRI-compatible flow loop system for comparison of multimodality imaging techniques to assess prosthetic valve function.
引用
收藏
页码:13 / 24
页数:12
相关论文
共 22 条
[1]   The potential of myocardial perfusion scintigraphy for risk stratification of asymptomatic patients with type 2 diabetes [J].
Bax, Jeroen J. ;
Bonow, Robert O. ;
Tschoepe, Diethelm ;
Inzucchi, Silvio E. ;
Barrett, Eugene .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (04) :754-760
[2]   Comprehensive Echocardiographic Assessment of Normal Mitral Medtronic Hancock II, Medtronic Mosaic, and Carpentier-Edwards Perimount Bioprostheses Early after Implantation [J].
Blauwet, Lori A. ;
Malouf, Joseph F. ;
Connolly, Heidi M. ;
Hodge, David O. ;
Evans, Katie N. ;
Herges, Regina M. ;
Sundt, Thoralf M., III ;
Miller, Fletcher A., Jr. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2010, 23 (06) :656-666
[3]  
GARVER D, 1995, TEX HEART I J, V22, P86
[4]   Prosthetic heart valve evaluation by magnetic resonance imaging [J].
Hasenkam, JM ;
Ringgaard, S ;
Houlind, K ;
Botnar, RM ;
Stodkilde-Jorgensen, H ;
Boesiger, P ;
Pedersen, EM .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1999, 16 (03) :300-305
[5]   ACCF/ACR/SCCT/SCMR/ASNC/NASCI/SCAI/SIR 2006 appropriateness criteria for cardiac computed tomography and cardiac magnetic resonance imaging - A report of the American College of Cardiology Foundation Quality Strategic Directions Committee Appropriateness Criteria Working Group, American College of Radiology, Society of Cardiovascular Computed Tomography, Society for Cardiovascular Magnetic Resonance, American Society of Nuclear Cardiology, North American Society for Cardiac Imaging, Society for Cardiovascular Angiography and Interventions, and Society of Interventional Radiology [J].
Hendel, Robert C. ;
Patel, Manesh R. ;
Kramer, Christopher M. ;
Poon, Michael ;
Hendel, Robert C. ;
Brindis, Ralph G. ;
Hendel, Robert C. ;
Douglas, Pamela S. ;
Peterson, Eric D. ;
Wolk, Michael J. ;
Allen, Joseph M. ;
Patel, Manesh R. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (07) :1475-1497
[6]   Direct Assessment of Normal Mechanical Mitral Valve Orifice Area by Real-Time 3D Echocardiography [J].
Krim, Selim R. ;
Vivo, Rey P. ;
Patel, Ankit ;
Xu, Jiaqiong ;
Igo, Stephen R. ;
Zoghbi, William A. ;
Little, Stephen H. .
JACC-CARDIOVASCULAR IMAGING, 2012, 5 (05) :478-483
[7]   In Vitro Assessment of Flow Patterns and Turbulence Intensity in Prosthetic Heart Valves Using Generalized Phase-Contrast MRI [J].
Kvitting, John-Peder Escobar ;
Dyverfeldt, Petter ;
Sigfridsson, Andreas ;
Franzen, Stefan ;
Wigstrom, Lars ;
Bolger, Ann F. ;
Ebbers, Tino .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (05) :1075-1080
[8]   Turbulence Characteristics Downstream of a New Trileaflet Mechanical Heart Valve [J].
Li, Chi-Pei ;
Chen, Sheng-Fu ;
Lo, Chi-Wen ;
Lu, Po-Chien .
ASAIO JOURNAL, 2011, 57 (03) :188-196
[9]   In vitro validation of real-time three-dimensional color Doppler echocardiography for direct measurement of proximal isovelocity surface area in mitral regurgitation [J].
Little, Stephen H. ;
Igo, Stephen R. ;
Pirat, Bahar ;
McCulloch, Marti ;
Hartley, Craig J. ;
Nose, Yukihiko ;
Zoghbi, William A. .
AMERICAN JOURNAL OF CARDIOLOGY, 2007, 99 (10) :1440-1447
[10]   Three-Dimensional Color Doppler Echocardiography for Direct Measurement of Vena Contracta Area in Mitral Regurgitation In Vitro Validation and Clinical Experience [J].
Little, Stephen H. ;
Pirat, Bahar ;
Kumar, Rahul ;
Igo, Stephen R. ;
McCulloch, Marti ;
Hartley, Craig J. ;
Xu, Jiaqiong ;
Zoghbi, William A. .
JACC-CARDIOVASCULAR IMAGING, 2008, 1 (06) :695-704