Comparison of native myocardial T1 and T2 mapping at 1.5T and 3T in healthy volunteers Reference values and clinical implications

被引:57
|
作者
Granitz, Marcel [1 ]
Motloch, Lukas J. [2 ]
Granitz, Christina [2 ]
Meissnitzer, Matthias [1 ]
Hitzl, Wolfgang [3 ]
Hergan, Klaus [1 ]
Schlattau, Alexander [1 ]
机构
[1] Paracelsus Med Univ, Dept Radiol, Mullner Hauptstr 48, A-5020 Salzburg, Austria
[2] Paracelsus Med Univ, Dept Internal Med 2, Salzburg, Austria
[3] Paracelsus Med Univ, Res Off Biostat, Salzburg, Austria
关键词
Cardiac; MRI; Normal; Field strength; CARDIOVASCULAR MAGNETIC-RESONANCE; T-1; AGE; VALIDATION; FIBROSIS; DISEASE;
D O I
10.1007/s00508-018-1411-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Myocardial native T1 and T2 mapping are promising techniques for quantitative assessment of diffuse myocardial pathologies; however, due to conflicting data regarding normal values, routine clinical implementation of this method is still challenging. Methods To evaluate this situation during daily clinical practice the characteristics of normal values obtained in 60 healthy volunteers who underwent magnetic resonance imaging (MRI) scans on 1.5T and 3T scanners were studied. The T1 modified look-locker inversion recovery (MOLLI; 5(3)3; modified for higher heart rates) and T2 navigator gated black-blood prepared gradient-spin-echo (GraSE) sequences were used. Results While age and body mass index did not affect relaxation times, a gender and heart rate dependency was found showing higher T1 and T2 values in females, whereas at higher heart rates a prolongation of T1 and a shortening of T2 relaxation times was found. Particularly prone to artifacts were T2 measurements at 3T and the inferolateral wall. In the individual setting mean relaxation times for T1 were 995.8 +/- 30.9 ms at 1.5T and 1183.8 +/- 37.5 ms at 3T and 55.8 +/- 2.8 ms at 1.5T and 51.6 +/- 3 ms at 3T for T2 indicating a high dependency of reference values on MRI protocol when compared to the literature. Furthermore, as presumed mean T1 and T2 values correlated in the same individual. Conclusions The T1 and T2 relaxation times depend on physiological factors and especially on MRI protocols. Therefore, reference values should be validated individually in every radiological institution before implementing mapping protocols in daily clinical practice. Correlation of mean T1 and T2 values in the same proband at both field strengths indicates intraindividual reproducibility.
引用
收藏
页码:143 / 155
页数:13
相关论文
共 50 条
  • [31] Effect of Gd-DTPA-BMA on blood and myocardial T1 at 1.5T and 3T in humans
    Sharma, P
    Socolow, J
    Patel, S
    Pettigrew, RI
    Oshinski, JN
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 23 (03) : 323 - 330
  • [32] NON-CONTRAST ASSESSMENT OF MYOCARDIAL VIABILITY IN CHRONIC MYOCARDIAL INFARCTION BY NATIVE T1 AND T2 MAPPING AT 1.5T CMR: COMPARISON WITH LATE GADOLINIUM ENHANCEMENT TECHNIQUE
    Dastidar, Amardeep Ghosh
    Pontecorboli, Giulia
    Harries, Iwan
    Moret, Charlie
    Morgan, Gareth
    De Garate, Estefania
    Baritussio, Anna
    Bucciarelli-Ducci, Chiara
    HEART, 2017, 103 : A2 - A2
  • [33] Quantitative contrast ratio comparison between T1 (TSE at 1.5T, FLAIR at 3T), magnetization prepared rapid gradient echo and subtraction imaging at 1.5T and 3T
    Downs, Richard Keith
    Bashir, Mariah Haider
    Ng, Chin Kau
    Heidenreich, Jens Olaf
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2013, 3 (03) : 141 - 146
  • [34] Myocardial T1 values in healthy volunteers measured with saturation method using adaptive recovery times for T1 mapping (SMART1Map) at 1.5 T and 3 T
    Shunsuke Matsumoto
    Shigeo Okuda
    Yoshitake Yamada
    Tatsuya Suzuki
    Akihiro Tanimoto
    Atsushi Nozaki
    Masahiro Jinzaki
    Heart and Vessels, 2019, 34 : 1889 - 1894
  • [35] Myocardial T1 values in healthy volunteers measured with saturation method using adaptive recovery times for T1 mapping (SMART1Map) at 1.5 T and 3 T
    Matsumoto, Shunsuke
    Okuda, Shigeo
    Yamada, Yoshitake
    Suzuki, Tatsuya
    Tanimoto, Akihiro
    Nozaki, Atsushi
    Jinzaki, Masahiro
    HEART AND VESSELS, 2019, 34 (11) : 1889 - 1894
  • [36] The relationship between myocardial and hepatic T2 and T2*at 1.5T and 3T MRI in normal and iron-overloaded patients
    Kritsaneepaiboon, Supika
    Ina, Natee
    Chotsampancharoen, Thirachit
    Roymanee, Supaporn
    Cheewatanakornkul, Sirichai
    ACTA RADIOLOGICA, 2018, 59 (03) : 355 - 362
  • [37] Myocardial T1 Mapping at 3 T: Reference Values and Influencing Factors
    Zhao, Lei
    Li, Song-Nan
    Wang, Lu
    Lin, Hui
    Wen, Song-Nan
    Ning, Man
    Ma, Xiao-Hai
    Yu, Rong-Hui
    Long, De-Yong
    Tang, Ri-Bo
    Sang, Cai-Hua
    Jiang, Chen-Xi
    Du, Xin
    Dong, Jian-Zeng
    Fan, Zhan-Ming
    Ma, Chang-Sheng
    CIRCULATION, 2015, 132
  • [38] Myocardial T1 and T2 Mapping: Techniques and Clinical Applications
    Kim, Pan Ki
    Hong, Yoo Jin
    Im, Dong Jin
    Suh, Young Joo
    Park, Chul Hwan
    Kim, Jin Young
    Chang, Suyon
    Lee, Hye-Jeong
    Hur, Jin
    Kim, Young Jin
    Choi, Byoung Wook
    KOREAN JOURNAL OF RADIOLOGY, 2017, 18 (01) : 113 - 131
  • [39] Comparison of different cardiovascular magnetic resonance sequences for native myocardial T1 mapping at 3T
    Teixeira, Tiago
    Hafyane, Tarik
    Stikov, Nikola
    Akdeniz, Cansu
    Greiser, Andreas
    Friedrich, Matthias G.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2016, 18 : 1 - 12
  • [40] Comparison of different cardiovascular magnetic resonance sequences for native myocardial T1 mapping at 3T
    Tiago Teixeira
    Tarik Hafyane
    Nikola Stikov
    Cansu Akdeniz
    Andreas Greiser
    Matthias G. Friedrich
    Journal of Cardiovascular Magnetic Resonance, 18