Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children

被引:23
作者
Garnier, R. P. [1 ]
van der Spoel, A. G. E. [1 ]
Sibarani-Ponsen, R. [1 ]
Markhorst, D. G. [2 ]
Boer, C. [1 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Dept Anesthesiol, NL-1081 HV Amsterdam, Netherlands
[2] Vrije Univ Amsterdam, Med Ctr, Pediat Intens Care Unit, Inst Cardiovasc Res, NL-1081 HV Amsterdam, Netherlands
关键词
arterial pressure monitors; equipment and supplies; medical devices; paediatrics; peripheral arterial catheterization; FINGER BLOOD-PRESSURE; PERIPHERAL VASOCONSTRICTION; INTRAARTERIAL; FEASIBILITY; VARIABILITY;
D O I
10.1093/bja/aes295
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We compared Nexfin non-invasive arterial pressure measurements using a novel small finger cuff with intra-arterial pressure in the paediatric setting in order to establish the level of agreement between both methods. The study included 41 children aged 216 yr admitted for surgery or paediatric intensive care with an intra-arterial catheter as part of standard monitoring. Values of systolic (SAP), diastolic (DAP), and mean arterial pressure (MAP) were obtained simultaneously from the intra-arterial catheter and the non-invasive Nexfin monitor. Data were analysed using intra-class correlation (ICC) coefficients and the BlandAltman analysis. A non-invasive arterial pressure signal was obtained in the majority of patients. The reproducibility of arterial pressure measurements over time by both non-invasive and invasive techniques was high, with ICC coefficients ranging from 0.94 to 0.98. The BlandAltman analysis for SAP, DAP, and MAP revealed a bias with 95 limits of agreement of 13.5 (39.7; 12.8), 0.2 (12.8; 13.2), and 2.6 (17.7; 12.5) mm Hg, respectively. Linear regression suggested a weak correlation of SAP and the bias between intra-arterial and Nexfin SAP measurements (intercept 4.9 mm Hg, 0.29; P0.01). Nexfin non-invasive arterial pressure measurements are feasible in paediatric patients. Nexfin accurately reflects the intra-arterial MAP and DAP curves, but seems to underestimate SAP compared with intra-arterial pressure. These results suggest that Nexfin may be used in low-to-moderate risk children without severe systemic hypotension, who require beat-to-beat haemodynamic monitoring but do not have an indication for invasive measurements.
引用
收藏
页码:609 / 615
页数:7
相关论文
共 19 条
[1]   Noninvasive assessment of blood pressure variability in preterm infants [J].
Andriessen, P ;
Schoffelen, RLM ;
Berendsen, RCM ;
De Beer, NAM ;
Oei, SG ;
Wijn, PFF ;
Blanco, CE .
PEDIATRIC RESEARCH, 2004, 55 (02) :220-223
[2]   Feasibility of Noninvasive continuous finger arterial blood pressure measurements in very young children, aged 0-4 years [J].
Andriessen, Peter ;
Schraa, Olaf ;
Van Den Bosch-Ruis, Wendy ;
Ten Harkel, Derk Jan ;
Settels, Jos J. ;
Oetomo, Sidarto Bambang ;
Blanco, Carlos E. .
PEDIATRIC RESEARCH, 2008, 63 (06) :691-696
[3]  
Association for the Advancement of Medical Instrumentation, 1993, ANSI AAMI SP, P40
[4]   Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease [J].
Bogert, L. W. J. ;
Wesseling, K. H. ;
Schraa, O. ;
Van Lieshout, E. J. ;
de Mol, B. A. J. M. ;
van Goudoever, J. ;
Westerhof, B. E. ;
van Lieshout, J. J. .
ANAESTHESIA, 2010, 65 (11) :1119-1125
[5]   Assessment of short-term blood pressure variability in anesthetized children: A comparative study between intraarterial and finger blood pressure [J].
Constant, I ;
Laude, D ;
Elghozi, JL ;
Murat, I .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (3-4) :205-214
[6]   EFFECTS OF PERIPHERAL VASOCONSTRICTION ON THE BLOOD-PRESSURE IN THE FINGER, MEASURED CONTINUOUSLY BY A NEW NONINVASIVE METHOD (THE FINAPRES) [J].
DORLAS, JC ;
NIJBOER, JA ;
BUTIJN, WT ;
VANDERHOEVEN, GMA ;
SETTELS, JJ ;
WESSELING, KH .
ANESTHESIOLOGY, 1985, 62 (03) :342-345
[7]   Feasibility of using finger arterial pressure in neonates [J].
Drouin, E ;
Gournay, V ;
Calamel, J ;
Mouzard, A ;
Roze, JC .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1997, 77 (02) :F139-F140
[8]   Continuous non-invasive finger arterial pressure monitoring reflects intra-arterial pressure changes in children undergoing cardiac surgery [J].
Hofhuizen, C. M. ;
Lemson, J. ;
Hemelaar, A. E. A. ;
Settels, J. J. ;
Schraa, O. ;
Singh, S. K. ;
van der Hoeven, J. G. ;
Scheffer, G. J. .
BRITISH JOURNAL OF ANAESTHESIA, 2010, 105 (04) :493-500
[9]   THE FINAPRES 2300E FINGER CUFF - THE INFLUENCE OF CUFF APPLICATION ON THE ACCURACY OF BLOOD-PRESSURE MEASUREMENT [J].
JONES, RDM ;
KORNBERG, JP ;
ROULSON, CJ ;
VISRAM, AR ;
IRWIN, MG .
ANAESTHESIA, 1993, 48 (07) :611-615
[10]   The Reliability of Continuous Noninvasive Finger Blood Pressure Measurement in Critically III Children [J].
Lemson, Joris ;
Hofhuizen, Charlotte M. ;
Schraa, Olaf ;
Settels, Jos J. ;
Scheffer, Gert Jan ;
van der Hoeven, Johannes G. .
ANESTHESIA AND ANALGESIA, 2009, 108 (03) :814-821