Optimal Location of Pulse Compression in Coded Excitation for Medical Ultrasound Imaging

被引:0
|
作者
Kim, Hoonmin [1 ]
Yoon, Changhan [1 ]
Lee, Wooyeol [1 ]
Yoo, Yangmo [1 ]
机构
[1] Sogang Univ, Dept Elect Engn, Seoul, South Korea
关键词
Coded Excitation; Pulse Compression; Optimal Location; Barker; Golay; weighted Chirp;
D O I
10.1109/ULTSYM.2012.0415
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Coded excitation methods can enhance signal-to-noise ratio (SNR) in medical ultrasound imaging. Specifically, the location of pulse compression determines the computational complexity and image quality. Thus, the optimal location of pulse compression is important for coded excitation-based medical ultrasound imaging systems. In this paper, to determine the optimal location of pulse compression, the axial resolution with three different types of codes (i.e. Barker, Golay, and weighted Chirp) was examined. The pulse compression filter can be applied to various locations, i.e., behind analog-to-digital converter (ADC), beamformer (BF), quadrature demodulator (QD). In the Field II simulation, the codes (i.e., Barker, Golay and weighted Chirp with 13, 16 and 16 cycles, respectively) are used. On the other hand, for phantom experiments, prebeamformed RF data from Barker and Golay codes were captured by a 4-MHz convex probe connected to a commercial ultrasound machine with a research package (SonixTouch, Ultrasonix, Vancouver, BC, Canada.) For all cases, the best image quality was given when pulse compression was placed behind ADC, but it is difficult to be implemented due to high hardware complexity. Behind BF and QD, for Barker and Golay, the degradation in axial resolution was observed due to nonlinear sampling during dynamic receive focusing. This degradation was minimal for Golay. On the other hand, Chirp is robust to the nonlinear sampling artifact regardless the location of pulse compression. Thus, the optimal location of pulse compression for Chirp is the behind beamformer that yields the smaller hardware complexity.
引用
收藏
页码:1655 / 1657
页数:3
相关论文
共 50 条
  • [1] Combined Pulse Compression and Adaptive Beamforming in Coded Excitation Ultrasound Medical Imaging
    Toosi, Tahereh Kahookar
    Behnam, Hamid
    PROCEEDINGS OF THE 2009 INTERNATIONAL CONFERENCE ON SIGNAL PROCESSING SYSTEMS, 2009, : 210 - 214
  • [2] An Efficient Pulse Compression Method of Chirp-Coded Excitation in Medical Ultrasound Imaging
    Yoon, Changhan
    Lee, Wooyoul
    Chang, Jin Ho
    Song, Tai-Kyong
    Yoo, Yangmo
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2013, 60 (10) : 2225 - 2229
  • [3] Parameter optimization of pulse compression in ultrasound imaging systems with coded excitation
    Behar, V
    Adam, D
    ULTRASONICS, 2004, 42 (10) : 1101 - 1109
  • [4] A New Dynamic Complex Baseband Pulse Compression Method for Chirp-Coded Excitation in Medical Ultrasound Imaging
    Kang, Jinbum
    Kim, Yeajin
    Lee, Wooyoul
    Yoo, Yangmo
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (11) : 1698 - 1710
  • [5] Potential of coded excitation in medical ultrasound imaging
    Misaridis, TX
    Gammelmark, K
    Jorgensen, CH
    Lindberg, N
    Thomsen, AH
    Pedersen, MH
    Jensen, JA
    ULTRASONICS, 2000, 38 (1-8) : 183 - 189
  • [6] An Improved Chirp Coded Excitation Based on Compression Pulse Weighting Method in Endoscopic Ultrasound Imaging
    Wang, Ninghao
    Yang, Chen
    Xu, Jie
    Shi, Wei
    Huang, Wenchang
    Cui, Yaoyao
    Jian, Xiaohua
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (03) : 446 - 452
  • [7] Chirp-Coded Pulse Excitation for Ultrasound Elasticity Imaging
    Peng, Hui
    Liu, Dong C.
    2010 4TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING (ICBBE 2010), 2010,
  • [8] Coded excitation and nonlinear pulse compression in pulse-inversion fundamental imaging
    Shen, Che-Chou
    Cheng, Yun-Chien
    Li, Pai-Chi
    2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 1744 - 1747
  • [9] MEDICAL ULTRASOUND IMAGING USING PULSE-COMPRESSION
    RAO, N
    MEHRA, S
    ELECTRONICS LETTERS, 1993, 29 (08) : 649 - 651
  • [10] Coded excitation with sigma-delta sampling for medical ultrasound imaging
    Yoo, Yang Mo
    Cao, Ji
    Koh, Liang Mong
    Kim, Yongmin
    MEDICAL IMAGING 2006: ULTRASONIC IMAGING AND SIGNAL PROCESSING, 2006, 6147