Anterior Chamber Depth Measurement in Phakic and Pseudophakic Eyes

被引:43
|
作者
Su, Po-Fang [1 ]
Lo, Andy Y. [1 ]
Hu, Chao-Yu [1 ]
Chang, Shu-Wen [1 ,2 ]
机构
[1] Far Eastern Mem Hosp, Dept Ophthalmol, Taipei 220, Taiwan
[2] Natl Taiwan Univ Hosp, Dept Ophthalmol, Taipei, Taiwan
关键词
anterior chamber depth; Pentacam; IOLMaster; A-scan ultrasonography; blue light-filtering IOLs;
D O I
10.1097/OPX.0b013e31818e8ceb
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To compare anterior chamber depth (ACD) measurements using two non-contact optical devices, Pentacam and IOLMaster, and a contact device, ultrasonic A-scan in phakic and pseudophakic eyes. Methods. Ninety phakic and 94 pseudophakic eyes were enrolled in this prospective Study. The difference between ACD measurements by the three devices was analyzed using the repeated-measures analysis of variance, and agreement among the three measurements was investigated. The accuracy of detecting the anterior lens surface of the intraocular lens (IOL) with or without blue-blocker was also assessed. Results. In phakic eyes, the Pentacam measured the deepest ACD with the smallest standard deviation (3.26 +/- 0.41, 3.20 +/- 0.45, and 3.12 +/- 0.44 mm measured by Pentacam, IOLMaster, and A-scan, respectively, p < 0.001). In contrast, the Pentacam measurement of ACD had the largest standard deviation in pseudophakic eyes (4.05 +/- 0.58, 4.06 +/- 0.46, and 3.81 +/- 0.41 mm by Pentacam, IOL-Master, and A-scan, respectively, p < 0.001). The Pentacam instrument failed to correctly identify the anterior lens surface in 26.4% of IOLs without blue-blocker and in 58.5% of blue light-filtering IOLs (p = 0.016). Manual correction of the Pentacam image and subsequent ACD measurement improved the agreement between Pentacam and the other two devices. After manual adjustment, the average ACD value increased, and the standard deviation decreased significantly (4.34 +/- 0.28 mm). Conclusions. ACD measurements by Pentacam, IOLMaster, and A-scan were in better agreement in phakic eyes when compared with pseudophakic eyes. In pseudophakic eyes, the 95% limits of agreements between Pentacam and IOLMaster as well as Pentacam and A-scan were unsatisfactory. IOLs with blue-blocker might interfere with ACD measurements. However, this error can be corrected manually in the Pentacam examination.
引用
收藏
页码:1193 / 1200
页数:8
相关论文
共 50 条
  • [41] Distribution of Anterior Chamber Angle Width and Correlation With Age, Refraction, and Anterior Chamber Depth-The Gutenberg Health Study
    Schuster, Alexander K.
    Pfeiffer, Norbert
    Nickels, Stefan
    Schulz, Andreas
    Hoehn, Rene
    Wild, Philipp S.
    Binder, Harald
    Muenzel, Thomas
    Beutel, Manfred E.
    Vossmerbaeumer, Urs
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (08) : 3740 - 3746
  • [42] Effects of atropine on anterior chamber depth and anterior chamber inflammation after primary trabeculectomy
    Orengo-Nania, S
    El-Harazi, SM
    Oram, O
    Feldman, RM
    Chuang, AZ
    Gross, RL
    JOURNAL OF GLAUCOMA, 2000, 9 (04) : 303 - 310
  • [43] Changes in anterior chamber volume after implantation of posterior chamber phakic intraocular lens in high myopia
    Zhu, Yi
    Zhu, Haobin
    Jia, Yan
    Zhou, Jibo
    BMC OPHTHALMOLOGY, 2018, 18
  • [44] Anterior Segment Characteristics of Eyes with Anterior Chamber Depth Less than 2.8 mm and Axial Length Greater than 25 mm
    Niu, Lingling
    Luo, Xiumei
    Chen, Xueli
    Wang, Xiaoying
    Zhou, Xingtao
    Qian, Yishan
    OPHTHALMOLOGY AND THERAPY, 2023, 12 (02) : 1195 - 1206
  • [45] Anterior segment optical coherence tomography for evaluation of changes in anterior chamber angle and depth after intraocular lens implantation in eyes with glaucoma
    Dawczynski, J.
    Koenigsdoerffer, E.
    Augsten, R.
    Strobel, J.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2007, 17 (03) : 363 - 367
  • [46] Phakic Intraocular Lens Implantation: Refractive Outcome and Safety in Patients with Anterior Chamber Depth between 2.8 and 3.0 versus ≥3.0 mm
    Guerra, Marta G.
    Silva, Andreia M. M.
    Marques, Sara H. M.
    Melo, Sofia H.
    Povoa, Joao A.
    Lobo, Conceicao
    Neto Murta, Joaquim
    OPHTHALMIC RESEARCH, 2017, 57 (04) : 239 - 246
  • [47] Anterior Segment Characteristics of Eyes with Anterior Chamber Depth Less than 2.8 mm and Axial Length Greater than 25 mm
    Lingling Niu
    Xiumei Luo
    Xueli Chen
    Xiaoying Wang
    Xingtao Zhou
    Yishan Qian
    Ophthalmology and Therapy, 2023, 12 : 1195 - 1206
  • [48] Diurnal change of anterior chamber depth in rabbits
    Larsson, LI
    Brubaker, RF
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 1995, 73 (06): : 534 - 536
  • [49] Changes in anterior chamber depth following vitrectomy
    Li Yang
    Yang Cheng-xun
    Qing Guo-ping
    Wei Wen-bin
    CHINESE MEDICAL JOURNAL, 2013, 126 (19) : 3701 - 3704
  • [50] Anterior chamber depth estimation by Scheimpflug photography
    Holmén, JB
    Ekesten, B
    Lundgren, B
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 2001, 79 (06): : 576 - 579