Measurements of endotracheal tube cuff contact pressure using fibre Bragg gratings

被引:3
作者
Hernandez, F. U. [1 ]
Correia, R. [1 ]
Korposh, S. [1 ]
Morgan, S. P. [1 ]
Hayes-Gill, B. R. [1 ]
James, S. W. [2 ]
Evans, D. [3 ]
Norris, A. [3 ]
机构
[1] Univ Nottingham, Elect Syst & Opt Res Div, Appl Opt Grp, Nottingham NG7 2RD, Notts, England
[2] Cranfield Univ, Sch Engn, Dept Engn Photon, Cranfield MK43 0AL, Beds, England
[3] Queens Med Ctr, Dept Anaesthesia, Nottingham NG7 2UH, England
来源
24TH INTERNATIONAL CONFERENCE ON OPTICAL FIBRE SENSORS | 2015年 / 9634卷
关键词
Fibre Bragg Grating (FBG); intracuff pressure; contact pressure; endotracheal tube (ETT);
D O I
10.1117/12.2195118
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An optical fibre Bragg grating (FBG) was used to measure local strain (due to contact pressure) at the interface of a cuffed endotracheal tube (ETT) tested in a tracheal model. The tracheal model consisted of a corrugated tube. Two FBG sensors written in a single optical fibre were attached to the outside wall of the cuff of the ETT. Intracuff endotracheal pressure was measured using a digital manometer, while the contact pressure between the model trachea and the ETT was measured using Flexiforce sensors. Changes in the Bragg wavelengths in response to the inflation of the cuff of the ETT, and concomitant pressure increase, were observed to be dependent on the location of the FBGs at the corrugations, i.e., the annular peaks and troughs of the corrugated tube. The performance of both contact pressure sensors FBG and Flexiforce suggests that FBG technology is better suited to this application as it allows the measurement of contact pressures on non-uniform surfaces such as in the tracheal model.
引用
收藏
页数:4
相关论文
共 5 条