Short-scan-time multi-slice diffusion MRI of the mouse cervical spinal cord using echo planar imaging

被引:17
作者
Callot, Virginie [1 ]
Duhamel, Guillaume [1 ]
Cozzone, Patrick J. [1 ]
Kober, Frank [1 ]
机构
[1] Univ Aix Marseille 2, CRMBM, UMR CNRS 6612, Fac Med Marseille, F-13385 Marseille 05, France
关键词
mouse; spinal cord; MRI; echo planar imaging; diffusion; high field; fiber tracking;
D O I
10.1002/nbm.1274
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Mouse spinal cord (SC) diffusion-weighted imaging (DWI) provides important information on tissue morphology and structural changes that may occur during pathologies such as multiple sclerosis or SC injury. The acquisition scheme of the commonly used DWI techniques is based on conventional spin-echo encoding, which is time-consuming. The purpose of this work was to investigate whether the use of echo planar imaging (EPI) would provide good-quality diffusion MR images of mouse SC, as well as accurate measurements of diffusion-derived metrics, and thus enable diffusion tensor imaging (DTI) and highly resolved DWI within reasonable scan times. A four-shot diffusion-weighted spin-echo EPI (SE-EPI) sequence was evaluated at 11.75 T oil a group of healthy mice (n = 10). SE-EPI-derived apparent diffusion coefficients of gray and while matter were compared with those obtained using a conventional spin-echo sequence (c-SE) to validate the accuracy of the method. To take advantage of the reduction in acquisition time offered by the EPI sequence, multi-slice DTI acquisitions were performed covering the cervical segments (six slices, six diffusion-encoding, directions. three b values) within 30 min (vs 2 h for c-SE). Front these measurements. fractional anisotropy and mean diffusivities were calculated. and fiber tracking along the C1 to C6 cervical segments was performed. In addition, high-resolution images (74 x 94 mu m(2)) were acquired within 5 min per direction. Clear delineation of gray and white matter and identical apparent diffusion coefficient values were obtained, with a threefold reduction in acquisition time compared with c-SE. While overcoming the difficulties associated with high spatially and temporally resolved DTI measurements, the present SE-EPI approach permitted identification of reliable quantitative parameters with a reproducibility compatible with the detection of pathologies. The SE-EPI method may be particularly valuable when multiple sets of images from the SC are needed, in cases of rapidly evolving, conditions, to decrease the duration of anesthesia or to improve MR exploration by including additional MR measurements. Copyright (C) 2008 John Wiley & Sons, Ltd.
引用
收藏
页码:868 / 877
页数:10
相关论文
共 29 条
  • [1] MR microscopy of transgenic mice that spontaneously acquire experimental allergic encephalomyelitis
    Ahrens, ET
    Laidlaw, DH
    Readhead, C
    Brosnan, CF
    Fraser, SE
    Jacobs, RE
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (01) : 119 - 132
  • [2] [Anonymous], 2003, NEUROIMAGE
  • [3] The basis of anisotropic water diffusion in the nervous system - a technical review
    Beaulieu, C
    [J]. NMR IN BIOMEDICINE, 2002, 15 (7-8) : 435 - 455
  • [4] Simple, low-cost multipurpose RF coil for MR microscopy at 9.4 T
    Bilgen, M
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (04) : 937 - 940
  • [5] Magnetic resonance imaging of mouse spinal cord
    Bilgen, M
    Al-Hafez, B
    Berman, NEJ
    Festoff, BW
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (05) : 1226 - 1231
  • [6] Imaging corticospinal tract connectivity in injured rat spinal cord using manganese-enhanced MRI
    Bilgen M.
    [J]. BMC Medical Imaging, 6 (1)
  • [7] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [8] Nuclear magnetic resonance microimaging of mouse spinal cord in vivo
    Bonny, JM
    Gaviria, M
    Donnat, JP
    Jean, B
    Privat, A
    Renou, JP
    [J]. NEUROBIOLOGY OF DISEASE, 2004, 15 (03) : 474 - 482
  • [9] Toward accurate diagnosis of white matter pathology using diffusion tensor imaging
    Budde, Matthew D.
    Kim, Joong Hee
    Liang, Hsiao-Fang
    Schmidt, Robert E.
    Russell, John H.
    Cross, Anne H.
    Song, Sheng-Kwei
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (04) : 688 - 695
  • [10] In vivo mouse spinal cord imaging using echo-planar imaging at 11.75 T
    Callot, Virginie
    Duhamel, Guillaume
    Cozzone, Patrick J.
    [J]. MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2007, 20 (04) : 169 - 173