We investigated a new intramedullary locking nail that allows the distal interlocking screws to be locked to the nail. We compared fixation using this new implant with fixation using either a conventional nail or a locking plate in a laboratory simulation of an osteoporotic fracture of the distal femur. A total of 15 human cadaver femora were used to simulate an AO 33-A3 fracture pattern. Paired specimens compared fixation using either a locking or non-locking retrograde nail, and using either a locking retrograde nail or a locking plate. The constructs underwent cyclical loading to simulate single-leg stance up to 125 000 cycles. Axial and torsional stiffness and displacement, cycles to failure and modes of failure were recorded for each specimen. When compared with locking plate constructs, locking nail constructs had significantly longer mean fatigue life (75 800 cycles (SD 33 900) vs 12 800 cycles (SD 6100); p = 0.007) and mean axial stiffness (220 N/mm (SD 80) vs 70 N/mm (SD 18); p = 0.005), but lower mean torsional stiffness (2.5 Nm/degrees (SD 0.9) vs 5.1 Nm/degrees (SD 1.5); p = 0.008). In addition, in the nail group the mode of failure was either cut-out of the distal screws or breakage of nails, and in the locking plate group breakage of the plate was always the mode of failure. Locking nail constructs had significantly longer mean fatigue life than non-locking nail constructs (78 900 cycles (SD 25 600) vs 52 400 cycles (SD 22 500); p = 0.04). The new locking retrograde femoral nail showed better stiffness and fatigue life than locking plates, and superior fatigue life to non-locking nails, which may be advantageous in elderly patients.
引用
收藏
页码:114 / 121
页数:8
相关论文
共 23 条
[1]  
[Anonymous], 1994, WHO Technical Report Series 844
[2]   Comparison of the LISS and a retrograde-inserted supracondylar intramedullary nail for fixation of a periprosthetic distal femur fracture proximal to a total knee arthroplasty [J].
Bong, MR ;
Egol, KA ;
Koval, KJ ;
Kummer, FJ ;
Su, ET ;
Iesaka, K ;
Bayer, J ;
Di Cesare, PE .
JOURNAL OF ARTHROPLASTY, 2002, 17 (07) :876-881
[3]   Immediate weight-bearing after treatment of a comminuted fracture of the femoral shaft with a statically locked intramedullary nail [J].
Brumback, RJ ;
Toal, TR ;
Murphy-Zane, MS ;
Novak, VP ;
Belkoff, SM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (11) :1538-1544
[4]  
Cornell C, 2011, HSS J, V7, P164, DOI DOI 10.1007/S11420-010-9194-8
[5]   Internal forces and moments in the femur during walking [J].
Duda, GN ;
Schneider, E ;
Chao, EYS .
JOURNAL OF BIOMECHANICS, 1997, 30 (09) :933-941
[6]   Minimal-invasive treatment of distal femoral fractures with the LISS (Less Invasive Stabilization System) - A prospective study of 30 fractures with a follow up of 20 months [J].
Fankhauser, F ;
Gruber, G ;
Schippinger, G ;
Boldin, C ;
Hofer, HP ;
Grechenig, W ;
Szyszkowitz, R .
ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (01) :56-60
[7]   The development of the distal femur Less Invasive Stabilization System (LISS) [J].
Frigg, R ;
Appenzeller, A ;
Christensen, R ;
Frenk, A ;
Gilbert, S ;
Schavan, R .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 :24-31
[8]   Experience of the T2 supracondylar nail in distal femoral fractures [J].
Ghandour, A. ;
Cosker, T. D. A. ;
Kadambande, S. S. ;
Pallister, I. ;
Russell, I. D. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 (10) :1019-1025
[9]   Distal Femoral Fixation: A Biomechanical Comparison of Trigen Retrograde Intramedullary (IM) Nail, Dynamic Condylar Screw (DCS), and Locking Compression Plate (LCP) Condylar Plate [J].
Heiney, Jake P. ;
Barnett, Michael D. ;
Vrabec, Gregory A. ;
Schoenfeld, Andrew J. ;
Baji, Avinash ;
Njus, Glen O. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2009, 66 (02) :443-449
[10]   Treatment of Distal Femoral Fracture by Minimally Invasive Percutaneous Plate Osteosynthesis: Comparison Between the Dynamic Condylar Screw and the Less Invasive Stabilization System [J].
Kao, Feng C. ;
Tu, Yuan K. ;
Su, Juin Y. ;
Hsu, Kuo Y. ;
Wu, Chin H. ;
Chou, Ming C. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2009, 67 (04) :719-726