Validation of a Clinical PET Scanner Using Monte Carlo Simulation Code: MCNP5

被引:0
|
作者
Musarudin, Marianie [1 ]
Saripan, M. Iqbal [1 ]
Mashohor, Syamsiah [1 ]
Saad, Wira Hidayat Mohd [1 ]
Hashim, Suhairul [2 ]
Nordin, Abdul Jalil [3 ]
机构
[1] Univ Putra Malaysia, Fac Engn, Serdang 43400, Malaysia
[2] Univ Teknol Malaysia, Fac Sci, Skudai, Johor, Malaysia
[3] Univ Putra Malaysia, Ctr Diagnost Nucl Imaging, Serdang 43400, Malaysia
来源
2013 8TH EUROSIM CONGRESS ON MODELLING AND SIMULATION (EUROSIM) | 2013年
关键词
validation; Monte Carlo; PET; WIRE-MESH COLLIMATOR; ATTENUATION CORRECTION; SPATIAL-RESOLUTION; BLOCK DETECTORS; PERFORMANCE; INSTRUMENTATION; DESIGN;
D O I
10.1109/EUROSIM.2013.16
中图分类号
TP301 [理论、方法];
学科分类号
081202 ;
摘要
This paper presents a realistic PET scanner simulation using Monte Carlo code, version MCNP5. The objective of our study was to verify the code used for this simulation by comparing the results obtained from the simulation with those obtained from the measurement did on a real PET scanner. This study will provide a basic benchmark for our further study on PET imaging. We modeled the scanner based on the physical specification of Siemens Biograph TruePoint PET scanner. We recorded the generated list-mode data which contains all the information needed to model the PET processes for instance coincidence photon detection. To account for the statistical fluctuations occur in the detector and photomultiplier tube, a Gaussian energy blurring model was applied to the energy deposited in the detector. The scatter and attenuation correction to correct the effect of scattered and attenuated events also took into account in this study. All of these post-simulation processes were done using a program developed using matlab. To validate the simulation, the simulated and measured energy and spatial resolution were compared. We have successfully modeled a PET system based on MCNP5. We also verified that this simulation result in a good agreement data with the real imaging.
引用
收藏
页码:36 / 41
页数:6
相关论文
共 50 条
  • [1] Uncertainty analysis in environmental radioactivity measurements using the Monte Carlo code MCNP5
    Gallardo, S.
    Querol, A.
    Ortiz, J.
    Rodenas, J.
    Verdu, G.
    Villanueva, J. F.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 116 : 214 - 218
  • [2] Using lattice tools and unfolding methods for hpge detector efficiency simulation with the Monte Carlo code MCNP5
    Querol, A.
    Gallardo, S.
    Rodenas, J.
    Verdu, G.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 116 : 219 - 225
  • [3] Modeling and experimental verification of Biograph TruePoint PET/CT using MCNP5
    Waeleh, Nazreen
    Saripan, M. Iqbal
    Musarudin, Marianie
    Mashohor, Syamsiah
    Saad, Fathinul Fikri Ahmad
    2020 IEEE-EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES 2020): LEADING MODERN HEALTHCARE TECHNOLOGY ENHANCING WELLNESS, 2021, : 319 - 323
  • [4] Validation of a Monte Carlo simulation of the Inveon PET scanner using GATE
    Lu, Lijun
    Zhang, Houjin
    Bian, Zhaoying
    Ma, Jianhua
    Feng, Qiangjin
    Chen, Wufan
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 828 : 170 - 175
  • [5] Evaluation of an extrapolation chamber for dosimetry in computed tomography beams using Monte Carlo code (MCNP5)
    Castro, Maysa C.
    Silva, Natalia F.
    Santos, Lucas R.
    Cintra, Felipe B.
    Caldas, Linda V. E.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 167
  • [6] Validation of a Monte Carlo model of the uEXPLORER total-body PET scanner using GATE code
    EL Katib, Mahmoud
    Chakir, El Mahjoub
    Sebihi, Rajaa
    Saikouk, Hind
    Nhila, Oussama
    RADIATION PHYSICS AND CHEMISTRY, 2023, 210
  • [7] Monte Carlo calculations of electrons impinging on a copper target: A comparison of EGSnrc, Geant4 and MCNP5
    Archambault, John Paul
    APPLIED RADIATION AND ISOTOPES, 2018, 132 : 129 - 134
  • [8] Extensions of the MCNP5 and TRIPOLI4 Monte Carlo Codes for Transient Reactor Analysis
    Hoogenboom, J. Eduard
    Sjenitzer, Bart L.
    SNA + MC 2013 - JOINT INTERNATIONAL CONFERENCE ON SUPERCOMPUTING IN NUCLEAR APPLICATIONS + MONTE CARLO, 2014,
  • [9] Design of a Monte Carlo model based on dual-source computed tomography (DSCT) scanners for dose and image quality assessment using the Monte Carlo N-Particle (MCNP5) code
    Chantzi, Stefania
    Papanastasiou, Emmanouil
    Athanasopoulou, Christina
    Molyvda-Athanasopoulou, Elisavet
    Bamidis, Panagiotis
    Siountas, Anastasios
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2020, 26 (01): : 11 - 20
  • [10] A review of radiation dosimetry applications using the MCNP Monte Carlo code
    Solberg, TD
    DeMarco, JJ
    Chetty, IJ
    Mesa, AV
    Cagnon, CH
    Li, AN
    Mather, KK
    Medin, PM
    Arellano, AR
    Smathers, JB
    RADIOCHIMICA ACTA, 2001, 89 (4-5) : 337 - 355