Neutron measurements with ultra-thin 3D silicon sensors in a radiotherapy treatment room using a Siemens PRIMUS linac

被引:23
作者
Guardiola, C. [1 ]
Gomez, F. [2 ]
Fleta, C. [1 ]
Rodriguez, J. [1 ]
Quirion, D. [1 ]
Pellegrini, G. [1 ]
Lousa, A. [3 ]
Martinez-de-Olcoz, L. [3 ]
Pombar, M. [4 ]
Lozano, M. [1 ]
机构
[1] Inst Microelect Barcelona IMB CNM CSIC, E-08193 Barcelona, Spain
[2] Univ Santiago de Compostela, Fac Fis, Dept Fis Particulas, E-15782 Santiago De Compostela, Spain
[3] Univ Barcelona, Fac Fis, Dept Fis Aplicada & Opt, E-08028 Barcelona, Spain
[4] Hosp Clin Univ Santiago, E-15782 Santiago De Compostela, Spain
关键词
MONTE-CARLO;
D O I
10.1088/0031-9155/58/10/3227
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The accurate detection and dosimetry of neutrons in mixed and pulsed radiation fields is a demanding instrumental issue with great interest both for the industrial and medical communities. In recent studies of neutron contamination around medical linacs, there is a growing concern about the secondary cancer risk for radiotherapy patients undergoing treatment in photon modalities at energies greater than 6 MV. In this work we present a promising alternative to standard detectors with an active method to measure neutrons around a medical linac using a novel ultra-thin silicon detector with 3D electrodes adapted for neutron detection. The active volume of this planar device is only 10 mu m thick, allowing a high gamma rejection, which is necessary to discriminate the neutron signal in the radiotherapy peripheral radiation field with a high gamma background. Different tests have been performed in a clinical facility using a Siemens PRIMUS linac at 6 and 15 MV. The results show a good thermal neutron detection efficiency around 2% and a high gamma rejection factor.
引用
收藏
页码:3227 / 3242
页数:16
相关论文
共 21 条
[1]  
AAPM, 1986, 19 AAPM
[2]   Neutron spectra and dosimetric features around an 18 MV Linac accelerator [J].
Barquero, R ;
Mendez, R ;
Vega-Carrillo, HR ;
Iñiguez, MP ;
Edwards, TM .
HEALTH PHYSICS, 2005, 88 (01) :48-58
[3]   The physics of solid-state neutron detector materials and geometries [J].
Caruso, A. N. .
JOURNAL OF PHYSICS-CONDENSED MATTER, 2010, 22 (44)
[4]  
Garcia Fuste M J, 2010, THESIS U AUTONOMA BA
[5]   A new active method for the measurement of slow-neutron fluence in modern radiotherapy treatment rooms [J].
Gomez, F. ;
Iglesias, A. ;
Sanchez Doblado, F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (04) :1025-1039
[6]   Ultra-thin 3D silicon sensors for neutron detection [J].
Guardiola, C. ;
Fleta, C. ;
Pellegrini, G. ;
Garcia, F. ;
Quirion, D. ;
Rodriguez, J. ;
Lozano, M. .
JOURNAL OF INSTRUMENTATION, 2012, 7
[7]   Geant4 and MCNPX simulations of thermal neutron detection with planar silicon detectors [J].
Guardiola, C. ;
Amgarou, K. ;
Garcia, F. ;
Fleta, C. ;
Quirion, D. ;
Lozano, M. .
JOURNAL OF INSTRUMENTATION, 2011, 6
[8]  
Guardiola C, 2011, P 8 SPAN C EL DEV
[9]  
Guardiola C, 2012, Patent Specification, Patent No. [P201231604, 201231604]
[10]  
ICRP, 1992, ANN ICRP, V33