Soft X-ray radiation damage in EM-CCDs used for Resonant Inelastic X-ray Scattering

被引:6
|
作者
Gopinath, D. [1 ,2 ]
Soman, M. [1 ]
Holland, A. [1 ]
Keelan, J. [1 ]
Hall, D. [1 ]
Holland, K. [2 ]
Colebrook, D. [2 ]
机构
[1] Open Univ, Ctr Elect Imaging, Walton Hall, Milton Keynes MK7 6AA, Bucks, England
[2] XCAM Ltd, 2 Stone Circle Rd, Northampton NN3 8RF, England
来源
JOURNAL OF INSTRUMENTATION | 2018年 / 13卷
关键词
Radiation damage to detector materials (solid state); X-ray detectors; Spectrometers; RESOLUTION; DETECTORS; NOISE; RIXS;
D O I
10.1088/1748-0221/13/02/C02027
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Advancement in synchrotron and free electron laser facilities means that X-ray beams with higher intensity than ever before are being created. The high brilliance of the X-ray beam, as well as the ability to use a range of X-ray energies, means that they can be used in a wide range of applications. One such application is Resonant Inelastic X-ray Scattering (RIXS). RIXS uses the intense and tuneable X-ray beams in order to investigate the electronic structure of materials. The photons are focused onto a sample material and the scattered X-ray beam is diffracted off a high resolution grating to disperse the X-ray energies onto a position sensitive detector. Whilst several factors affect the total system energy resolution, the performance of RIXS experiments can be limited by the spatial resolution of the detector used. Electron-Multiplying CCDs (EM-CCDs) at high gain in combination with centroiding of the photon charge cloud across several detector pixels can lead to sub-pixel spatial resolution of 2-3 mu m. X-ray radiation can cause damage to CCDs through ionisation damage resulting in increases in dark current and/or a shift in flat band voltage. Understanding the effect of radiation damage on EM-CCDs is important in order to predict lifetime as well as the change in performance over time. Two CCD-97s were taken to PTB at BESSY II and irradiated with large doses of soft X-rays in order to probe the front and back surfaces of the device. The dark current was shown to decay over time with two different exponential components to it. This paper will discuss the use of EM-CCDs for readout of RIXS spectrometers, and limitations on spatial resolution, together with any limitations on instrument use which may arise from X-rayinduced radiation damage.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] The use of CCDs and EM-CCDs for future soft X-ray Spectrometers
    Tutt, James H.
    Holland, Andrew D.
    Murray, Neil J.
    Hall, David J.
    Endicott, James
    McEntaffer, Randall L.
    SPACE TELESCOPES AND INSTRUMENTATION 2012: ULTRAVIOLET TO GAMMA RAY, 2012, 8443
  • [2] The use of EM-CCDs on high resolution soft X-ray spectrometers
    Tutt, James H.
    Holland, Andrew D.
    Murray, Neil J.
    Harriss, Richard D.
    Hall, David J.
    Soman, Matt
    McEntaffer, Randall L.
    Endicott, James
    UV, X-RAY, AND GAMMA-RAY SPACE INSTRUMENTATION FOR ASTRONOMY XVII, 2011, 8145
  • [3] Resonant inelastic soft X-ray scattering
    Eberhardt, W
    Luning, J
    Rubensson, JE
    APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1997, 65 (02): : 89 - 89
  • [4] Resonant inelastic soft x-ray scattering of Be chalcogenides
    Eich, D
    Fuchs, O
    Groh, U
    Weinhardt, L
    Fink, R
    Umbach, E
    Heske, C
    Fleszar, A
    Hanke, W
    Gross, EKU
    Bostedt, C
    Von Buuren, T
    Franco, N
    Terminello, LJ
    Keim, M
    Reuscher, G
    Lugauer, H
    Waag, A
    PHYSICAL REVIEW B, 2006, 73 (11):
  • [5] Improving the spatial resolution of a soft X-ray Charge Coupled Device used for Resonant Inelastic X-ray Scattering
    Soman, M. R.
    Hall, D. J.
    Tutt, J. H.
    Murray, N. J.
    Holland, A. D.
    Schmitt, T.
    Raabe, J.
    Schmitt, B.
    JOURNAL OF INSTRUMENTATION, 2011, 6
  • [6] X-ray resonant inelastic scattering
    Ma, YJ
    JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA, 1996, 79 : 131 - 134
  • [7] Resonant inelastic x-ray scattering
    Platzman, PM
    Isaacs, ED
    PHYSICAL REVIEW B, 1998, 57 (18) : 11107 - 11114
  • [8] Resonant inelastic X-ray scattering
    Nature Reviews Methods Primers, 4 (1):
  • [9] X-ray resonant inelastic scattering
    Ma, Y.
    Journal of Electron Spectroscopy and Related Phenomena, 1996, 79
  • [10] Resonant inelastic X-ray scattering
    de Groot, Frank M. F.
    Haverkort, Maurits W.
    Elnaggar, Hebatalla
    Juhin, Amelie
    Zhou, Ke-Jin
    Glatzel, Pieter
    NATURE REVIEWS METHODS PRIMERS, 2024, 4 (01):