摘要
PURPOSE: To quantify the cut quality of lamellar dissections made with the femtosecond laser using atomic force microscopy (AFM). SETTING: Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA. DESIGN: Experimental study. METHODS: Experiments were performed on 3 pairs of human cadaver eyes. The cornea was thinned to physiologic levels by placing the globe, cornea side down, in 25% dextran for 24 hours. The eyes were reinflated to normal pressures by injecting a balanced salt solution into the vitreous cavity. The eyes were placed in a holder, the epithelium was removed, and the eyes were cut with a Visumax femtosecond laser. The energy level was 180 nJ for the right eye and 340 nJ for the left eye of each pair. The cut depths were 200 mu m, 300 mu m, and 400 mu m, with the cut depth maintained for both eyes of each pair. A 12.0 mm trephination was then performed. The anterior portion of the lamellar surface was placed in a balanced salt solution and imaged with AFM. As a control, the posterior surface was placed in 2% formalin and imaged with environmental scanning electron microscopy (SEM). Four quantitative parameters (root-mean-square deviation, average deviation, skewness, kurtosis) were calculated from the AFM images. RESULTS: From AFM, the 300 mu m low-energy cuts were the smoothest. Similar results were seen qualitatively in the environmental SEM images. CONCLUSION: Atomic force microscopy provided quantitative information on the quality of lamellar dissections made using a femtosecond laser, which is useful in optimizing patient outcomes in refractive and lamellar keratoplasty surgeries. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned. J Cataract Refract Surg 2013; 39:110-117 (C) 2012 ASCRS and ESCRS
引用
收藏
页码:110 / 117
页数:8
相关论文
共 27 条
[1]   Nanoscale topography of the corneal epithelial basement membrane and Descemet's membrane of the human [J].
Abrams, GA ;
Schaus, SS ;
Goodman, SL ;
Nealey, PF ;
Murphy, CJ .
CORNEA, 2000, 19 (01) :57-64
[2]   ATOMIC FORCE MICROSCOPE [J].
BINNIG, G ;
QUATE, CF ;
GERBER, C .
PHYSICAL REVIEW LETTERS, 1986, 56 (09) :930-933
[3]   Preparation and hydration control of corneal tissue strips for experimental use [J].
Borja, D ;
Manns, F ;
Lamar, P ;
Rosen, A ;
Fernandez, V ;
Parel, JM .
CORNEA, 2004, 23 (01) :61-66
[4]   Histologic Evaluation of Human Posterior Lamellar Discs for Femtosecond Laser Descemet's Stripping Endothelial Keratoplasty [J].
Cheng, Yanny Y. Y. ;
Kang, Shin J. ;
Grossniklaus, Hans E. ;
Pels, Elisabeth ;
Duimel, Hans J. Q. ;
Frederik, Peter M. ;
Hendrikse, Fred ;
Nuijts, Rudy M. M. A. .
CORNEA, 2009, 28 (01) :73-79
[5]   ATOMIC-FORCE MICROSCOPY OF THE CORNEA AND SCLERA [J].
FULLWOOD, NJ ;
HAMMICHE, A ;
POLLOCK, HM ;
HOURSTON, DJ ;
SONG, M .
CURRENT EYE RESEARCH, 1995, 14 (07) :529-535
[6]   Comparison of the femtosecond laser (IntraLase) versus manual microkeratome (Moria ALTK) in dissection of the donor in endothelial keratoplasty - Initial study in eye bank eyes [J].
Jones, Yian Jin ;
Goins, Kenneth Mark ;
Sutphin, John E. ;
Mullins, Robert ;
Skeie, Jessica M. .
CORNEA, 2008, 27 (01) :88-93
[7]   Comparative micromorphologic in vitro porcine study of IntraLase and Femto LDV femtosecond lasers [J].
Kermani, Omid ;
Oberheide, Uwe .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (08) :1393-1399
[8]   The Applications of Atomic Force Microscopy to Vision Science [J].
Last, Julie A. ;
Russell, Paul ;
Nealey, Paul F. ;
Murphy, Christopher J. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (12) :6083-6094
[9]  
Lombardo M, 2012, CORNEA
[10]   Atomic force microscopy analysis of normal and photoablated porcine corneas [J].
Lombardo, Marco ;
De Santo, Maria Penelope ;
Lombardo, Giuseppe ;
Barberi, Riccardo ;
Serrao, Sebastiano .
JOURNAL OF BIOMECHANICS, 2006, 39 (14) :2719-2724