Patient-tailored plate for bone fixation and accurate 3D positioning in corrective osteotomy

被引:40
作者
Dobbe, J. G. G. [1 ]
Vroemen, J. C. [2 ]
Strackee, S. D. [2 ]
Streekstra, G. J. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Biomed Engn & Phys, NL-1105 AZ Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Plast Reconstruct & Hand Surg, NL-1105 AZ Amsterdam, Netherlands
关键词
Additive manufacturing; Computer-assisted planning; Custom implant; Cutting guide; Patient-specific treatment; MALUNITED FRACTURES; RADIUS; SIMULATION;
D O I
10.1007/s11517-012-0959-8
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A bone fracture may lead to malunion of bone segments, which gives discomfort to the patient and may lead to chronic pain, reduced function and finally to early osteoarthritis. Corrective osteotomy is a treatment option to realign the bone segments. In this procedure, the surgeon tries to improve alignment by cutting the bone at, or near, the fracture location and fixates the bone segments in an improved position, using a plate and screws. Three-dimensional positioning is very complex and difficult to plan, perform and evaluate using standard 2D fluoroscopy imaging. This study introduces a new technique that uses preoperative 3D imaging to plan positioning and design a patient-tailored fixation plate that only fits in one way and realigns the bone segments as planned. The method is evaluated using artificial bones and renders realignment highly accurate and very reproducible (d (err) < 1.2 +/- A 0.8 mm and phi (err) < 1.8A degrees A A +/- A 2.1A degrees). Application of a patient-tailored plate is expected to be of great value for future corrective osteotomy surgeries.
引用
收藏
页码:19 / 27
页数:9
相关论文
共 36 条
[1]   Computer-assisted distal radius osteotomy [J].
Athwal, GS ;
Ellis, RE ;
Small, CF ;
Pichora, DR .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2003, 28A (06) :951-958
[2]  
Bilic R, 2005, Acta Chir Orthop Traumatol Cech, V72, P375
[3]   Detection of In Vivo Dynamic 3-D Motion Patterns in the Wrist Joint [J].
Carelsen, Bart ;
Jonges, Remmet ;
Strackee, Simon D. ;
Maas, Mario ;
van Kemenade, Patricia ;
Grimbergen, Cornelis A. ;
van Herk, Marcel ;
Streekstra, Geert J. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (04) :1236-1244
[4]   Accuracy in planar cutting of bones: an ISO-based evaluation [J].
Cartiaux, Olivier ;
Paul, Laurent ;
Docquier, Pierre-Louis ;
Francq, Bernard G. ;
Raucent, Benoit ;
Dombre, Etienne ;
Banse, Xavier .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2009, 5 (01) :77-84
[5]   CAD/CAM and rapid prototyped scaffold construction for bone regenerative medicine and surgical transfer of virtual planning: A pilot study [J].
Ciocca, L. ;
De Crescenzio, F. ;
Fantini, M. ;
Scotti, R. .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2009, 33 (01) :58-62
[6]   COMPLICATIONS OF COLLES FRACTURES [J].
COONEY, WP ;
DOBYNS, JH ;
LINSCHEID, RL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (04) :613-619
[7]  
Croitoru H, 2001, Comput Aided Surg, V6, P160, DOI 10.1002/igs.1019
[8]  
Cronier P., 2010, ORTHOP TRAUMATOL-SUR, V96, pS17, DOI DOI 10.1016/J.OTSR.2010.03.008
[9]   Computer-assisted and patient-specific 3-D planning and evaluation of a single-cut rotational osteotomy for complex long-bone deformities [J].
Dobbe, J. G. G. ;
du Pre, K. J. ;
Kloen, P. ;
Blankevoort, L. ;
Streekstra, G. J. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2011, 49 (12) :1363-1370
[10]   Computer-Assisted Planning and Navigation for Corrective Distal Radius Osteotomy, Based on Pre- and Intraoperative Imaging [J].
Dobbe, J. G. G. ;
Strackee, S. D. ;
Schreurs, A. W. ;
Jonges, R. ;
Carelsen, B. ;
Vroemen, J. C. ;
Grimbergen, C. A. ;
Streekstra, G. J. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (01) :182-190