Improvement of γ-ray energy resolution of LaBr3:Ce3+ scintillation detectors by Sr2+ and Ca2+ co-doping

被引:135
作者
Alekhin, M. S. [1 ]
de Haas, J. T. M. [1 ]
Khodyuk, I. V. [1 ]
Kramer, K. W. [2 ]
Menge, P. R. [3 ]
Ouspenski, V. [4 ]
Dorenbos, P. [1 ]
机构
[1] Delft Univ Technol, Fac Sci Appl, Dept Radiat Sci & Technol, NL-2629 JB Delft, Netherlands
[2] Univ Bern, Dept Chem & Biochem, CH-3012 Bern, Switzerland
[3] St Gobain Crystals, Hiram, OH 44234 USA
[4] St Gobain Res, F-93303 Aubervilliers, France
关键词
PROPORTIONALITY; CE3+; PERFORMANCE;
D O I
10.1063/1.4803440
中图分类号
O59 [应用物理学];
学科分类号
摘要
Commercially available LaBr3:5% Ce3+ scintillators show with photomultiplier tube readout about 2.7% energy resolution for the detection of 662 keV gamma-rays. Here we will show that by co-doping LaBr3: Ce3+ with Sr2+ or Ca2+ the resolution is improved to 2.0%. Such an improvement is attributed to a strong reduction of the scintillation light losses that are due to radiationless recombination of free electrons and holes during the earliest stages (1-10 ps) inside the high free charge carrier density parts of the ionization track. (C) 2013 AIP Publishing LLC. [http://dx.doi.org/10.1063/1.4803440]
引用
收藏
页数:4
相关论文
共 22 条
[1]   Scintillation Properties of and Self Absorption in SrI2:Eu2+ [J].
Alekhin, Mikhail S. ;
de Haas, Johan T. M. ;
Kraemer, Karl W. ;
Dorenbos, Pieter .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (05) :2519-2527
[2]  
[Anonymous], [No title captured]
[3]  
[Anonymous], P INT S NUCL EL VERS
[4]   Eu2+-doped Ba2Csl5, a new high-performance scintillator [J].
Bourret-Courchesne, E. D. ;
Bizarri, G. ;
Borade, R. ;
Yan, Z. ;
Hanrahan, S. M. ;
Gundiah, G. ;
Chaudhry, A. ;
Canning, A. ;
Derenzo, S. E. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 612 (01) :138-142
[5]   The CMS experiment at the CERN LHC [J].
Chatrchyan, S. ;
Hmayakyan, G. ;
Khachatryan, V. ;
Sirunyan, A. M. ;
Adam, W. ;
Bauer, T. ;
Bergauer, T. ;
Bergauer, H. ;
Dragicevic, M. ;
Eroe, J. ;
Friedl, M. ;
Fruehwirth, R. ;
Ghete, V. M. ;
Glaser, P. ;
Hartl, C. ;
Hoermann, N. ;
Hrubec, J. ;
Haensel, S. ;
Jeitler, M. ;
Kastner, K. ;
Krammer, M. ;
de Abril, I. Magrans ;
Markytan, M. ;
Mikulec, I. ;
Neuherz, B. ;
Noebauer, T. ;
Oberegger, M. ;
Padrta, M. ;
Pernicka, M. ;
Porth, P. ;
Rohringer, H. ;
Schmid, S. ;
Schreiner, T. ;
Stark, R. ;
Steininger, H. ;
Strauss, J. ;
Taurok, A. ;
Uhl, D. ;
Waltenberger, W. ;
Walzel, G. ;
Widl, E. ;
Wulz, C. -E. ;
Petrov, V. ;
Prosolovich, V. ;
Chekhovsky, V. ;
Dvornikov, O. ;
Emeliantchik, I. ;
Litomin, A. ;
Makarenko, V. ;
Marfin, I. .
JOURNAL OF INSTRUMENTATION, 2008, 3
[6]   Strontium and barium iodide high light yield scintillators [J].
Cherepy, Nerine J. ;
Hull, Giulia ;
Drobshoff, Alexander D. ;
Payne, Stephen A. ;
Van Loef, Edgar ;
Wilson, Cody M. ;
Shah, Kanai S. ;
Roy, Utpal N. ;
Burger, Arnold ;
Boatner, Lynn A. ;
Choong, Woon-Seng ;
Moses, William W. .
APPLIED PHYSICS LETTERS, 2008, 92 (08)
[7]   Advances in yield calibration of scintillators [J].
de Haas, Johan T. M. ;
Dorenbos, Pieter .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) :1086-1092
[8]   Non-proportionality in the scintillation response and the energy resolution obtainable with scintillation crystals [J].
Dorenbos, P ;
deHaas, JTM ;
vanEijk, CWE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (06) :2190-2202
[9]   Fundamental Limitations in the Performance of Ce3+-, Pr3+-, and Eu2+-Activated Scintillators [J].
Dorenbos, Pieter .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (03) :1162-1167
[10]   ALKALI HALIDE SCINTILLATION COUNTERS [J].
HOFSTADTER, R .
PHYSICAL REVIEW, 1948, 74 (01) :100-101