Reproducibility of prostate volume measurements from transrectal ultrasonography by an automated and a manual technique

被引:34
作者
Aarnink, RG
DelaRosette, JJMCH
Debruyne, FMJ
Wijkstra, H
机构
[1] Department of Urology, University Hospital, Nijmegen
[2] University Hospital, Department of Urology, Biomedical Engineering Unit, 6500 HB Nijmegen
来源
BRITISH JOURNAL OF UROLOGY | 1996年 / 78卷 / 02期
关键词
step-section planimetry; reproducibility; prostate volume; ultrasonography;
D O I
10.1046/j.1464-410X.1996.08213.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objective To assess the reproducibility of estimates of prostate volume determined by planimetry from transrectal ultrasonography (TRUS) images. Materials and methods Two sequential sessions of images obtained by TRUS were obtained from 30 patients, with the ultrasound probe removed and inserted between the sessions. The stored images were outlined, both manually and by computer, and measurements of prostate volume obtained planimetrically. In addition, the ability of the urologist to accurately draw the contour was assessed by outlining predefined contours. Results The mean (so) variability of manual outlining between sessions was 3.5 (3.4)%, within one session was 1.7 (1.3)% and of computer outlining between sessions was 4.3 (3.8)%. Comparing the results of manual and computer outlining showed a mean (so) variability of 7.5 (5.6)%, with larger values obtained from computer outlining. The mean (so) variation in manually outlining predefined contours was 1.4 (1.4)%. Conclusions The variability of computer outlining was slightly higher than expected theoretically. The within-session variability was higher than the variation caused by errors in outlining predefined contours, indicating that the interpretation of TRUS images differed with time. Automated determination of prostate volume can save time during clinical investigation and the variability is within the clinically acceptable range of 5%.
引用
收藏
页码:219 / 223
页数:5
相关论文
共 9 条
  • [1] AARNICK RG, 1996, IN PRESS PROSTATE
  • [2] A PRACTICAL CLINICAL METHOD FOR CONTOUR DETERMINATION IN ULTRASONOGRAPHIC PROSTATE IMAGES
    AARNINK, RG
    GIESEN, RJB
    HUYNEN, AL
    DELAROSETTE, JJMCH
    DEBRUYNE, FMJ
    WIJKSTRA, H
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (08) : 705 - 717
  • [3] PLANIMETRIC VOLUMETRY OF THE PROSTATE - HOW ACCURATE IS IT
    AARNINK, RG
    GIESEN, RJB
    DELAROSETTE, JJMCH
    HUYNEN, AL
    DEBRUYNE, FMJ
    WIJKSTRA, H
    [J]. PHYSIOLOGICAL MEASUREMENT, 1995, 16 (03) : 141 - 150
  • [4] Aarnink RG, 1996, EUR UROL, V29, P399
  • [5] AUTOMATED PROSTATE VOLUME DETERMINATION WITH ULTRASONOGRAPHIC IMAGING
    AARNINK, RG
    HUYNEN, AL
    GIESEN, RJB
    DELAROSETTE, JJMCH
    DEBRUYNE, FMJ
    WIJKSTRA, H
    [J]. JOURNAL OF UROLOGY, 1995, 153 (05) : 1549 - 1554
  • [6] ULTRASONICALLY DETERMINED PATTERNS OF ENLARGEMENT IN BENIGN PROSTATIC HYPERPLASIA
    COLLINS, GN
    LEE, RJ
    RUSSELL, EB
    RAAB, GM
    HEHIR, M
    [J]. BRITISH JOURNAL OF UROLOGY, 1993, 71 (04): : 451 - 456
  • [7] DENIS LJ, 1995, BRIT J UROL, V76, P17
  • [8] STONE N N, 1991, Journal of Endourology, V5, P251, DOI 10.1089/end.1991.5.251
  • [9] DETERMINATION OF PROSTATE VOLUME BY TRANSRECTAL ULTRASOUND
    TERRIS, MK
    STAMEY, TA
    [J]. JOURNAL OF UROLOGY, 1991, 145 (05) : 984 - 987