Virtual simulation with Sim&Size software for Pipeline Flex Embolization: evaluation of the technical and clinical impact

被引:26
作者
Piergallini, Lorenzo [1 ,2 ]
Cagnazzo, Federico [2 ]
Conte, Giorgio [3 ]
Dargazanli, Cyril [2 ]
Derraz, Imad [2 ]
Lefevre, Pierre-Henri [2 ]
Gascou, Gregory [2 ]
Mantilla, Daniel [2 ]
Riquelme, Carlos [2 ]
Bonafe, Alain [2 ]
Costalat, Vincent [2 ]
机构
[1] Hosp Civils Lyon, Neuroradiol, F-69002 Lyon, France
[2] CHU Montpellier, Neuroradiol, Montpellier, France
[3] Fdn IRCCS Ca Granda Osped Maggiore Milano Policli, Neuroradiol, Milan, Italy
关键词
aneurysm; flow diverter; device; brain; FLOW-DIVERTOR; DEPLOYMENT; ANEURYSMS; DEVICE;
D O I
10.1136/neurintsurg-2020-015813
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Introduction During flow diversion, the choice of the length, diameter, and location of the deployed stent are critical for the success of the procedure. Sim&Size software, based on the three-dimensional rotational angiography (3D-RA) acquisition, simulates the release of the stent, suggesting optimal sizing, and displaying the degree of the wall apposition. Objective To demonstrate technical and clinical impacts of the Sim&Size simulation during treatment with the Pipeline Flex Embolization Device. Methods Consecutive patients who underwent aneurysm embolization with Pipeline at our department were retrospectively enrolled (January 2015-December 2017) and divided into two groups: treated with and without simulation. Through univariate and multivariate models, we evaluated: (1) rate of corrective intervention for non-optimal stent placement, (2) duration of intervention, (3) radiation dose, and (4) stent length. Results 189 patients, 95 (50.2%) without and 94 (49.7%) with software assistance were analyzed. Age, sex, comorbidities, aneurysm characteristics, and operator's experience were comparable among the two groups. Procedures performed with the software had a lower rate of corrective intervention (9% vs 20%, p=0.036), a shorter intervention duration (46 min vs 52 min, p=0.002), a lower median radiation dose (1150 mGy vs 1558 mGy, p<0.001), and a shorter stent length (14 mm vs 16 mm, p<0.001). Conclusions In our experience, the use of the virtual simulation during Pipeline treatment significantly reduced the need for corrective intervention, the procedural time, the radiation dose, and the length of the stent.
引用
收藏
页码:968 / 973
页数:6
相关论文
共 19 条
[1]   Comparison of effective doses obtained from dose-area product and air kerma measurements in interventional radiology [J].
Bor, D ;
Sancak, T ;
Olgar, T ;
Elcim, Y ;
Adanali, A ;
Sanlidilek, U ;
Akyar, S .
BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916) :315-322
[2]   Endovascular Treatment of Intracranial Aneurysms With Flow Diverters A Meta-Analysis [J].
Brinjikji, Waleed ;
Murad, Mohammad H. ;
Lanzino, Giuseppe ;
Cloft, Harry J. ;
Kallmes, David F. .
STROKE, 2013, 44 (02) :442-447
[3]   Treatment of Wide-Neck Intracranial Aneurysms with the Woven EndoBridge Device Associated with Stenting: A Single-Center Experience [J].
Cagnazzo, F. ;
Ahmed, R. ;
Dargazanli, C. ;
Lefevre, P-H ;
Gascou, G. ;
Derraz, I ;
Kalmanovich, S. A. ;
Riquelme, C. ;
Bonafe, A. ;
Costalat, V .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (05) :820-826
[4]   Patency of the supraclinoid internal carotid artery branches after flow diversion treatment. A meta-analysis [J].
Cagnazzo, Federico ;
Lefevre, Pierre-Henri ;
Mantilla, Daniel ;
Rouchaud, Aymeric ;
Morganti, Riccardo ;
Perrini, Paolo ;
Di Carlo, Davide ;
Dargazanli, Cyril ;
Gascou, Gregory ;
Riquelme, Carlos ;
Bonafe, Alain ;
Costalat, Vincent .
JOURNAL OF NEURORADIOLOGY, 2019, 46 (01) :9-14
[5]   CLINICAL EVENTS FOLLOWING NEUROANGIOGRAPHY - A PROSPECTIVE-STUDY [J].
DION, JE ;
GATES, PC ;
FOX, AJ ;
BARNETT, HJM ;
BLOM, RJ .
STROKE, 1987, 18 (06) :997-1004
[6]   Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections [J].
Fischer, Sebastian ;
Vajda, Zsolt ;
Perez, Marta Aguilar ;
Schmid, Elisabeth ;
Hopf, Nikolai ;
Baezner, Hansjoerg ;
Henkes, Hans .
NEURORADIOLOGY, 2012, 54 (04) :369-382
[7]   Silent embolism after stent-assisted coiling of cerebral aneurysms: diffusion-weighted MRI study of 75 cases [J].
Hahnemann, M. L. ;
Ringelstein, A. ;
Sandalcioglu, I. E. ;
Goericke, S. ;
Moenninghoff, C. ;
Wanke, I. ;
Forsting, M. ;
Sure, U. ;
Schlamann, M. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2014, 6 (06) :461-465
[8]  
Harrell F. E., 2001, REGRESSION MODELING
[9]   Cellular Mechanisms of Aneurysm Occlusion after Treatment with a Flow Diverter [J].
Kadirvel, Ramanathan ;
Ding, Yong-Hong ;
Dai, Daying ;
Rezek, Issa ;
Lewis, Debra A. ;
Kallmes, David F. .
RADIOLOGY, 2014, 270 (02) :394-399
[10]   Computer modeling of deployment and mechanical expansion of neurovascular flow diverter in patient-specific intracranial aneurysms [J].
Ma, Ding ;
Dargush, Gary F. ;
Natarajan, Sabareesh K. ;
Levy, Elad I. ;
Siddiqui, Adnan H. ;
Meng, Hui .
JOURNAL OF BIOMECHANICS, 2012, 45 (13) :2256-2263