Accuracy of automated patient positioning in CT using a 3D camera for body contour detection

被引:44
作者
Booij, Ronald [1 ]
Budde, Ricardo P. J. [1 ]
Dijkshoorn, Marcel L. [1 ]
van Straten, Marcel [1 ]
机构
[1] Erasmus MC, Dept Radiol & Nucl Med, POB 2240, NL-3000 CA Rotterdam, Netherlands
关键词
Tomography; x-ray computed; Health physics; Radiation dosage; Diagnostic imaging; IMAGE NOISE; COMPUTED-TOMOGRAPHY; IMPACT; MDCT;
D O I
10.1007/s00330-018-5745-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo assess the accuracy of a 3D camera for body contour detection and patient positioning in CT compared to routine manual positioning by radiographers.Methods and materialsFour hundred twenty-three patients that underwent CT of the head, thorax, and/or abdomen on a scanner with manual table height selection and 254 patients on a scanner with table height suggestion by a 3D camera were retrospectively included. Within the camera group, table height suggestion was based on infrared body contour detection and fitting of a scalable patient model to the 3D data. Proper positioning was defined as the ideal table height at which the scanner isocenter coincides with the patient's isocenter. Patient isocenter was computed by automatic skin contour extraction in each axial image and averaged over all images. Table heights suggested by the camera and selected by the radiographer were compared with the ideal height.ResultsMedian (interquartile range) absolute table height deviation in millimeter was 12.0 (21.6) for abdomen, 12.2 (12.0) for head, 13.4 (17.6) for thorax-abdomen, and 14.7 (17.3) for thorax CT scans positioned by radiographers. The deviation was significantly less (p<0.01) for the 3D camera at 6.3 (6.9) for abdomen, 9.5 (6.8) for head, 6.0 (6.1) for thorax-abdomen, and 5.4 (6.4)mm for thorax.ConclusionA 3D camera for body contour detection allows for accurate patient positioning, thereby outperforming manual positioning done by radiographers, resulting in significantly smaller deviations from the ideal table height. However, radiographers remain indispensable when the system fails or in challenging cases.Key Points center dot A 3D camera for body contour detection allows for accurate patient positioning.center dot A 3D camera outperformed radiographers in patient positioning in CT.center dot Deviation from ideal table height was more extreme for patients positioned by radiographers for all body parts.
引用
收藏
页码:2079 / 2088
页数:10
相关论文
共 18 条
[1]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[2]   Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies [J].
Habibzadeh, M. A. ;
Ay, M. R. ;
Asl, A. R. Kamali ;
Ghadiri, H. ;
Zaidi, H. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (03) :191-199
[3]   Variability of MDCT Dose Due to Technologist Performance: Impact of Posteroanterior Versus Anteroposterior Localizer Image and Table Height With Use of Automated Tube Current Modulation [J].
Harri, Peter A. ;
Moreno, Courtney C. ;
Nelson, Rendon C. ;
Fani, Negar ;
Small, William C. ;
Duong, Anh ;
Tang, Xiangyang ;
Applegate, Kimberly E. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (02) :377-386
[4]   Effect of Patient Centering on Patient Dose and Image Noise in Chest CT [J].
Kaasalainen, Touko ;
Palmu, Kirsi ;
Reijonen, Vappu ;
Kortesniemi, Mika .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (01) :123-130
[5]   Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT-phantom study [J].
Kaasalainen, Touko ;
Palmu, Kirsi ;
Lampinen, Anniina ;
Kortesniemi, Mika .
PEDIATRIC RADIOLOGY, 2013, 43 (06) :673-684
[6]   Application of automatic vertical positioning software to reduce radiation exposure in multidetector row computed tomography of the chest [J].
Li, Jianhai ;
Udayasankar, Unni K. ;
Toth, Thomas L. ;
Small, William C. ;
Kalra, Mannudeep K. .
INVESTIGATIVE RADIOLOGY, 2008, 43 (06) :447-452
[7]   Automatic patient centering for MDCT: Effect on radiation dose [J].
Li, Jianhai ;
Udayasankar, Unni K. ;
Toth, Thomas L. ;
Seamans, John ;
Small, William C. ;
Kalra, Mannudeep K. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (02) :547-552
[8]   The effects of patient positioning when interpreting CT dose metrics: A phantom study [J].
Marsh, Rebecca M. ;
Silosky, Michael S. .
MEDICAL PHYSICS, 2017, 44 (04) :1514-1524
[9]   Estimating patient dose from CT exams that use automatic exposure control: Development and validation of methods to accurately estimate tube current values [J].
McMillan, Kyle ;
Bostani, Maryam ;
Cagnon, Christopher H. ;
Yu, Lifeng ;
Leng, Shuai ;
McCollough, Cynthia H. ;
McNitt-Gray, Michael F. .
MEDICAL PHYSICS, 2017, 44 (08) :4262-4275
[10]   Trends and patterns of computed tomography scan use among children in The Netherlands: 1990-2012 [J].
Meulepas, Johanna M. ;
Smets, Anne M. J. B. ;
Nievelstein, Rutger A. J. ;
Gradowska, Patrycja ;
Verbeke, Jonathan ;
Holscher, Herma C. ;
Rutten, Matthieu J. C. M. ;
Kieft, Mariette ;
Ronckers, Cecile M. ;
Hauptmann, Michael .
EUROPEAN RADIOLOGY, 2017, 27 (06) :2426-2433