Accuracy of automated patient positioning in CT using a 3D camera for body contour detection

被引:41
|
作者
Booij, Ronald [1 ]
Budde, Ricardo P. J. [1 ]
Dijkshoorn, Marcel L. [1 ]
van Straten, Marcel [1 ]
机构
[1] Erasmus MC, Dept Radiol & Nucl Med, POB 2240, NL-3000 CA Rotterdam, Netherlands
关键词
Tomography; x-ray computed; Health physics; Radiation dosage; Diagnostic imaging; IMAGE NOISE; COMPUTED-TOMOGRAPHY; IMPACT; MDCT;
D O I
10.1007/s00330-018-5745-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo assess the accuracy of a 3D camera for body contour detection and patient positioning in CT compared to routine manual positioning by radiographers.Methods and materialsFour hundred twenty-three patients that underwent CT of the head, thorax, and/or abdomen on a scanner with manual table height selection and 254 patients on a scanner with table height suggestion by a 3D camera were retrospectively included. Within the camera group, table height suggestion was based on infrared body contour detection and fitting of a scalable patient model to the 3D data. Proper positioning was defined as the ideal table height at which the scanner isocenter coincides with the patient's isocenter. Patient isocenter was computed by automatic skin contour extraction in each axial image and averaged over all images. Table heights suggested by the camera and selected by the radiographer were compared with the ideal height.ResultsMedian (interquartile range) absolute table height deviation in millimeter was 12.0 (21.6) for abdomen, 12.2 (12.0) for head, 13.4 (17.6) for thorax-abdomen, and 14.7 (17.3) for thorax CT scans positioned by radiographers. The deviation was significantly less (p<0.01) for the 3D camera at 6.3 (6.9) for abdomen, 9.5 (6.8) for head, 6.0 (6.1) for thorax-abdomen, and 5.4 (6.4)mm for thorax.ConclusionA 3D camera for body contour detection allows for accurate patient positioning, thereby outperforming manual positioning done by radiographers, resulting in significantly smaller deviations from the ideal table height. However, radiographers remain indispensable when the system fails or in challenging cases.Key Points center dot A 3D camera for body contour detection allows for accurate patient positioning.center dot A 3D camera outperformed radiographers in patient positioning in CT.center dot Deviation from ideal table height was more extreme for patients positioned by radiographers for all body parts.
引用
收藏
页码:2079 / 2088
页数:10
相关论文
共 50 条
  • [1] Accuracy of automated patient positioning in CT using a 3D camera for body contour detection
    Ronald Booij
    Ricardo P.J. Budde
    Marcel L. Dijkshoorn
    Marcel van Straten
    European Radiology, 2019, 29 : 2079 - 2088
  • [2] Automated patient positioning in CT using a 3D camera for body contour detection: accuracy in pediatric patients
    Ronald Booij
    Marcel van Straten
    Andreas Wimmer
    Ricardo P.J. Budde
    European Radiology, 2021, 31 : 131 - 138
  • [3] Automated patient positioning in CT using a 3D camera for body contour detection: accuracy in pediatric patients
    Booij, Ronald
    van Straten, Marcel
    Wimmer, Andreas
    Budde, Ricardo P. J.
    EUROPEAN RADIOLOGY, 2021, 31 (01) : 131 - 138
  • [4] Influence of breathing state on the accuracy of automated patient positioning in thoracic CT using a 3D camera for body contour detection
    Booij, Ronald
    van Straten, Marcel
    Wimmer, Andreas
    Budde, Ricardo P. J.
    EUROPEAN RADIOLOGY, 2022, 32 (01) : 442 - 447
  • [5] Influence of breathing state on the accuracy of automated patient positioning in thoracic CT using a 3D camera for body contour detection
    Ronald Booij
    Marcel van Straten
    Andreas Wimmer
    Ricardo P. J. Budde
    European Radiology, 2022, 32 : 442 - 447
  • [6] Patient Positioning in CT using a 3D Camera
    Graewert, Stephanie
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2020, 192 (02):
  • [7] Use of a 3D camera for automated patient positioning for chest-abdomen-pelvis CT scans: Effect on positioning accuracy and patient dose
    Nelson, R.
    Harkin, S.
    Iball, G.
    RADIOGRAPHY, 2024, 30 (04) : 1060 - 1067
  • [8] Radiation dose reduction, improved isocenter accuracy and CT scan time savings with automatic patient positioning by a 3D camera
    Dane, Bari
    O'Donnell, Thomas
    Liu, Shu
    Vega, Emilio
    Mohammed, Sharon
    Singh, Vivek
    Kapoor, Ankur
    Megibow, Alec
    EUROPEAN JOURNAL OF RADIOLOGY, 2021, 136
  • [9] Automated CT Lung Cancer Screening Workflow Using 3D Camera
    Teixeira, Brian
    Singh, Vivek
    Tamersoy, Birgi
    Prokein, Andreas
    Kapoor, Ankur
    MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION, MICCAI 2023, PT VII, 2023, 14226 : 423 - 431
  • [10] Accuracy of contour lines using 3D bands
    Mozas-Calvache, Antonio T.
    Antonio Urena-Camara, Manuel
    Luis Perez-Garcia, Jose
    INTERNATIONAL JOURNAL OF GEOGRAPHICAL INFORMATION SCIENCE, 2013, 27 (12) : 2362 - 2374