Radiological and Clinical Differences between Tinavi Orthopedic Robot and O-Arm Navigation System in Thoracolumbar Screw Implantation for Reconstruction of Spinal Stability

被引:25
作者
Du, Jinpeng [1 ]
Gao, Lin [1 ]
Huang, Dageng [1 ]
Shan, Lequn [1 ]
Wang, Wentao [1 ]
Fan, Yong [1 ]
Hao, Dingjun [1 ]
Yan, Liang [1 ]
机构
[1] Xi An Jiao Tong Univ, Dept Spine Surg, Honghui Hosp, Xian, Shaanxi, Peoples R China
来源
MEDICAL SCIENCE MONITOR | 2020年 / 26卷
关键词
Bone Screws; Dimensional Measurement Accuracy; Patient Navigation; Robotics; Spine; C-ARM; ACCURACY; PLACEMENT; ASSISTANCE; INSERTION; FIXATION; LUMBAR;
D O I
10.12659/MSM.924770
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: Pedicle screw fixation is one of the most commonly used methods in spine surgery. We introduce a surgical robot system from China based on 3-dimensional fluoroscopy imaging and compare it with the commonly used O-arm navigation system. We study the differences in accuracy, safety, and clinical effect in auxiliary pedicle screw fixation. Material/Methods: Patients who underwent thoracolumbar internal fixation in our hospital from 2017 to 2019 were divided into a robot and navigation group according to whether surgery was assisted by the Tinavi orthopedic robot or O-arm navigation system. Imaging data of patients were searched from the image system and accuracy of screw implantation was measured by Rampersaud A to D grade classification. Deviation sagittal, deviation transversal, and facet joint violation were also measured and calculated. Results: In total, 306 patients were included: 136 patients in the robot group with 760 screws implanted; 166 patients in the navigation group with 908 screws implanted. The accuracy of "perfect" and "clinically acceptable" pedicle screw implantation was 96.2% and 99.6%, respectively, in the robot group and 90.5% and 96.7%, respectively, in the navigation group, with a significant difference between the 2 groups (P<0.05). The sagittal and transversal deviations in the robot group were significantly less than those in the navigation group (P<0.05). Conclusions: The Tinavi orthopedic robot can significantly improve surgical accuracy and safety of pedicle screw fixation, as compared with that of O-arm navigation technology, without increasing complications. It shows great potential in clinical application.
引用
收藏
页数:10
相关论文
共 28 条
[1]   Clinical Acceptance and Accuracy Assessment of Spinal Implants Guided With SpineAssist Surgical Robot Retrospective Study [J].
Devito, Dennis P. ;
Kaplan, Leon ;
Dietl, Rupert ;
Pfeiffer, Michael ;
Horne, Dale ;
Silberstein, Boris ;
Hardenbrook, Mitchell ;
Kiriyanthan, George ;
Barzilay, Yair ;
Bruskin, Alexander ;
Sackerer, Dieter ;
Alexandrovsky, Vitali ;
Stueer, Carsten ;
Burger, Ralf ;
Maeurer, Johannes ;
Gordon, Donald G. ;
Schoenmayr, Robert ;
Friedlander, Alon ;
Knoller, Nachshon ;
Schmieder, Kirsten ;
Pechlivanis, Ioannis ;
Kim, In-Se ;
Meyer, Bernhard ;
Shoham, Moshe .
SPINE, 2010, 35 (24) :2109-2115
[2]  
Fan MX, 2003, J CLIN ORTHOP RES, P213
[3]   Comparison of Accuracy of Pedicle Screw Insertion Among 4 Guided Technologies in Spine Surgery [J].
Fan, Yong ;
Du, Jinpeng ;
Zhang, Jianan ;
Liu, Shichang ;
Xue, Xukai ;
Huang, Yunfei ;
Zhang, Jing ;
Hao, Dingjun .
MEDICAL SCIENCE MONITOR, 2017, 23 :5960-5968
[4]   Clinically relevant complications related to pedicle screw placement in thoracolumbar surgery and their management: a literature review of 35,630 pedicle screws [J].
Gautschi, Oliver P. ;
Schatlo, Bawarjan ;
Schaller, Karl ;
Tessitore, Enrico .
NEUROSURGICAL FOCUS, 2011, 31 (04)
[5]  
Han X, 2019, J NEUROSURG SPINE
[6]   Clinical Assessment of Percutaneous Lumbar Pedicle Screw Placement Using the O-Arm Multidimensional Surgical Imaging System [J].
Houten, John K. ;
Nasser, Rani ;
Baxi, Nrupen .
NEUROSURGERY, 2012, 70 (04) :990-995
[7]   Robotic-assisted pedicle screw placement: lessons learned from the first 102 patients [J].
Hu, Xiaobang ;
Ohnmeiss, Donna D. ;
Lieberman, Isador H. .
EUROPEAN SPINE JOURNAL, 2013, 22 (03) :661-666
[8]   Minimally Invasive Robotic Versus Open Fluoroscopic-guided Spinal Instrumented Fusions A Randomized Controlled Trial [J].
Hyun, Seung-Jae ;
Kim, Ki-Jeong ;
Jahng, Tae-Ahn ;
Kim, Hyun-Jib .
SPINE, 2017, 42 (06) :353-358
[9]  
Kim HJ, 2016, INT J MED ROBOT, V13, P3
[10]   Comparison of adjacent segment degeneration after successful posterolateral fusion with unilateral or bilateral pedicle screw instrumentation: a minimum 10-year follow-up [J].
Kim, Tae-Hwan ;
Lee, Byung H. ;
Moon, Seong-Hwan ;
Lee, Seung-Hwan ;
Lee, Hwan-Mo .
SPINE JOURNAL, 2013, 13 (10) :1208-1216