Flow volume and shunt quantification in pediatric congenital heart disease by real-time magnetic resonance velocity mapping -: A validation study

被引:64
作者
Körperich, H
Gieseke, J
Barth, P
Hoogeveen, R
Esdorn, H
Peterschröder, A
Meyer, H
Beerbaum, P
机构
[1] Ruhr Univ Bochum, Heart & Diabet Ctr Northrhine Westfalia, Clin Congenital Heart Dis, D-32545 Bad Oeynhausen, Germany
[2] Ruhr Univ Bochum, Heart & Diabet Ctr Northrhine Westfalia, Inst Magnet Resonance Imaging, D-32545 Bad Oeynhausen, Germany
[3] Philips Med Syst, Best, Netherlands
关键词
magnetic resonance imaging; blood flow; heart defects; congenital; pediatrics; shunts;
D O I
10.1161/01.CIR.0000126494.66859.A2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Flow quantification in real time by phase-contrast MRI (PC-MRI) may provide unique hemodynamic information in congenital heart disease, but available techniques have important limitations. We sought to validate a novel real-time magnetic resonance flow sequence in children. Methods and Results-In 14 pediatric patients (mean age 5.2 +/- 2.0 years) with cardiac left-to-right shunt, pulmonary (Q(p)) and aortic (Q(s)) flow rates were determined by nontriggered free-breathing real-time PC-MRI with single-shot echo-planar imaging combined with sensitivity encoding, which yielded 25 phase images per second at 2.7 x 2.7-mm in-plane resolution (field of view 30x34 cm(2)). Over a 9.5-second period that included 2 to 5 respiratory cycles, 16.6+/-2.6 subsequent stroke volumes (range 13 to 22) were acquired in each vessel. Results were compared with conventional retrospectively ECG-gated PC-MRI. Mean Q(p)/Q(s) by conventional PC-MRI was 1.91+/-0.64, and it was 1.94+/-0.68 (mean+/-SD) by real-time PC-MRI. For blood flow rate through pulmonary artery and aorta, we found differences of 2% to 3% (Bland-Altman analysis), with lower limits of agreement of -11% to -13% (mean-2 SD) and upper limits of 18% to 19% (mean+2 SD), which demonstrated good agreement between both methods. Mean difference for Q(p)/Q(s) was 1%, with limits of agreement ranging between -18% and 22% (mean+/-2 SD). High repeatability but some flow overestimation was observed in vitro (pulsatile flow phantom) with real-time PC-MRI, whereas conventional PC-MRI was accurate. Beat-to-beat stroke-volume variation was 6.1+/-2.3% in vivo and 3.7+/-0.3% in vitro. Conclusions-Beat-to-beat quantification of pulmonary and aortic flows and hence left-to-right shunt within a few seconds is reliable by nontriggered real-time PC-MRI with echo-planar imaging and sensitivity encoding. Good spatial/temporal resolution and a large field of view may render the sequence valuable for multiple applications in congenital heart disease.
引用
收藏
页码:1987 / 1993
页数:7
相关论文
共 25 条
  • [1] DOPPLER ECHOCARDIOGRAPHIC ANALYSIS OF CARDIAC FLOW DURING THE MUELLER MANEUVER
    ANDREAS, S
    WERNER, GS
    SOLD, G
    WIEGAND, V
    KREUZER, H
    [J]. EUROPEAN JOURNAL OF CLINICAL INVESTIGATION, 1991, 21 (01) : 72 - 76
  • [2] Bakker CJG, 1999, JMRI-J MAGN RESON IM, V9, P119, DOI 10.1002/(SICI)1522-2586(199901)9:1<119::AID-JMRI16>3.3.CO
  • [3] 2-6
  • [4] Rapid left-to-right shunt quantification in children by phase-contrast magnetic resonance imaging combined with sensitivity encoding
    Beerbaum, P
    Körperich, H
    Gieseke, J
    Barth, P
    Peuster, M
    Meyer, H
    [J]. CIRCULATION, 2003, 108 (11) : 1355 - 1361
  • [5] Beerbaum P, 2001, CIRCULATION, V103, P2476
  • [6] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [7] ANALYSIS OF FLOW EFFECTS IN ECHO-PLANAR IMAGING
    BUTTS, K
    RIEDERER, SJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (03): : 285 - 293
  • [8] SIGNAL-TO-NOISE IN PHASE-ANGLE RECONSTRUCTION - DYNAMIC-RANGE EXTENSION USING PHASE REFERENCE OFFSETS
    CONTURO, TE
    SMITH, GD
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) : 420 - 437
  • [9] THEORETICAL ASPECTS OF MOTION SENSITIVITY AND COMPENSATION IN ECHO-PLANAR IMAGING
    DUERK, JL
    SIMONETTI, OP
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (06): : 643 - 650
  • [10] Phase-contrast echo-planar MR imaging: Real-time quantification of flow and velocity patterns in the thoracic vessels induced by Valsalva's maneuver
    Eichenberger, AC
    Schwitter, J
    Mckinnon, GC
    Debatin, JF
    vonSchulthess, GK
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (06): : 648 - 655